Skip to main content
  • Original Paper
  • Open access
  • Published:

Similar outcomes including maximum knee flexion between mobile bearing condylar-stabilised and fixed bearing posterior-stabilised prosthesis: a case control study

Abstract

Purpose

Prosthesis design influences stability in total knee arthroplasty and may affect maximum knee flexion. Posterior-stabilised (PS) and condylar-stabilised (CS) designed prosthesis do not require a posterior-cruciate ligament to provide stability. The aim of the current study was to compare the range of motion (ROM) and clinical outcomes of patients undergoing cemented total knee arthroplasty (TKA) using either a PS or CS design prosthesis.

Methods

A total of 167 consecutive primary TKAs with a CS bearing (mobile deep-dish polyethylene) were retrospectively identified and compared to 332 primary TKA with a PS constraint, with similar design components from the same manufacturer. Passive ROM was assessed at last follow-up with use of a handheld goniometer. Clinical scores were assessed using Patient-Reported Outcome Measures (PROMs); International Knee Society (IKS) knee and function scores and satisfaction score. Radiographic assessment was performed pre and post operatively consisting of mechanical femorotibial angle (mFTA), femoral and tibial mechanical angles measured medially (FMA and TMA, respectively) on long leg radiographs, tibial slope and patella height as measured by the Blackburne-Peel index (BPI).

Results

Both groups had a mean follow-up of 3 years (range 2–3.7 years). Mean post-operative maximum knee flexion was 117° ± 4.9° in the PS group and 119° ± 5.2° in the CS group (p = 0.29). Postoperative IKS scores were significantly improved in both groups compared to preoperative scores (p < 0.01). The mean IKS score in the PS group was 170.9 ± 24.1 compared to 170.3 ± 22.5 in the CS group (p = 0.3). Both groups had similar radiographic outcomes as determined by coronal and sagittal alignment, tibial slope and posterior condylar offset ratio measurements. When considering the size of tibial slope change and posterior-condylar offset ratio, there was no differences between groups (p = 0.4 and 0.59 respectively). The PS group had more interventions for post-operative stiffness (arthrolysis or manipulation under anaesthesia) 8 (2.7%) compared to 1 (0.6%) in the CS group (p = 0.17).

Conclusion

Condylar-stabilised TKA have similar patient outcomes and ROM at a mean follow-up of 3 years compared to PS TKA. Highly congruent inserts could be used without compromising results in TKA at short term.

Level of evidence

Level IV, retrospective case control study.

Introduction

Achieving a functional range of motion (ROM) and a stable joint in the coronal and sagittal planes are critical goals of primary total knee arthroplasty (TKA). Prosthesis design influences stability in TKA [1] and may effect maximum knee flexion.

Posterior-stabilised (PS) designed implants are characterised by a box post-cam mechanism that substitutes the posterior cruciate ligament (PCL). This creates a larger flexion space, facilitating balancing and clearance of posterior osteophytes whilst maintaining stability through the post-cam mechanism [2, 3]. Increased femoral rollback in PS constraint is proposed to lead to greater range of flexion and a reduced prevalence of posterior tibial subluxation [3]. However, concerns exist regarding high stress imparted onto the cam-mechanism in PS designed implants, potentially leading to increased polyethylene wear, tibial loosening or fracture of the post [4]. Furthermore, bone stock is sacrificed due to the need for intercondylar bone resection [5, 6].

Implant design with low constraint such as a condylar constrained (CS) (ultra-congruent, deep dished, lipped liner) bearing insert theoretically offer stability through a highly conforming articulation and raised anterior and posterior lips [7]. Conflicting results have been published using CS prosthesis in TKA, with some studies previously reporting instability with the use of a CS implant [7,8,9], whilst others have reported good medium-term survival without increased risk of revision for instability [10]. Further, concerns remain that the increased conformity of deep-dished liners may come at the expense of flexion range and increased sheer forces across the polyethylene may be experienced due to the increased sagittal laxity observed compared to PS designed prosthesis [5, 6, 8, 11].

The aim of this study was to compare the ROM and clinical outcomes of patients undergoing cemented TKA using a either a PS or a mobile CS TKA from the same manufacturer. The authors’ hypothesis was that when tibial slope and PCOR were controlled for, no difference in ROM or patient outcomes would be observed between groups.

