Skip to main content
  • Original Paper
  • Open access
  • Published:

An anatomically shaped medial meniscus prosthesis is able to partially restore the contact mechanics of the meniscectomized knee joint

Abstract

Purpose

The aim of this study was to determine whether a flexible medial meniscus prosthesis is more capable of sharing loads with the direct tibiofemoral cartilage contact than the stiffer first-generation prosthesis. Additionally, the effect of the prosthesis on the tibial pressure distribution after total meniscectomy was investigated.

Methods

In an artificial knee joint, the relative amounts of load transferred through both meniscus prostheses and the direct tibiofemoral contact were assessed with pressure-sensitive sensors.

Additionally, six cadaveric knee joints were loaded in a physiological environment. Tibial contact pressures were measured with an intact native meniscus, after total meniscectomy and after implantation of the second-generation meniscus prosthesis.

Results

Whereas the first generation of the meniscus prosthesis transferred virtually all the load from femur to tibia, the second-generation prosthesis allowed for load sharing with the direct tibiofemoral contact.

No differences in load sharing were found between the native meniscus and the second-generation meniscus prosthesis. The prosthesis decreased peak and mean pressures on the medial tibial cartilage compared to meniscectomy. No significant differences in pressure were found between the native meniscus and the meniscus prosthesis.

Conclusions

The second-generation meniscus prosthesis presented in this study can share loads with the direct tibiofemoral contact, a characteristic that the first-generation prosthesis did not have.

The flexible meniscus prosthesis significantly reduces the contact pressures on the medial tibial plateau after total meniscectomy. Although the biomechanical performance of the native meniscus could not be reproduced completely, the meniscus prosthesis may have the potential to relieve post-meniscectomy pain symptoms.

Introduction

The most common knee injuries are meniscal tears [15], which can be caused either by acute trauma or by degenerative processes [6]. This results in over 500,000 partial or total meniscectomies in the U.S. [1, 16] and Europe [17] every year. However, the removal of meniscal tissue, is a strong predictor of knee osteoarthritis and pain [4, 5, 18]. Tibiofemoral contact pressures are known to increase after (sub-)total meniscectomy [2, 14, 15, 21, 31], which may be one of the causes of the pain experienced by patients after undergoing meniscectomy. Meniscal allograft transplantation (MAT) decreases tibiofemoral contact pressures [11, 27] and provides an off-the-shelf possibility to relieve pain and delay the onset of knee osteoarthritis [23]. However, the availability of size-matched meniscal allografts is limited and the long-term clinical effects may be affected by extrusion [13, 23] or shrinkage of the allograft [12, 19].

An anatomically shaped, polycarbonate urethane (PCU) meniscus prosthesis was developed, based on the average meniscus geometry obtained from MRI scans of 35 healthy subjects [26]. It was designed to restore the biomechanical function of the native medial meniscus without the disadvantages of MAT. Pre-clinical studies showed that prototypes of the prosthesis were able to reduce the tibiofemoral pressures relative to the meniscectomized knee joint, to similar levels as an allograft [10, 27]. None of these studies, however, involved testing in a wet environment at body temperature, i.e. in physiological conditions.

The first generation of the prosthesis was a two-component implant, with a stiff reinforcing core to limit extrusion and a soft cartilage-contacting layer (Fig. 1). Two titanium fixation screws were developed to secure the prosthesis to the anatomical attachment sites of the meniscal horns on the tibial plateau. This resulted in a relatively rigid prosthesis, which requires a perfect fit and position due to the lack of geometrical adaptability, to conform with the femoral surface geometry which changes during flexion and extension. This configuration was evaluated during a first-in-human clinical investigation in five patients, who had previously undergone meniscectomy [24]. The high in-plane stiffness of the prosthesis, in combination with the rigid fixation technique by means of screws, resulted in a deficit of knee flexion and extension and persistent pain in all patients. Four out of five prostheses were explanted, of which three were broken. It was concluded that, due to the rigidity of the prosthesis and its fixation, the compressive loads were transferred entirely through the prosthesis and not through the direct tibiofemoral contact [24]. This lack of load sharing increased the compressive and circumferential stresses in the prosthesis and ultimately resulted in structural failure of the prosthesis. Furthermore, it was expected that in the relatively high temperature inside the human body, the water uptake of the PCU would result in a more flexible prosthesis [7]. Less adaptation of the prosthesis than expected may have contributed to the failure mechanism described in the first-in-human clinical investigation. Earlier studies investigating the effect of water uptake and temperature on the prosthesis are lacking.

