Skip to main content

Table 2 Mean and standard deviation of the biomechanical outcomes from the cadaveric study

From: An anatomically shaped medial meniscus prosthesis is able to partially restore the contact mechanics of the meniscectomized knee joint

Force

Condition

Medial peak pressure (MPa)

Medial mean pressure (MPa)

Medial contact area (mm2)

Load sharinga

(%)

Lateral mean pressure (MPa)

500 N

Native

1.4 ± 0.6

0.4 ± 0.1

543 ± 102

61 ± 28

0.4 ± 0.1

Meniscectomy

3.4 ± 1.3

0.9 ± 0.3

296 ± 92

-

0.4 ± 0.1

Prosthesis

2.1 ± 0.7

0.5 ± 0.2

335 ± 51

59 ± 24

0.4 ± 0.0

1000 N

Native

2.4 ± 1.2

0.6 ± 0.2

607 ± 114

51 ± 27

0.7 ± 0.1

Meniscectomy

5.5 ± 1.8

1.6 ± 0.5

329 ± 123

-

0.6 ± 0.1

Prosthesis

3.7 ± 1.6

0.9 ± 0.3

405 ± 91

43 ± 14

0.7 ± 0.1

  1. a Percentage of medial compartment load transferred through the meniscus or the prosthesis