Methods

Patients

A monocentric retrospective analysis of consecutive patients who underwent primary PS TKA from the same manufacturer, between January 2018 to November 2019 was performed. All TKA were performed by a senior surgeon with either a CS or PS knee prosthesis and were included if they had minimum follow-up of 2 years.

Demographics

In the CS group, 1 patient passed away, 2 were lost to follow-up and 4 were excluded leaving 160 patients available for analysis. In the control group, 3 patients passed away, 4 were lost to follow-up and 9 were excluded leaving 316 PS knees with a post-cam mechanism for analysis. A complete flowchart summarising patient selection is illustrated in Fig. 1. Both groups were similar for all characteristics as reported in Table 1. Specifically, both groups were similar in pre-operative flexion range (116° control versus 116° study group).

Fig. 1
figure 1

Study flowchart of posterior-stabilised versus condylar-stabilised cohort selection

Table 1 Comparison of pre-operative patient demographics between posterior-stabilised and condylar-stabilised groups

Surgery

All surgeries were performed without tourniquet. Patients in the control group received a fixed bearing primary PS TKA with a post-cam mechanism (ANATOMIC®, Amplitude, Valence, France). Patients in the CS group received a deep-dish, mobile bearing TKA (SCORE II®, Amplitude, Valence, France). Both the mobile-bearing and fixed-bearing prostheses were manufactured by the same manufacturer. Except for the design of the articulating surface, the two prostheses were identical. Both femoral components have the same curvature in the sagittal plane.

A medial sub-vastus approach was used if pre-operative alignment was in varus and a lateral parapatellar approach for cases with valgus alignment. Surgery was performed using manual instrumentation and a measured resection technique was utilised in all cases. Sizing for the femur was done by posterior referencing. All femoral components were referenced from the posterior femoral condyle. Eccentric external femoral rotation of 3° relative to the posterior condylar axis (PCA) was performed for valgus aligned knees. All other knees had femoral components implanted in neutral rotation relative to the PCA. Balancing of gaps in extension and flexion was assessed manually after osteophyte clearance and removal of the PCL with spacers, and soft tissue releases performed as required. All components in both groups were cemented and the patella was electively resurfaced.

Clinical assessment

All patients had standardised postoperative follow-up at 2,12 months and annually after surgery. The International Knee Society score was collected [12] . Patient satisfaction was assessed categorised using a global clinical outcome: very satisfied, satisfied, disappointed or dissatisfied. Range of motion was recorded using a hand-held goniometer. The complication rate was evaluated at the last follow-up, including all reintervention procedures (component exchange, debridement and irrigation, mobilisation under anaesthesia and arthrolysis).

Radiographic assessment

All patients had a pre-operative and postoperative radiographic assessment at 2, 12 months and annually which included: anteroposterior view, lateral view, weight bearing view, patellar axial view and standing full length-radiographs. Axial views were performed using the Merchant method [13]. Patellar height was calculated with Blackburne–Peel index (BPI) [14]. PCOR was measured according to technique described by Johal et al. [15]. Mechanical femorotibial angle (mFTA), femoral and tibial mechanical angles measured medially (FMA and TMA, respectively) and tibial slope were measured [16] using the PACS system (Centricity Enterprise, GE Healthcare, Barrington, IL, USA). All radiographs were measured by two independent orthopaedic surgeons.

Statistical analysis

Statistical analysis was performed using SPSS (IBM, version 18.0). Baseline characteristics were described using mean and standard deviation for continuous measures. For non-parametric data, means were compared using Mann-Whitney test for continuous outcomes and Chi-square and Fisher Exact test for categorical outcomes. Continuous means were compared using independent T-test. Significance was set at p < 0.05 for all tests. A post hoc analysis was performed with a mean difference of maximum knee flexion of 2 points between groups, with a common standard deviation of 5 points, a power of 0.8 and an alpha risk of 0.05. A minimum sample size of 99 cases in each group was necessary for this study.

Ethics approval

This study had approval from the Advisory Committee on Research Information Processing in the Field of Health (CCTIRS), and IRB approval study number is 135–5265.