Fig. 1
figure 1

First-generation meniscus prosthesis, with a reinforcing core and soft outer layer

To overcome the issues that were identified during the first clinical investigation, the meniscus prosthesis and its fixation technique were adapted, resulting in a more flexible and adaptable device (Fig. 2). The first objective of this study was to assess the load sharing capabilities of the flexible second-generation prosthesis in an artificial knee joint. It was hypothesized that a more flexible prosthesis would allow for more load sharing with the direct tibiofemoral cartilage contact. In the same test, the effects of temperature and water uptake were investigated to obtain a better understanding of the material properties in different conditions. A more flexible prosthesis could potentially lead to excessive extrusion from the joint and, with that, possible loss of function. Therefore, the second objective of this study was to investigate the functional biomechanical performance of the second-generation medial meniscus prosthesis system in different cadaveric knee joints, measured by contact pressures on the tibial plateau. The prosthesis is expected to improve the contact mechanics compared to a total medial meniscectomy, in a similar way to the allograft.

Fig. 2
figure 2

Second-generation meniscus prosthesis, with strong horns, a flexible body and an anchoring tape

Materials and methods

This study consisted of two experiments in different test setups: In an artificial knee joint, the load sharing capabilities of the second-generation meniscus prosthesis (Fig. 2) were compared to the first generation (Fig. 1). The effects of water uptake of the meniscus prosthesis and environmental temperature were investigated as well. In the second test, axial loads were applied to cadaveric knee joints under physiological knee conditions. Contact mechanics on the medial and lateral tibial plateau were evaluated with pressure-sensitive sensors (Tekscan, South Boston, MA, USA, Fig. 3), in three situations (Fig. 4): With both intact native menisci (Fig. 5), after total medial meniscectomy and with the second-generation artificial medial meniscus prosthesis implanted.

Fig. 3
figure 3

Tekscan 4011 pressure-sensitive sensor

Fig. 4
figure 4

Schematic representation of the different situations tested in the cadaveric knee joints. a Native menisci. b Total medial meniscectomy. c Second-generation medial meniscus prosthesis. The arrows indicate the anterior (A), posterior (P), lateral (L) and medial (M) sides of the tibial plateau

Fig. 5
figure 5

Native medial meniscus, excised from a cadaveric knee joint

Meniscus prostheses

The rigid first generation of the medial meniscus prosthesis consists of a stiff core of Bionate® 75D polycarbonate urethane (PCU, DSM Biomedical, Berkeley, CA, USA), which extends into two meniscal horns of the same material. This reinforcing core is covered by a soft, cartilage-contacting outer layer of Bionate® II 80A PCU (Fig. 1). The horns of the meniscus prosthesis are fixed to the tibial plateau by using two dedicated fixation screws, made of a titanium alloy (Ti-6Al-7Nb).

For the design of the second-generation meniscus prosthesis (Fig. 2), two changes were implemented: Firstly, the stiff reinforcing core was removed, resulting in a single-component, soft and flexible meniscus body made of Bionate® II 80A PCU. The attachment horns of the prosthesis remained the same and are made of the stronger and stiffer Bionate® 75D PCU. Secondly, the titanium fixation screws were replaced by a polyethylene terephthalate (PET) anchoring tape. This tape is currently used clinically for fixation of ACL grafts and is secured on the anterior side of the tibia with polyether ether ketone (PEEK) anchoring screws.

All evaluated meniscus prostheses were manufactured by injection moulding of PCU, followed by EtO (ethylene oxide) sterilization. Since the approximately 1% water uptake is expected to strongly affect the mechanical properties of both materials, the wet-tested prostheses were pre-soaked for at least three weeks prior to testing [7]. Only average-sized (size 3) meniscus prostheses were used during this study, to match the selected cadaveric specimens.

Test setup—Artificial knee joint

As no test setup for assessing the load sharing between the meniscus prosthesis and the direct tibiofemoral contact has been described, an artificial setup was developed (Fig. 6). The medial tibial plateau of an average-sized knee joint was segmented from a CT-scan and machined from ultra-high molecular weight polyethylene (UHMWPE). The bottom side of the plateau was hollow, to reproduce the deformation that occurs in the tibial and femoral cartilage of the human knee joint [8]. The different meniscus prostheses could be placed on the tibial plateau by use of two pre-drilled fixation holes, suitable for both the screw and tape fixation techniques. The femoral part of the test setup consisted of an average-sized femur component of a total knee replacement, made of a cobalt chrome alloy (CoCr). Axial loads could be applied to the femur condyle by a hydraulic testing rig (MTS Systems, Eden Prairie, MN, USA). The test setup (Fig. 7) was placed inside a temperature-controlled water bath. The measurements were performed both dry at room temperature and in a water bath of 37 ± 2 °C.

