Finite element analysis of medial closing and lateral opening wedge osteotomies of the distal femur in relation to hinge fractures
Journal of Experimental Orthopaedics volume 10, Article number: 33 (2023)
Abstract
Purpose
Intraoperative hinge fractures in distal femur osteotomies represent a risk factor for loss of alignment and non-union. Using finite element analysis, the goal of this study was to investigate the influence of different hinge widths and osteotomy corrections on hinge fractures in medial closed-wedge and lateral open-wedge distal femur osteotomies.
Methods
The hinge was located at the proximal margin of adductor tubercle for biplanar lateral open-wedge and at the upper border of the lateral femoral condyle for biplanar medial closed-wedge distal femur osteotomies, corresponding to optimal hinge positions described in literature. Different hinge widths (5, 7.5, 10Â mm) were created and the osteotomy correction was opened/closed by 5, 7.5 and 10Â mm. Tensile and compressive strain of the hinge was determined in a finite element analysis and compared to the ultimate strain of cortical bone to assess the hinge fracture risk.
Results
Doubling the correction from 5 to 10 mm increased mean tensile and compressive strain by 50% for lateral open-wedge and 48% for medial closed-wedge osteotomies. A hinge width of 10 mm versus 5 mm showed increased strain in the hinge region of 61% for lateral open-wedge and 32% for medial closed-wedge osteotomies. Medial closed-wedge recorded a higher fracture risk compared to lateral open-wedge osteotomies due to a larger hinge cross-section area (60–67%) for all tested configurations. In case of a 5 mm hinge, medial closed-wedge recorded 71% higher strain in the hinge region compared to lateral open-wedge osteotomies.
Conclusion
Due to morphological features of the medial femoral condyle, finite element analysis suggests that lateral-open wedge osteotomies are the preferable option if larger corrections are intended, as a thicker hinge can remain without an increased hinge fracture risk.
Introduction
Genu valgum deformity is less common than varus malalignment, but no less relevant concerning unicompartmental joint wear [5, 8, 10, 38]. Moreover, valgus malalignment is associated with patellofemoral maltracking [9, 13, 14, 16, 32]. A hypoplastic lateral condyle of the femur is the most common cause for idiopathic valgus malalignment. The distal femoral osteotomy (DFO) is the preferable surgical approach therefore [45]. Hereby, a parallel joint line can be achieved by either a medial closed wedge (MCW) or lateral open wedge (LOW) technique [45, 46]. Independent of the technique used, DFO to treat lateral compartment disease was shown to improve clinical scores with good mid and long-term survivorship [5, 8, 24]. The most commonly reported drawbacks include a high rate of complications (9%) and a considerable number of delayed (4%) or non-unions (3%), in a systematic review with 372 DFOs, respectively [45]. The main reason cited for these complications is a fracture of the hinge, which leads to reduced axial and torsional stability, and therefore increased movement and stress across the osteotomy gap and at the bone-implant construct [2, 23, 35, 44, 45]. According to the literature, cortical hinge fractures occur in 43—57% of cases, slightly more frequent in open-wedge osteotomies [15, 18, 37, 44].
Previous research identified larger osteotomy gaps and a hinge location close to the opposite cortex as risk factors associated with unstable cortical hinge fractures [26, 27, 37, 44]. Moreover, based on bone density and soft tissue coverage, optimal hinge positions have been suggested [23, 44]. The pathomechanism of hinge fractures, however, is not yet fully understood. So far, biomechanical studies focused on modified joint loading forces [1, 47] or osteosynthesis material [36, 41, 43]. To this date, no biomechanical study exists, however, that systematically assessed risk factors of hinge fractures in varus creating DFOs.
Using Finite Element (FE) analysis, the goal of this study was to investigate previously-described parameters that are associated with hinge fractures. The hypothesis was that 1) a larger osteotomy gap and a larger hinge width increase the risk of intraoperative hinge fractures in DFO, as does 2) a higher cortical bone density of the hinge. Moreover, it was hypothesized that 3) there is a higher risk of hinge fractures in MCW compared to LOW.