Results

Patient characteristics

A total of 476 knees were included in the final analysis (160 CS and 316 PS). Mean follow-up was 36.6 months for the PS group and 37.1 months for the CS group (p = 0.07). The patella was resurfaced 84.6% of the time in the control group compared to 45.8% of the time in the study group (p < 0.001).

Clinical outcomes

Mean post-operative maximum knee flexion was 117° ± 5° in the PS group and 119° ± 5° in the CS group (p = 0.29). In both groups, postoperative IKS scores were significantly improved compared to preoperative scores (p < 0.01). No significant differences were observed between both groups for clinical scores or patient satisfaction. Specifically, all clinical outcomes are reported in Table 2.

Table 2 Comparison of post-operative clinical outcomes between posterior-stabilised and condylar-stabilised groups

Radiographic outcomes

Radiographic outcomes at last follow-up are summarised in Table 3. Both groups had similar coronal and sagittal alignment as determined by mFTA and tibial slope measurements. When considering the size of tibial slope change and PCOR, there were no differences between groups (p = 0.4 and 0.59 respectively).

Table 3 Comparison of post-operative radiographic outcomes between posterior-stabilised and condylar-stabilised groups

Complications and revisions

The PS group had more interventions for post-operative stiffness (arthrolysis or manipulation under anaesthesia) 8 (2.5%) compared to 1 (0.6%) in the CS group, but this difference was not significant (p = 0.17). When considering revision surgery for any revision, the PS group had 14 (4.4%) revisions compared to 3 (1.9%) (p = 0.19). Specifically, one patient in the PS control group underwent revision for instability during the follow-up period.

Discussion

The most important finding of this study was the use of a mobile bearing CS designed polyethylene liner did not compromise ROM or patient outcomes compared to a PS fixed-bearing designed implant when used for primary TKA. Similar outcomes for flexion range were achieved without leading to a difference in re-intervention for stiffness, revision rate for instability or indeed any reason at short-term follow-up.

A proposed advantage of PS constrained TKA is improved ROM and facilitation of gap balancing whilst providing AP stability [17]. Conversely, CS TKA features a deep dished polyethylene insert which has been proposed to lead to impingement of the femur on the posterior lip causing a subsequent reduction in flexion [18]. Several studies have examined ROM in CS designed TKA reporting variable results, but have not controlled for tibial slope or PCOR [18,19,20,21,22,23,24,25,26,27]. Both of these factors have been reported to influence flexion range at least in PS designed TKA [28]. The current study controlled for both of these factors and demonstrated that ROM is not negatively affected by the use of a mobile bearing CS type prosthesis compared to a PS design. Furthermore, use of a CS implant did not result in an increased re-intervention for stiffness following primary TKA compared to a PS designed implant.

Concerns exist regarding the stability of CS designed implants. Previously it has been observed that deep-dish designed inserts have greater AP translation than PS designed implants [18, 19]. This could potentially lead to instability, or increased sheer forces on the tibia. Interestingly, mobile bearing CS designed TKA in one study were observed to provide more mid-flexion AP stability than a PS fixed bearing designed TKA [29]. In the current study one revision was performed for instability and this was in the PS group.

The results of the current study are supported by recent literature which have not demonstrated poorer outcomes with deep-dished polyethylene liners in primary TKA (Table 4). Three recent RCT’s have reported no difference in outcomes between PS and CS designed TKA. Specifically, Akti et al. found no significant differences in KSS or isokinetic performance scores between prosthesis designs [7]. Furthermore, two of these RCT’s have demonstrated no difference in functional outcomes or ROM intra-operatively, at 1 [24] and 5-year follow-up [35]. In a much larger study consisting of over 3000 TKA, Yacovelli et al. recently compared functional outcomes between patients undergoing primary TKA using fixed bearing CS versus PS designed implant. The authors found similar functional and survival outcomes [36], however ROM was not reported. Stirling et al. compared 54 CS TKA to 364 CR TKAs and found similar functional results, including no difference in flexion range [35]. Whilst these studies used a fixed bearing CS design and often included smaller numbers than the present study, the results are similar (Table 4). Whilst differences in design features are seemingly subtle, functional results comparing a mobile bearing deep-dished liner to a similar PS designed femoral implant until now have not been well described previously. The present study demonstrates comparable clinical outcome to a PS prosthesis.