Fig. 6
figure 6

Second-generation meniscus prosthesis in the artificial knee joint

Fig. 7
figure 7

First-generation meniscus prosthesis and sensor in the artificial knee joint

Test setup—Cadaveric knee joints

The femoral and tibial shafts of all specimens were shortened to approximately 12 and 10 cm respectively, to avoid any unnecessary effects of bending. Soft tissue was removed from the ends of the bones to enable potting of the bones with polymethyl methacrylate (PMMA) bone cement in stainless steel cups, which could be mounted in the test setup (Fig. 8). The setup was positioned in a water bath with a temperature of 37° ± 2 °C, to try to mimic the physiological conditions inside the knee joint [9]. It was decided to perform all tests in full extension, for better reproducibility of the alignment of the joints during testing and to maintain knee stability. The tibial cup was mounted in a dedicated support, using a steel rod oriented in the anteroposterior direction to make varus/valgus rotation of the knee joint possible. The axis of rotation was located approximately 5 mm medially of the centre, to allow for the unequal physiological load distribution between the medial and lateral compartment [9, 20]. Anteroposterior and mediolateral translations were not possible in this setup. Axial loads could be applied to the femoral bone cup by the MTS hydraulic testing rig.

Fig. 8
figure 8

Schematic representation of the cadaveric knee test setup: 1 Axial load applied by the testing rig. 2 Femoral bone cup, restricted in all directions except for proximodistal translation. 3 Tibial bone cup and 4 Steel rod of ø8mm, allowing small anteroposterior translations and 14° of varus and valgus rotation. 5 Support for the steel rod, restricted in all directions

Specimens

After obtaining ethical approval from the Anatomy Department, fresh frozen human cadaveric knee joints were obtained from the Radboud University Medical Centre (Nijmegen, The Netherlands). Anteroposterior and mediolateral radiographs were used to select one right and five left average-sized knee joints, to match the size of the pressure-sensitive sensors and the available meniscus prostheses. Specimens were excluded if they showed significant signs of osteoarthritis, such as osteophytes and joint space narrowing. The absence of osteoarthritis was confirmed during the surgical procedure.

Cadaveric surgery

All surgical activities were performed by an experienced orthopaedic knee surgeon (TvT). Firstly, all excessive skin, fat and muscle tissue was removed from the cadaveric knee joints. To improve visibility and to prevent folding of the pressure-sensitive sensor during insertion, the patella and the anterior and posterior parts of the joint capsule were removed. Care was taken to leave the collateral and cruciate ligaments intact, to maintain stability and physiological alignment of the knee joint. The meniscal horn attachments and the meniscocapsular and meniscotibial ligaments were left intact, while the periphery of the native medial and lateral menisci was detached from the tibial plateau. In this way, the pressure-sensitive sensor could be inserted underneath the menisci without folding. Detachment of the circumferential fixation is not expected to affect kinematics of the native meniscus [28], which is therefore expected to maintain its functional performance.

After the native condition was tested, the medial meniscus was completely removed by cutting both horn attachments and the attachment to the remainder of the joint capsule. After insertion of the pressure-sensitive sensor, the meniscectomy measurements were performed.

Subsequently, the second-generation medial meniscus prosthesis was implanted. Firstly, both anchoring tapes were led through the fixation holes of the prosthesis horns. A bone tunnel was drilled from the posterior horn attachment of the native meniscus to the anteromedial aspect of the tibia. By pulling the anchoring tape, the prosthesis was inserted into the knee joint, and the posterior horn was positioned above the tunnel. The optimal location of the anterior drill hole was determined with the knee in extension, thus preventing impingement or extension deficit. After drilling the anterior bone tunnel to the anterolateral aspect of the tibia, both anchoring tapes were tensioned and fixed in the tunnel by a PEEK anchoring screw. The pressure-sensitive sensor was positioned under the meniscus prosthesis before performing the final measurements, as shown in Fig. 9.

Fig. 9
figure 9

Second-generation meniscus prosthesis and sensor in a cadaveric knee joint

Loading protocol – Cadaveric knee joints

Before each series of measurements in the cadaveric knee joints, three pre-conditioning cycles of 1000 N were applied to allow the knee joint to find its natural alignment. Subsequently, the measurements were performed at the end of 120 s of constant axial loading of 500 and 1000 N respectively. Although 1000 N is below physiological loads, the load was not increased any further, to prevent breakage of the cadaveric bones, which has occurred in pilot experiments. In between load cycles, the joint was always unloaded for at least 30 s.

Loading protocol – Artificial knee joint

Approximately 70% of the physiological load is transferred through the medial joint compartment [20, 32]. Therefore, axial loads of 350 and 700 N were applied in the artificial knee joint, which only consist of a medial tibial plateau, to allow comparison with the 500 and 1000 N applied in the cadaveric tests.

Pressure measurements

For this study, piezoelectric pressure mapping sensors of type 4011 (Fig. 3) were used for the measurements in both the artificial and the cadaveric knee joints. Prior to use, the sensors were pre-conditioned and calibrated by applying five different pressures between 0 and 7 MPa. In between the measurements of the different meniscal conditions, additional calibration measurements were performed to evaluate and correct for any possible loss of sensor sensitivity.