Methods
Development of DFO models
Based on 61 femora from routine postmortem CT scans, a previously published statistical shape model of the human femora was used [12]. To create this model, a non-rigid registration algorithm [29] was used and 20 principal components were defined to represent 99% of the shape variance [12]. Using Autodesk Inventor (Autodesk Inc., San Rafael, USA), biplanar DFO were created with a horizontal osteotomy angle of 25° for MCW and 18° for LOW, respectively (Fig. 1), with a vertical osteotomy angle of 70° [35]. The coronal plane was defined by a plane intersecting the medial and lateral condyle and the greater trochanter and was used as a reference for the horizontal osteotomy angle. The hinge axis was orientated orthogonal to the coronal plane, corresponding to the sagittal plane. Hinge location was defined according to most recent literature at the proximal margin of the adductor tubercle for LOW [44], and at the upper border of the lateral femoral condyle for MCW [23], respectively. After performed DFO, the femur was sectioned 200 mm proximal to the joint line to reduce computational effort (see Fig. 1).
Osteotomy configurations and loading conditions for MCW and LOW, fixed support (B) at the proximal end of the segmented femur and remote displacement (A). Higher meshed hinge area of cortical bone in scope from a posterior view. Mean compressive and tensile principal strain was read out for this region. MCW: Medial closed wedge; LOW: Lateral open wedge
Simulated factors of interest
To investigate the influence of correction quantity on hinge fractures, wedge sizes of 5 mm, 7.5 mm, and 10 mm were simulated (Fig. 1). In case of a closing-wedge osteotomy, the gap was defined as 1.5 mm respecting the saw blade thickness. Next, hinge widths of 5 mm, 7.5 mm, and 10 mm were imitated, reflecting the range of hinge width in current DFO techniques [44]. From the outer surface of the statistical femora shape model a volume element was created with a constant thickness for the entire model to represent cortical bone, the remaining inner volume from the statistical shape model was considered cancellous bone. Two variables for cortical thickness were investigated: 3 mm (physiological) and 1.5 mm (pathological/osteoporotic) [6, 19].
Finite element analysis and material properties
For FE analysis, Ansys Workbench (Version R1, Ansys Inc., Canonsburg, USA) was used to create a mesh using tetragonal elements with an approximate element size of 1 mm and 0.5 mm in the hinge area. A convergence analysis was performed and an aspect ratio of < 4 was recorded for 95% of its elements. The analysis was performed without the use of a bone-void filler due to the following reasons [17, 22]: 1) maximal size of the osteotomy gap was set to 10 mm, and 2) to simulate the worst-case scenario for hinge loading. Cortical bone was considered transversely isotropic (Elastic Modulus (E): Ex = Ey = 11.5 GPa, Ez = 17 GPa; Poisson's ratio (v): vxy = 0.51, vxz = vyz = 0.31 GPa), whereas cancellous bone was modeled as a linear isotropic material property (E = 2.13 GPa and v = 0.3) [21].
Loading and boundary conditions
Contrary to the femur condyles, the sectioned femur shaft was considered fully constrained, simulating the patient lying on his back during the procedure. Therefore, the proximal femur remained rigid and the distal osteotomy surface was distracted (LOW) with a remote displacement vector in axial direction (5, 7.5 and 10 mm) at the border of the lateral distal osteotomy surface. For MCW, the distal osteotomy surface was reduced by applying a remote displacement vector in axial direction (5, 7.5 and 10 mm) on the medial condyle, in line with the medial distal osteotomy surface (Fig. 1).
Outcome measures
A hinge region was defined as the area of interest, corresponding to the area where hinge fractures usually occur [42] (Fig. 1). The mesh element number for the cortical bone volume showed a congruence of 96% averaged for all hinge widths (Table 1). Higher stress and strain in the cortical than the cancellous bone is expected: cortical bone has a higher Young's modulus and consequently carries more load on wedge opening or closing. Though cancellous bone responses are included in this study, the prevention of unwanted cortical bone fracture is of greatest interest and thus cortical bone was the main focus. Because bone fractures are strain-determined [31], primary outcome measure was the cortical bone maximum principal strain and minimum principal strain in the hinge area. According to the crack tip stress theory [7], for an infinitesimally fine crack tip (zero radius) wedge opening would cause stress at the apex of the crack tip to approach infinity. Crack tip proximity can give falsely elevated peak stresses and strains, therefore, the mean principal strain in the hinge region was selected as an outcome measure in preference. In order to establish hinge fracture potential, a risk for fracture (RF) was calculated. RF was defined as the ratio between either mean tensile or mean compressive strain in the hinge region and the corresponding ultimate strain [3, 39]:
εmax: is the mean tensile/compressive strain in the hinge region
εlim: is the ultimate strain. The ultimate strain for bone is different under compressive and tensile conditions.