Table 4 Comparative studies reporting range of motion and functional outcome for condylar-stabilised (CS) prosthesis in primary total knee arthroplasty

The indications for choosing between a CS or PS design prosthesis appear to be based on proposed advantages or disadvantages rather than outcomes. In the present study, mobile- and fixed-bearing designs were compared showing no differences in terms of radiological and clinical results, including maximum knee flexion. This is in line with several meta-analyses and randomized controlled trial assessing clinical and radiological results between fixed- and mobile- bearing for PS TKA [37,38,39,40,41]. PS designed prosthesis have previously demonstrated less AP translation than CS prosthesis [8], however sacrifice more bone stock, generate more polyethylene wear particles [42, 43] and may have increased risk of aseptic loosening [4, 44]. The present study did not find significant differences between both groups considering revision rate for aseptic loosening. This is consistent with the literature comparing fixed- and mobile-bearing for PS implants [45,46,47]. Despite these differences, neither design has demonstrated inferior outcomes. Furthermore, concerns about the increased AP translation seen in kinematic studies possibly leading to an increased revision rate of CS TKA compared to PS is not supported by recent registry data [48].

Results comparing mobile bearing ultra-congruent design TKA to a fixed bearing PS design implant are limited. One study has compared a mobile bearing CS design to a fixed bearing PS prosthesis for primary TKA and found similar ROM between groups to the present study [29]. Two studies have previously reported outcomes comparing two mobile bearing designs (CS versus PS) [21], with one reporting reduced range of motion using the CS compared to mobile-bearing PS TKAs (126° vs. 131°) [34]. The findings of the current study did not demonstrate clinically meaningful differences in ROM and indeed outcomes post TKA may be related to additional factors such as balancing [49].

This study has several limitations. Firstly, it was limited by the retrospective nature of the study. Secondly, the follow-up period for the present study is comparatively short and long-term data is required comparing outcomes between deep-dished liner to PS or CR TKA polyethylene inserts. Nonetheless, the primary aim of the present study was to compare the range of motion of these two designs, and the minimum 24-month follow-up is sufficient to address this question. Finally, it is important to note that this study only represents results for this specific posterior-stabilised and condylar-stabilised prostheses.

Conclusion

Condylar-stabilised TKA have similar patient outcomes and ROM at a mean follow-up of 3 years compared to PS TKA. Highly congruent inserts could be used without compromising results in TKA at short term.

References

  1. Migliorini F, Eschweiler J, Tingart M, Rath B (2019) Posterior-stabilized versus cruciate-retained implants for total knee arthroplasty: a meta-analysis of clinical trials. Eur J Orthop Surg Traumatol 29:937–946

    Article  PubMed  Google Scholar 

  2. Bercik MJ, Joshi A, Parvizi J (2013) Posterior cruciate-retaining versus posterior-stabilized total knee arthroplasty: a meta-analysis. J Arthroplast 28:439–444

    Article  Google Scholar 

  3. Berend KR, Lombardi AV Jr, Adams JB (2013) Which total knee replacement implant should I pick? Correcting the pathology: the role of knee bearing designs. Bone Joint J 95-B:129–132

    Article  CAS  PubMed  Google Scholar 

  4. Sappey-Marinier E, Shatrov J, Batailler C, Schmidt A, Servien E, Marchetti E et al (2021) Restricted kinematic alignment may be associated with increased risk of aseptic loosening for posterior-stabilized TKA: a case-control study. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-021-06714-5

  5. Mazzucchelli L, Deledda D, Rosso F, Ratto N, Bruzzone M, Bonasia DE et al (2016) Cruciate retaining and cruciate substituting ultra-congruent insert. Ann Transl Med 4:2

    PubMed  PubMed Central  Google Scholar 

  6. Nicholls RL, Schirm AC, Jeffcote BO, Kuster MS (2007) Tibiofemoral force following total knee arthroplasty: comparison of four prosthesis designs in vitro. J Orthop Res 25:1506–1512

    Article  PubMed  Google Scholar 

  7. Akti S, Karakus D, Sezgin EA, Cankaya D (2021) No differences in clinical outcomes or isokinetic performance between cruciate-substituting ultra-congruent and posterior stabilized total knee arthroplasties: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 29:3443–3449