From the obtained pressure maps, the peak pressure, mean pressure and contact area on the medial tibial plateau were determined. Furthermore, the load sharing ratio between the medial meniscus or medial meniscus prosthesis and the direct tibiofemoral cartilage contact, i.e. the percentage of the medial compartment load that is transferred through the meniscus, was estimated from the pressure maps. Finally, the pressures on the lateral tibial plateau of the cadaveric knee joints were determined, to assess a potential shift in the load distribution from one joint compartment to another.

Statistical analysis

Based on a previous study [27] and preliminary results from pilot testing, a sample size calculation was performed to determine the number of cadaveric specimens required. Using a power of 80% and a significance level of 0.05, six knees are required to detect contact pressure differences of 1.0 MPa, assuming a standard deviation of 0.5 MPa.

Linear mixed models were used to study the effect of meniscal condition (i.e. native meniscus, meniscectomy and prosthesis) of the medial knee compartment on contact mechanics. Medial peak pressure, mean pressure and contact area were analysed separately. In addition, the percentage of the load transferred through the meniscus or prosthesis and directly through the cartilage was analysed. Finally, the lateral mean pressure was analysed. The models included specimen as a random factor and all other variables that applied to the outcome measure under analysis (i.e. meniscal condition and axial load) as fixed factors. A random intercept was included to account for the specimen’s individual response to each experimental condition. Interaction terms between the fixed factors were also evaluated. Pairwise comparisons between the different levels of the fixed variables were performed by Tukey’s tests that were Bonferroni-corrected to account for multiple comparisons. 95% confidence intervals were determined and P-values below 0.05 were considered statistically significant. Statistical analyses were performed using R (version 4.1.2; R Foundation for Statistical Computing, Vienna, Austria).

Results

Test in artificial knee joint

No excessive extrusion was observed visually for any of the prostheses during the test, meaning that most of the prosthesis surface maintains in contact with the tibial plateau and femur condyle.

The results listed in Table 1 show that the first-generation meniscus prosthesis, with the rigid reinforcing core, transfers virtually all the load from artificial femur to artificial tibia. This is also the case when the flexible second-generation prosthesis is tested dry at room temperature. When this prosthesis is pre-soaked and tested under more physiological conditions, the load distribution shifts towards the direct tibiofemoral contact, especially when increasing the applied axial load on the femur. As this does not occur with the first-generation meniscus prosthesis, the difference in the relative amount of load transferred through the meniscus prosthesis is large, with 99% for the first generation and 36% for the second generation respectively at 700 N.

Table 1 Load sharinga in the artificial knee setup

Test in cadaveric knee joints

In general, the size of the meniscus prosthesis matched well with the size of the native meniscus. The average anteroposterior (AP) length (mean ± standard deviation) of the native menisci was 49.1 ± 2.7 mm, where the AP dimension of the meniscus prosthesis is 48.9 mm. This results in a mean absolute dimensional difference in AP length of 3.7 ± 2.6%.

Under the highest axial load applied in this study, i.e. 1000 N, total medial meniscectomy increases the peak pressures compared to the intact native meniscus from 2.4 to 5.5 MPa. After implantation of the medial meniscus prostheses, the peak pressure is decreased to 3.7 MPa. The average peak pressure was higher for the meniscus prosthesis than for the native meniscus, but this difference of 1.3 MPa was not statistically significant.

Under an axial load of 500 N, the medial mean pressure in the meniscectomized knee (0.9 MPa) was also higher than with an intact native meniscus (0.4 MPa) and higher than with an implanted meniscus prosthesis (0.5 MPa). The mean pressures increased when increasing the load to 1000 N, to 0.6 MPa for the native meniscus, 1.6 MPa for total meniscectomy and 0.9 MPa for the meniscus prosthesis. For both axial loading conditions, no statistically significant differences were found between the native meniscus and the meniscus prosthesis.

The contact area on the medial tibial plateau was largest with the native meniscus in place, with 543 and 607mm2 under 500 and 1000 N, respectively. The contact area decreased after total medial meniscectomy, to 296 and 329mm2. A statistically significant increase compared to meniscectomy was not found for the meniscus prosthesis, with contact areas of 335 and 405mm2, respectively.

No significant differences were found in the ratio between the loads transferred through the medial native meniscus or meniscus prosthesis and through the direct tibiofemoral contact. Both the native meniscus (from 61 to 51%) and the meniscus prosthesis (from 59 to 43%) seem to transfer a relatively smaller amount of load when the total axial load is increased from 500 to 1000 N, but none of the differences was statistically significant.

No significant effect of the meniscal conditions on the mean contact pressures on the lateral tibial plateau was observed. This was the case for both 500 and 1000 N of applied axial load. Detailed results are shown in Fig. 10, Tables 2 and 3.