According to Bayraktar et al. [3], the ultimate compressive strain is εlim-compressive = 0.0104, whereas the ultimate tensile strain was 70% of that value (εlim-tensile = 0.0073). RF values greater than 1 indicated a certain fracture and RF values lower than 1 indicated no fracture occurrence.
Results
Amount of correction and cortical thickness
For LOW (Fig. 2), there was compressive strain at the outer cortex, with maximal compressive strain proximal to the hinge. Tensile strain occurred at the inner cortex. The behavior of strain was reversed in case of MCW (Fig. 2). The larger the correction, the larger the strain. In LOW, doubling the correction from 5 to 10 mm increased mean tensile and compressive strain by 50% (Table 2). In MCW the mean tensile and compressive strain increased by 48% for a 10 mm instead of a 5 mm closed-wedge (Table 2). The linear correlation is shown in Fig. 3.
LOW (upper figure) and MCW (lower figure) with different hinge widths of 5, 7.5, and 10Â mm. The cross-section area of the hinge is visible in the upper row. Lower row shows strain distribution in the hinge area. LOW: maximum (blue) and minimum (red) principle compressive strain. MCW: maximum (red) and minimum (blue) principle tensile strain. MCW: Medial closed wedge; LOW: Lateral open wedge
Upper figures: Linear correlation between strain increase (%) and amount of correction for LOW (left) and MCW (right). No difference is shown between different cortical thicknesses (1.5 versus 3Â mm). Lower figures: Tensile and compressive strain (%) plotted for different hinge widths for LOW (left) and MCW (right), for cortical thickness of 1.5 and 3Â mm, and averaged for all correction amounts. MCW: Medial closed wedge; LOW: Lateral open wedge
A 3Â mm compared to a 1.5Â mm cortical thickness recorded a 4% and 8% higher tensile and compressive strain in LOW and MCW, respectively.
Hinge width
A 10 mm hinge width produced the largest mean tensile and compressive strain for LOW and MCW (Fig. 3), because the strain is experienced over a larger volume of bone (Fig. 2 and 5). In LOW, a hinge of 10 mm recorded 61% higher mean strain compared to a hinge width of 5 mm (Table 2). In MCW, a hinge of 10 mm compared to 5 mm recorded 32% higher mean strain respectively (Table 2).
Strain in LOW versus MCW
Comparing LOW and MCW with a 5Â mm hinge regarding strain encountered in the hinge region (Table 3), MCW recorded 71% higher mean strain. Hinge width and strain difference was inversely proportional, so that strain difference between MCW and LOW declined with wider hinges. Thus, strain was only 40% higher in MCW compared to LOW in case of a 10Â mm hinge (Table 3).
Fracture risk
In LOW, a hinge fracture was recorded only in case of a 10 mm hinge in combination with a 10 mm open-wedge correction for tensile strain on the inner cortex (Table 4). In MCW, a RF > 1 was calculated for each simulation (Table 4). Independent of the hinge width, every scenario with a 10 mm closed-wedge correction showed an RF of > 1. For a 7.5 mm or 5 mm correction, the cortical bone failed if the hinge was 7.5 mm or more (Table 4).
Synopsis
Due to the morphology of the femur, the cross-section area of the hinge is significantly bigger in MCW compared to LOW (Table 5). This results in a higher risk of hinge fractures in MCW, especially in case of larger corrections. If a LOW is performed with the goal of 10Â mm correction, the hinge should not exceed a thickness of 7.5Â mm. For MCW, however, the hinge needs to be thinned out more extensively, since a 7.5Â mm closed-wedge correction with a 5Â mm hinge was the only scenario where no hinge fracture occurred.