    Article  PubMed  Google Scholar 

  8. Bae JH, Yoon JR, Sung JH, Shin YS (2018) Posterior-stabilized inserts are preferable to cruciate-substituting ultracongruent inserts due to more favourable kinematics and stability. Knee Surg Sports Traumatol Arthrosc 26:3300–3310

    Article  PubMed  Google Scholar 

  9. Sur YJ, Koh IJ, Park SW, Kim HJ, In Y (2015) Condylar-stabilizing tibial inserts do not restore anteroposterior stability after total knee arthroplasty. J Arthroplast 30:587–591

    Article  Google Scholar 

  10. Czekaj J, Fary C, Gaillard T, Lustig S (2017) Does low-constraint mobile bearing knee prosthesis give satisfactory results for severe coronal deformities? A five to twelve year follow up study. Int Orthop 41:1369–1377

    Article  PubMed  Google Scholar 

  11. Raja BS, Gowda AKS, Ansari S, Choudhury AK, Kalia RB (2021) Comparison of functional outcomes, femoral rollback and sagittal stability of anterior-stabilized versus posterior-stabilized Total knee Arthroplasty: a systematic review and Meta-analysis of randomized controlled trials. Indian J Orthop 55:1076–1086

    Article  Google Scholar 

  12. Insall JN, Dorr LD, Scott RD, Scott WN (1989) Rationale of the knee society clinical rating system. Clin Orthop Relat Res 248:13–14

    Article  Google Scholar 

  13. Merchant AC, Mercer RL, Jacobsen RH, Cool CR (1974) Roentgenographic analysis of patellofemoral congruence. J Bone Joint Surg Am 56:1391–1396

    Article  CAS  PubMed  Google Scholar 

  14. Blackburne JS, Peel TE (1977) A new method of measuring patellar height. J Bone Joint Surg Br 59:241–242

    Article  CAS  Google Scholar 

  15. Johal P, Hassaballa MA, Eldridge JD, Porteous AJ (2012) The posterior condylar offset ratio. Knee 19:843–845

    Article  PubMed  Google Scholar 

  16. Sarmah SS, Patel S, Hossain FS, Haddad FS (2012) The radiological assessment of total and unicompartmental knee replacements. J Bone Joint Surg Br 94:1321–1329

    Article  CAS  PubMed  Google Scholar 

  17. Harato K, Bourne RB, Victor J, Snyder M, Hart J, Ries MD (2008) Midterm comparison of posterior cruciate-retaining versus -substituting total knee arthroplasty using the genesis II prosthesis. A multicenter prospective randomized clinical trial. Knee 15:217–221

    Article  PubMed  Google Scholar 

  18. Han HS, Kang SB (2020) Anterior-stabilized TKA is inferior to posterior-stabilized TKA in terms of postoperative posterior stability and knee flexion in osteoarthritic knees: a prospective randomized controlled trial with bilateral TKA. Knee Surg Sports Traumatol Arthrosc 28:3217–3225

    Article  PubMed  Google Scholar 

  19. Fritzsche H, Beyer F, Postler A, Lutzner J (2018) Different intraoperative kinematics, stability, and range of motion between cruciate-substituting ultracongruent and posterior-stabilized total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 26:1465–1470

    Article  PubMed  Google Scholar 

  20. Jang SW, Kim MS, Koh IJ, Sohn S, Kim C, In Y (2019) Comparison of anterior-stabilized and posterior-stabilized Total knee Arthroplasty in the same patients: a prospective randomized study. J Arthroplast 34:1682–1689

    Article  Google Scholar 

  21. Kim TW, Lee SM, Seong SC, Lee S, Jang J, Lee MC (2016) Different intraoperative kinematics with comparable clinical outcomes of ultracongruent and posterior stabilized mobile-bearing total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 24:3036–3043

    Article  PubMed  Google Scholar 

  22. Laskin RS, Maruyama Y, Villaneuva M, Bourne R (2000) Deep-dish congruent tibial component use in total knee arthroplasty: a randomized prospective study. Clin Orthop Relat Res 380:36–44