Fig. 10
figure 10

Mean and standard deviation of the biomechanical outcome of the test in cadaveric knee joints, split into medial peak pressure, medial mean pressure, medial contact area, load sharing between the meniscus or prosthesis and the direct tibiofemoral contact, and lateral mean pressure

Table 2 Mean and standard deviation of the biomechanical outcomes from the cadaveric study
Table 3 Differences in biomechanical outcomes between the different conditions in the cadaveric study

Discussion

This study showed that a flexible, anatomically shaped meniscus prosthesis is able to significantly reduce peak and mean pressures in a knee joint after total medial meniscectomy. This improvement is achieved by load sharing between the meniscus prosthesis and the direct tibiofemoral cartilage contact. This load sharing phenomenon did not occur with the stiff first-generation prosthesis, as demonstrated in the test in the artificial knee joint. No significant differences in pressures on the medial tibial plateau between the native meniscus and the second-generation meniscus prosthesis could be demonstrated in this study. Despite the absence of the reinforcing core in the body, the flexible prosthesis was not extruded from the knee joint and therefore sufficiently contributed to the load transfer from the femoral condyle to the tibial plateau during initial use.

The magnitude of peak pressures, mean pressures and contact areas in the native and the meniscectomized knee joint are within the range of contact pressures reported in earlier studies [2, 14, 21, 27, 31]. The main difference with these studies is that the current study was performed in a more physiological environment (i.e. wet and 37 °C), in order to try to mimic the conditions in the knee joint. The importance of testing under these conditions is demonstrated by the test in the artificial knee joint, where the relative amount of load transferred trough the meniscus prosthesis considerably decreased compared to the dry test at room temperature.

The anterior location of the high-pressure areas (Fig. 11), is consistent with findings during unicompartmental arthroplasties, where typically the tibial cartilage is found to be eroded in the anteromedial area of the tibial plateau [30]. This may be an indication that long-term exposure to high pressures may lead to cartilage degeneration and pain [25]. Although a direct causal relationship between contact pressure and pain has never been demonstrated, meniscal allografts have proven to both decrease tibial contact pressures [11, 27] and relieve pain [23]. An alternative, free-floating medial meniscus prosthesis also demonstrated to decrease tibial contact pressures [22] and reduce pain in patients with partial meniscectomy [33]. Based on the comparison with alternative treatments and the outcomes of this study, the meniscus prosthesis system presented here is expected to relieve pain in patients who have undergone (sub-)total medial meniscectomy. The positive results of the patient with the remaining first-generation prosthesis in the first-in-human clinical investigation corroborate this hypothesis [24]. The outcomes of the clinical investigation suggest that even the stiff version of the meniscus prosthesis may eventually adapt to the geometry of the knee joint through permanent deformation after two years of continued loading. The results indicate that the meniscus prosthesis has the potential to reduce pain in the affected knee compartment, provided that it has a good fit and is positioned correctly [24].

Fig. 11
figure 11

Representative example of the pressure distribution in the different conditions in a cadaveric knee joint

The load sharing ratio between the meniscus or prosthesis and the direct tibiofemoral contact (i.e. the percentage of the medial compartment load that is transferred through the meniscus) found in this study was within the expected range [5, 29]. At different loads, no differences in load sharing capabilities were found between the native meniscus and the second-generation meniscus prosthesis. This finding may indicate a similar behaviour in terms of extrusion, i.e. elastic deformation caused by the axial load on the femur condyle in combination with the wedge-shaped cross-section of the meniscus and the prosthesis. In the cadaveric study, approximately half of the load was carried by the direct tibiofemoral contact, while the results of the test in the artificial knee joint showed that hardly any load sharing occurred with the stiff first generation of the prosthesis.

This study has several limitations, as mimicking the complex interaction between tibia, femur and the deformable meniscus is highly challenging from an experimental point of view. One of the main limitations of the test in the artificial knee joint is the lack of flexibility in the metal femur and the plastic tibia. Although the deformation of the center of the UHMWE tibial plateau simulates the physiological deformation to some extent [8], the rigid material does not adapt to the contacting geometry as the natural cartilage would. The same applies to the metal femoral component. These limitations probably lead to higher pressures in the direct tibiofemoral contact, as a result of a smaller contact area. It is therefore recommended to only use an artificial knee setup, like the one introduced in this study, for direct and qualitative comparison between different test cases. For assessment of in vivo mechanical performance and comparison with other meniscal conditions (e.g. meniscectomy), cadaveric studies are preferred.

Ideally, the cadaveric test performed in this study would also have included the previous, stiffer version of the prosthesis. Unfortunately, it was not feasible to subsequently use the screw fixation and the tape fixation methods in the same knee joint, due to the different drill holes required for both techniques. Therefore, it was decided to perform the first part of this study (i.e. the comparison between the flexible and the rigid prosthesis) in an artificial knee joint, which ensures reproducible results.