Discussion
The most important findings of this study were that a higher correction amount as well as a larger hinge width increase the risk for a hinge fracture in DFO. Moreover, due to morphological features of the medial femoral condyle, the fracture risk is higher in MCW compared to LOW.
The hinge fracture being an accepted risk factor for malunion and loss of correction after DFO [15, 18], substantial efforts have been made to mitigate the risk of unstable cortical hinge fractures in DFO [23, 37, 44]. This study is in accordance with recent literature where technical and constitutional risk factors were investigated and larger osteotomy gaps were associated with an increased risk of hinge fractures [26, 30, 34, 37, 44]. However, different results were reported regarding the role of hinge width. In HTO, Ogawa et al. [34] and Nakamura et al. [30] showed that a deep osteotomy involving both anterior and posterior cortices reduces the risk of a hinge fracture compared to a shallow osteotomy. Brinkman et al. [4] suggested an optimal hinge width of 10 mm. In LOW DFO, a mean hinge width of 8.8 ± 4.2 mm showed significantly less hinge fractures compared to 7.2 ± 5.1 mm in the study of Winkler et al. [44]. Nonetheless, this finding could not be confirmed for lateral closing-wedge DFO by the same study group [37]. From a biomechanical point of view however, the hinge should be less than the currently established 10 mm. According to the newly gained insights provided by this study, a remaining hinge of 5 mm records the lowest amount of strain and thus carries the lowest risk of fracture. Furthermore, a smaller hinge requires less force to apply the correction according to Hooke's law and makes the procedure more controllable. The elastic deformation of the hinge can be described by the Hooke's law and is inversely proportional to the elastic modulus of cortical bone and the hinge-cross section, thus also depending on the cortical thickness of the hinge region. This study recorded a 4% respectively 8% tensile and compressive strain decrease for 1.5 mm instead of 3 mm cortical thickness in the hinge region. Kim et al. [23] proposed an ideal lateral hinge position for MCW at the upper edge of the lateral condyle with the rational of not only having stable soft tissue coverage but also low cortical bone density and therefore increased capacity of deformation. These findings support the theory of aiming for hinge positions with low cortex thickness or low cortical density to reduce the fracture risk.
However, this model should not be applied to infinity by creating a remaining hinge as small as possible. At some point the remaining hinge will be so small that a microcrack created during sawing or correction will fracture the hinge, which might be a possible explanation for the above-mentioned findings in clinical studies. Such microcrack occurs at the apex of the hinge during DFO and the bone resists such stress-risers to much greater extent than anticipated on the basis of its elastic or elastoplastic properties [11, 25]. However, no method is currently available to achieve a balance between creating a more favorable strain environment and preserving cortical bone stock to minimize hinge fracture risk. Further cadaveric studies are needed to assess the relation of a thin remaining hinge and possible microcrack propagation.
Several studies have been performed to find the best approach for varus-producing DFO. MCW has the advantages of direct bony contact, which leads to inherent stability and reliable bony healing compared to the need for potential bone grafting in large LOW corrections [46]. However, nonunion rates of LOW do not appear to be inferior to MCW [28, 45]. Hardware irritation is less frequently reported in MCW compared to LOW, where the plate is placed directly underneath the iliotibial band, leading to a higher hardware removal rate [20, 45]. LOW, on the other hand, allows correction and adjustment of distraction in order to optimize the mechanical leg axis, and places the plate on the mechanically preferred tension side [46]. Nevertheless, according to the new insights found in this study, several additional mechanical properties must be assigned to either of two DFO options. Due to the morphology of the distal femur, which is responsible for larger remaining hinge cross-section area in MCW compared to LOW osteotomies, MCW bears a higher risk of hinge fractures. In detail, there was a higher risk of fracture for all tested MCW configurations. Therefore, LOW seems to be the preferable DFO if larger corrections are intended, as a thicker hinge can remain without increasing the risk of a hinge fracture. If MCW is performed, due to the potential advantage of direct bony contact, a smaller hinge width than the currently established 10Â mm should be achieved.