    Article  Google Scholar 

  23. Lee SS, Yeom J, Lee DH, Moon YW (2020) Similar outcomes between ultracongruent and posterior-stabilized insert in total knee arthroplasty: a propensity score-matched analysis. J Orthop Surg (Hong Kong) 28:2309499019893515

    Article  Google Scholar 

  24. Lutzner J, Beyer F, Dexel J, Fritzsche H, Lutzner C, Kirschner S (2017) No difference in range of motion between ultracongruent and posterior stabilized design in total knee arthroplasty: a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc 25:3515–3521

    Article  PubMed  Google Scholar 

  25. Lutzner J, Beyer F, Lutzner C, Riedel R, Tille E (2021) Ultracongruent insert design is a safe alternative to posterior cruciate-substituting total knee arthroplasty: 5-year results of a randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-021-06545-4

  26. Parsley BS, Conditt MA, Bertolusso R, Noble PC (2006) Posterior cruciate ligament substitution is not essential for excellent postoperative outcomes in total knee arthroplasty. J Arthroplast 21:127–131

    Article  Google Scholar 

  27. Uvehammer J, Karrholm J, Regner L, Carlsson L, Herberts P (2001) Concave versus posterior-stabilized tibial joint surface in total knee arthroplasty: randomized evaluation of 47 knees. J Arthroplast 16:25–32

    Article  CAS  Google Scholar 

  28. Kim JH (2013) Effect of posterior femoral condylar offset and posterior tibial slope on maximal flexion angle of the knee in posterior cruciate ligament sacrificing total knee arthroplasty. Knee Surg Relat Res 25:54–59

    Article  PubMed  PubMed Central  Google Scholar 

  29. Minoda Y, Ikebuchi M, Mizokawa S, Ohta Y, Nakamura H (2016) Mobile-bearing TKA improved the anteroposterior joint stability in mid-flexion range comparing to fixed-bearing TKA. Arch Orthop Trauma Surg 136:1601–1606

    Article  PubMed  Google Scholar 

  30. Hinarejos P, Leal-Blanquet J, Fraile-Suari A, Sanchez-Soler J, Torres-Claramunt R, Monllau JC (2021) Increased posterior translation but similar clinical outcomes using ultracongruent instead of posterior stabilized total knee arthroplasties in a prospective randomized trial. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-021-06710-9

  31. Kim MS, Koh IJ, Kim CK, Choi KY, Jeon JH, In Y (2021) Comparison of joint perception between posterior-stabilized and Ultracongruent Total knee Arthroplasty in the same patient. J Bone Joint Surg Am 103:44–52

    Article  PubMed  Google Scholar 

  32. Stirling P, Clement ND, MacDonald D, Patton JT, Burnett R, Macpherson GJ (2019) Early functional outcomes after condylar-stabilizing (deep-dish) versus standard bearing surface for cruciate-retaining total knee arthroplasty. Knee Surg Relat Res 31:3

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  33. Song EK, Lim HA, Joo SD, Kim SK, Lee KB, Seon JK (2017) Total knee arthroplasty using ultra-congruent inserts can provide similar stability and function compared with cruciate-retaining total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 25:3530–3535

    Article  PubMed  Google Scholar 

  34. Machhindra MV, Kang JY, Kang YG, Chowdhry M, Kim TK (2015) Functional outcomes of a new Mobile-bearing ultra-congruent TKA system: comparison with the posterior stabilized system. J Arthroplast 30:2137–2142

    Article  Google Scholar 

  35. Scott DF (2018) Prospective randomized comparison of posterior-stabilized versus condylar-stabilized Total knee Arthroplasty: final report of a five-year study. J Arthroplast 33:1384–1388

    Article  Google Scholar 

  36. Yacovelli S, Grau LC, Hozack WJ, Courtney PM (2021) Functional outcomes are comparable between posterior stabilized and cruciate-substituting Total knee Arthroplasty designs at short-term follow-up. J Arthroplast 36:986–990

    Article  Google Scholar 

  37. Abdel MP, Tibbo ME, Stuart MJ, Trousdale RT, Hanssen AD, Pagnano MW (2018) A randomized controlled trial of fixed- versus mobile-bearing total knee arthroplasty: a follow-up at a mean of ten years. Bone Joint J 100-B:925–929