The cadaveric experiments are also subject to other limitations. Due to the fixation to the tibial plateau (i.e. the presence of anchoring tapes), the pressure-sensitive sensor could not be placed as laterally on the medial plateau as desired, which hindered the measurement of pressures underneath the ends of the prosthesis horns. Therefore, the contact area in the situation with an implanted meniscus prosthesis is probably underestimated in this study. Based on the pressure maps as shown in Fig. 11, one could expect that the meniscus prosthesis does considerably increase the contact area relative to a knee joint that underwent total medial meniscectomy, but not to the level of the native meniscus.

The cadaveric test setup provides a certain amount of freedom in varus/valgus rotation, to allow self-alignment of the joint after intervention. The axis of this varus/valgus rotation is defined by the position and direction of the steel rod (Fig. 8) in the tibia. Due to the design of the test setup and the anatomy of the knee joint the rod was placed in the tibia, a couple of centimetres below the joint line. This impedes the freedom of varus/valgus rotation, which in combination with restricted mediolateral translations limits the self-alignment capabilities of the knee joint.

This may have resulted in a misjudgement of the effect of the different meniscus conditions on the load distribution between the medial and lateral compartment. This limitation could therefore be part of the explanation why no differences in lateral contact mechanics were found in this study. A similar study also found no effect of meniscectomy and implantation of a medial meniscus prosthesis on the lateral knee compartment, but varus/valgus rotation was completely fixed in this study [22].

This study only included axial loading in full extension, while the tibial contact pressures may be higher under different flexion angles [2, 14, 21]. Although the highest physiological loads during gait occur at only 15° of flexion [3, 9], full extension may prevent anteroposterior translation and varus/valgus rotation. Therefore, some differences in biomechanical outcomes may be expected in different flexion angles. In this study, flexion angles are limited by the experimental setup, which does not allow for testing under different flexion angles inside the water bath. Additionally, prolonged testing was not desirable, due to the deterioration of the cadaveric soft tissues in a warm and wet environment, and the potential effect of shear forces in flexed knee joints [31] and prolonged exposure to water [27] on the pressure-sensitive sensors.

Finally, to prevent failure of the cadaveric bones, which occurred during pilot testing, and to prevent overloading of the pressure-sensitive sensors, sub-physiological loads were applied. Higher axial loads would probably have led to higher contact pressures. The main interest, however, was focussing on the differences between the different meniscal situations, and it is not expected that higher loads and larger flexion angles would have a large impact on the relative differences found for all outcome measures of this study.

Conclusions

The anatomically shaped PCU medial meniscus prosthesis presented in this study was able to significantly improve the contact pressures on the tibial plateau of knee joints that underwent a total medial meniscectomy, by achieving load sharing between the prosthesis and the direct tibiofemoral cartilage contact. Although the biomechanical performance of the native meniscus could not be reproduced completely, the meniscus prosthesis has the potential to relieve pain in patients who have undergone previous (sub-)total medial meniscectomy, which needs to be substantiated in a clinical study.

Availability of data and materials

The dataset supporting the conclusions of this article will be made available after publication of the article.

Abbreviations

ACL:

Anterior cruciate ligament

AP:

Anteroposterior

CI:

Confidence interval

CoCr:

Cobalt chrome

EtO:

Ethylene oxide

MAT:

Meniscal allograft transplantation

NS:

Not significant

PCU:

Polycarbonate urethane

PEEK:

Polyether ether ketone

PET:

Polyethylene terephthalate

PMMA:

Polymethyl methacrylate

RT:

Room temperature

Ti-6Al-7Nb:

Titanium alloy, containing 6% aluminium and 7% niobium

UHMWPE:

Ultra-high molecular weight polyethylene

References

  1. Abrams GD, Frank RM, Gupta AK, Harris JD, McCormick FM, Cole BJ (2013) Trends in meniscus repair and meniscectomy in the United States, 2005–2011. Am J Sports Med 41:2333–2339

    Article  Google Scholar 

  2. Bedi A, Kelly NH, Baad M, Fox AJ, Brophy RH, Warren RF, Maher SA (2010) Dynamic contact mechanics of the medial meniscus as a function of radial tear, repair, and partial meniscectomy. J Bone Jt Surg Am 92:1398–1408

    Article  Google Scholar 

  3. Bergmann G, Bender A, Graichen F, Dymke J, Rohlmann A, Trepczynski A, Heller MO, Kutzner I (2014) Standardized loads acting in knee implants. PLoS One 9:e86035

    Article  Google Scholar 

  4. Chatain F, Adeleine P, Chambat P, Neyret P (2003) A comparative study of medial versus lateral arthroscopic partial meniscectomy on stable knees: 10-year minimum follow-up. Arthrosc J Arthrosc Relat Surg 19:842–849

    Article  CAS  Google Scholar 

  5. Drobnič M, Ercin E, Gamelas J, Papacostas ET, Slynarski K, Zdanowicz U, Spalding T, Verdonk P (2019) Treatment options for the symptomatic post-meniscectomy knee. Knee Surg Sports Traumatol Arthrosc Off J ESSKA 27:1817–1824

    Article  Google Scholar 

  6. Englund M, Guermazi A, Gale D, Hunter DJ, Aliabadi P, Clancy M, Felson DT (2008) Incidental meniscal findings on knee MRI in middle-aged and elderly persons. N Engl J Med 359:1108–1115

    Article  CAS  Google Scholar 

  7. Geary C, Birkinshaw C, Jones E (2008) Characterisation of Bionate polycarbonate polyurethanes for orthopaedic applications. J Mater Sci Mater Med 19:3355–3363

    Article  CAS  Google Scholar 

  8. Hosseini A, Van de Velde SK, Kozanek M, Gill TJ, Grodzinsky AJ, Rubash HE, Li G (2010) In-vivo time-dependent articular cartilage contact behavior of the tibiofemoral joint. Osteoarthritis Cartilage 18:909–916

    Article  CAS  Google Scholar 

  9. ISO 14243–1:2009 Implants for surgery - Wear of total knee-joint prostheses - Part 1: Loading and displacement parameters for wear-testing machines with load control and corresponding environmental conditions for test. International Organization for Standardization, Geneva. Retrieved from: https://www.iso.org/standard/44262.html

  10. Khoshgoftar M, Vrancken ACT, van Tienen TG, Buma P, Janssen D, Verdonschot N (2015) The sensitivity of cartilage contact pressures in the knee joint to the size and shape of an anatomically shaped meniscal implant. J Biomech 48:1427–1435

    Article  CAS  Google Scholar 

  11. Kim JG, Lee YS, Bae TS, Ha JK, Lee DH, Kim YJ, Ra HJ (2013) Tibiofemoral contact mechanics following posterior root of medial meniscus tear, repair, meniscectomy, and allograft transplantation. Knee Surg Sports Traumatol Arthrosc 21:2121–2125

    Article  Google Scholar 

  12. Lee BS, Chung JW, Kim JM, Cho WJ, Kim KA, Bin SI (2012) Morphologic Changes in Fresh-Frozen Meniscus Allografts Over 1 Year: A Prospective Magnetic Resonance Imaging Study on the Width and Thickness of Transplants. Am J Sports Med 40:1384–1391

    Article  Google Scholar 

  13. Lee DH, Lee CR, Jeon JH, Kim KA, Bin SI (2015) Graft Extrusion in Both the Coronal and Sagittal Planes Is Greater After Medial Compared With Lateral Meniscus Allograft Transplantation but Is Unrelated to Early Clinical Outcomes. Am J Sports Med 43:213–219

    Article  Google Scholar 

  14. Lee SJ, Aadalen KJ, Malaviya P, Lorenz EP, Hayden JK, Farr J, Kang RW, Cole BJ (2006) Tibiofemoral contact mechanics after serial medial meniscectomies in the human cadaveric knee. Am J Sports Med 34:1334–1344

    Article  Google Scholar 

  15. McDermott ID, Amis AA (2006) The consequences of meniscectomy. J Bone Joint Surg Br 88-B:1549–1556

    Article  Google Scholar 

  16. Millennium Research Group (2015) Orthopedic Soft Tissue Solutions US. Market Analysis, Toronto, Ontario

    Google Scholar 

  17. Millennium Research Group (2015) Orthopedic Sports Medicine EU. Market Analysis, Toronto, Ontario

    Google Scholar 

  18. Papalia R, Del Buono A, Osti L, Denaro V, Maffulli N (2011) Meniscectomy as a risk factor for knee osteoarthritis: a systematic review. Br Med Bull 99:89–106

    Article  Google Scholar 

  19. Rao AJ, Erickson BJ, Cvetanovich GL, Yanke AB, Bach BR, Cole BJ (2015) The Meniscus-Deficient Knee: Biomechanics, Evaluation, and Treatment Options. Orthop J Sports Med 3:2325967115611386

    Article  Google Scholar 

  20. Richards RE, Andersen MS, Harlaar J, van den Noort JC (2018) Relationship between knee joint contact forces and external knee joint moments in patients with medial knee osteoarthritis: effects of gait modifications. Osteoarthritis Cartilage 26:1203–1214

    Article  CAS  Google Scholar 

  21. Seitz AM, Lubomierski A, Friemert B, Ignatius A, Durselen L (2012) Effect of partial meniscectomy at the medial posterior horn on tibiofemoral contact mechanics and meniscal hoop strains in human knees. J Orthop Res 30:934–942

    Article  Google Scholar 

  22. Shemesh M, Shefy-Peleg A, Levy A, Shabshin N, Condello V, Arbel R, Gefen A (2020) Effects of a novel medial meniscus implant on the knee compartments: imaging and biomechanical aspects. Biomech Model Mechanobiol 19:2049–2059