    Article  CAS  PubMed  Google Scholar 

  38. Bo ZD, Liao L, Zhao JM, Wei QJ, Ding XF, Yang B (2014) Mobile bearing or fixed bearing? A meta-analysis of outcomes comparing mobile bearing and fixed bearing bilateral total knee replacements. Knee 21:374–381

    Article  PubMed  Google Scholar 

  39. Hantouly AT, Ahmed AF, Alzobi O, Toubasi A, Salameh M, Elmhiregh A et al (2021) Mobile-bearing versus fixed-bearing total knee arthroplasty: a meta-analysis of randomized controlled trials. Eur J Orthop Surg Traumatol. https://doi.org/10.1007/s00590-021-02999-x

  40. Li YL, Wu Q, Ning GZ, Feng SQ, Wu QL, Li Y et al (2014) No difference in clinical outcome between fixed- and mobile-bearing TKA: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 22:565–575

    Article  PubMed  Google Scholar 

  41. Sappey-Marinier E, Swan J, Maucort-Boulch D, Batailler C, Malatray M, Neyret P et al (2020) No significant clinical and radiological differences between fixed versus mobile bearing total knee replacement using the same semi-constrained implant type: a randomized controlled trial with mean 10 years follow-up. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-020-06346-1

  42. Furman BD, Lipman J, Kligman M, Wright TM, Haas SB (2008) Tibial post wear in posterior-stabilized knee replacements is design-dependent. Clin Orthop Relat Res 466:2650–2655

    Article  PubMed  PubMed Central  Google Scholar 

  43. Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB (2001) Tibial post wear in posterior stabilized total knee arthroplasty. An unrecognized source of polyethylene debris. J Bone Joint Surg Am 83:390–397

    Article  CAS  PubMed  Google Scholar 

  44. Jung WH, Jeong JH, Ha YC, Lee YK, Koo KH (2012) High early failure rate of the Columbus posterior stabilized high-flexion knee prosthesis. Clin Orthop Relat Res 470:1472–1481

    Article  PubMed  Google Scholar 

  45. Kim YH, Choi Y, Kim JS (2010) Osteolysis in well-functioning fixed- and mobile-bearing TKAs in younger patients. Clin Orthop Relat Res 468:3084–3093

    Article  PubMed  PubMed Central  Google Scholar 

  46. Kim YH, Park JW, Kim JS (2018) Comparison of high-flexion fixed-bearing and high-flexion Mobile-bearing Total knee Arthroplasties-a prospective randomized study. J Arthroplast 33:130–135

    Article  Google Scholar 

  47. Kim YH, Park JW, Kim JS (2019) The long-term results of simultaneous high-flexion Mobile-bearing and fixed-bearing Total knee Arthroplasties performed in the same patients. J Arthroplast 34:501–507

    Article  Google Scholar 

  48. Dalton P, Holder C, Rainbird S, Lewis PL (2021) Survivorship comparisons of Ultracongruent, cruciate-retaining and posterior-stabilized Tibial inserts using a single knee system design: results from the Australian orthopedic association National Joint Replacement Registry. J Arthroplast. https://doi.org/10.1016/j.arth.2021.11.001

  49. Wakelin EA, Shalhoub S, Lawrence JM, Keggi JM, DeClaire JH, Randall AL et al (2021) Improved total knee arthroplasty pain outcome when joint gap targets are achieved throughout flexion. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-021-06482-2

Download references

Funding

No funding was received for this study.

Author information

Authors and Affiliations

Authors

Contributions

The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Elliot Sappey-Marinier.

Ethics declarations

Competing interests

Prof. Sébastien Lustig has performed consultancy work for Medacta, Heraeus, Corin, Amplitude, Groupe Lépine, Depuy Synthes, Smith & Nephew, Stryker. Prof. Sébastien Lustig receives institutional research support from Corin and Amplitude. Prof. Sébastien Lustig is a board member of KSSTA, Maitrise Orthopédique and JBJS Am. The other authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Shatrov, J., Sappey-Marinier, E., Kafelov, M. et al. Similar outcomes including maximum knee flexion between mobile bearing condylar-stabilised and fixed bearing posterior-stabilised prosthesis: a case control study. J EXP ORTOP 9, 17 (2022). https://doi.org/10.1186/s40634-022-00456-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40634-022-00456-0

Keywords