    Article  Google Scholar 

  23. Southworth TM, Naveen NB, Tauro TM, Chahla J, Cole BJ (2020) Meniscal Allograft Transplants. Clin Sports Med 39:93–123

    Article  Google Scholar 

  24. van Tienen TG, van Minnen BS, Defoort KC, Emans PJ, Van de Groes SAW, Jutten LM, Pikaart RWE, Verdonschot N, Heesterbeek PJC (2022) Anatomical medial meniscus prosthesis: learnings from a first-in-man study [Under review by KSSTA]

    Google Scholar 

  25. Voloshin AS, Wosk J (1983) Shock absorption of meniscectomized and painful knees: a comparative in vivo study. J Biomed Eng 5:157–161

    Article  CAS  Google Scholar 

  26. Vrancken ACT, Crijns SPM, Ploegmakers MJM, O’Kane C, van Tienen TG, Janssen D, Buma P, Verdonschot N (2014) 3D geometry analysis of the medial meniscus - a statistical shape modeling approach. J Anat 225:395–402

    Article  CAS  Google Scholar 

  27. Vrancken ACT, Eggermont F, van Tienen TG, Hannink G, Buma P, Janssen D, Verdonschot N (2016) Functional biomechanical performance of a novel anatomically shaped polycarbonate urethane total meniscus replacement. Knee Surg Sports Traumatol Arthrosc 24:1485–1494

    Article  CAS  Google Scholar 

  28. Vrancken ACT, van Tienen TG, Hannink G, Janssen D, Verdonschot N, Buma P (2014) Releasing the circumferential fixation of the medial meniscus does not affect its kinematics. Knee 21:1033–1038

    Article  CAS  Google Scholar 

  29. Walker PS, Arno S, Bell C, Salvadore G, Borukhov I, Oh C (2015) Function of the medial meniscus in force transmission and stability. J Biomech 48:1383–1388

    Article  Google Scholar 

  30. White S, Ludkowski P (1991) Goodfellow J Anteromedial osteoarthritis of the knee. J Bone Joint Surg Br 73-B:582–586

    Article  Google Scholar 

  31. Willinger L, Lang JJ, Berthold D, Muench LN, Achtnich A, Forkel P, Imhoff AB, Burgkart R, von Deimling C (2020) Varus alignment aggravates tibiofemoral contact pressure rise after sequential medial meniscus resection. Knee Surg Sports Traumatol Arthrosc Off J ESSKA 28:1055–1063

    Article  Google Scholar 

  32. Winby CR, Lloyd DG, Besier TF, Kirk TB (2009) Muscle and external load contribution to knee joint contact loads during normal gait. J Biomech 42:2294–2300

    Article  CAS  Google Scholar 

  33. Zaslav KR, Farr J, Alfred R, Alley RM, Dyle M, Gomoll AH, Lattermann C, McKeon BP, Kaeding CC, Giel T, Hershman EB (2022) Treatment of post-meniscectomy knee symptoms with medial meniscus replacement results in greater pain reduction and functional improvement than non-surgical care. Knee Surg Sports Traumatol Arthrosc 30:1325–1335

    Article  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Richard van Swam for his help in conducting the experiments and Gerjon Hannink for his help with the statistical analysis.

Funding

This study was partially funded by ATRO Medical B.V.

Author information

Authors and Affiliations

Authors

Contributions

BvM participated in the design of the study, executed the mechanical tests, processed the results and drafted the manuscript. AvdV participated in the design of the study, specimen preparation and execution of the mechanical tests. SvdG participated in the design of the study and helped to interpret the data from a clinical point of view. NV provided the laboratory facilities and helped to interpret the outcomes from a biomechanical point of view. TvT participated in the design of the study, performed all surgical procedures and helped to interpret the outcomes. All authors read, reviewed and approved the final manuscript.

Corresponding author

Correspondence to Branco S. van Minnen.

Ethics declarations

Ethics approval and consent to participate

Ethical approval for the use of cadaveric material was obtained from the anatomy department of the Radboud university medical center. All donors have submitted a handwritten version of their will, signed by both the donor and a public notary.

Consent for publication

Not applicable.

Competing interests

> BvM and AvdV are employed at ATRO Medical B.V.

> TvT receives a management fee from ATRO Medical B.V.

> BvM, AvdV and TvT own stocks in ATRO Medical B.V.

> BvM, AvdV and TvT are inventors on applicable patents, but have no ownership.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

van Minnen, B.S., van der Veen, A.J., van de Groes, S.A.W. et al. An anatomically shaped medial meniscus prosthesis is able to partially restore the contact mechanics of the meniscectomized knee joint. J EXP ORTOP 9, 91 (2022). https://doi.org/10.1186/s40634-022-00531-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40634-022-00531-6