Overall, the findings of this study do not allow prediction whether a fractured hinge in LOW or MCW is more prone to instability. Due to the amount of remaining hinge material in MCW, a fracture might not propagate through the entire hinge and be more stable. This FE study focused on hinge fractures, but not on their extension, dislocation or potential correction loss. Figure 2 allows us to hypothesize that the hinge fracture in LOW resulting from tensile strain at the inner cortex spreads to the area of maximal compressive strain, which is proximal to the hinge. However, according to Winkler et al. [44], only 20% of hinge fractures seem to extend proximally (referred to as type 3 hinge fractures).
This study has some limitations. A femur statistical shape model with a constant cortical thickness was used that does not reflect the graduated trabecular structure of distal femur cancellous bone. Nonetheless, an anisotropic heterogeneous bone was modeled using isotropic and homogenous material properties and a linearly elastic analysis was performed, making the results applicable to the statistical shape femora model. This is a common method and does not discredit the differences found between geometries of both osteotomies [40]. A further limitation of this study is the fact that surrounding soft tissue, including tendons and muscles, was neglected. This might affect the findings, since some studies have suggested predestined hinge locations based on soft tissue coverage to reduce the risk of hinge fracture [23, 33]. Furthermore, the simulated load conditions reflect only the intraoperative situation of MCW and LOW DFO before plating.
Conclusion
LOW demonstrates a lower intraoperative hinge fracture risk compared to MCW in varus-producing DFO due to the morphological features of the medial femoral condyle. Therefore, LOW seems to be the preferable option if larger corrections are intended, as a thicker hinge can remain without increasing the risk of a hinge fracture. If MCW is performed, the currently established hinge width of approximately 10Â mm should be reduced. Due to the limitation of the study regarding microfracture propagation a recommendation regarding the ideal hinge width in clinical use can not be given.
References
Agneskirchner JD, Hurschler C, Stukenborg-Colsman C, Imhoff AB, Lobenhoffer P (2004) Effect of high tibial flexion osteotomy on cartilage pressure and joint kinematics: a biomechanical study in human cadaveric knees. Winner of the AGA-DonJoy Award 2004. Arch Orthop Trauma Surg 124:575–584
Batista BB, Volpon JB, Shimano AC, Kfuri M Jr (2015) Varization open-wedge osteotomy of the distal femur: comparison between locking plate and angle blade plate constructs. Knee Surg Sports Traumatol Arthrosc 23:2202–2207
Bayraktar HH, Morgan EF, Niebur GL, Morris GE, Wong EK, Keaveny TM (2004) Comparison of the elastic and yield properties of human femoral trabecular and cortical bone tissue. J Biomech 37:27–35
Brinkman J-M, Lobenhoffer P, Agneskirchner J, Staubli A, Wymenga A, Van Heerwaarden R (2008) Osteotomies around the knee: patient selection, stability of fixation and bone healing in high tibial osteotomies. The Journal of bone and joint surgery. British 90:1548–1557
Cameron JI, McCauley JC, Kermanshahi AY, Bugbee WD (2015) Lateral opening-wedge distal femoral osteotomy: pain relief, functional improvement, and survivorship at 5 years. Clin Orthop Relat Res 473:2009–2015
Chang G, Rajapakse CS, Babb JS, Honig SP, Recht MP, Regatte RR (2012) In vivo estimation of bone stiffness at the distal femur and proximal tibia using ultra-high-field 7-Tesla magnetic resonance imaging and micro-finite element analysis. J Bone Miner Metab 30:243–251
Csernátony Z, Kiss L, Manó S (2008) A new technique of wedge osteotomy to diminish undesirable fractures. Eur J Orthop Surg Traumatol 18:485–488
Dewilde TR, Dauw J, Vandenneucker H, Bellemans J (2013) Opening wedge distal femoral varus osteotomy using the Puddu plate and calcium phosphate bone cement. Knee Surg Sports Traumatol Arthrosc 21:249–254
Dickschas J, Ferner F, Lutter C, Gelse K, Harrer J, Strecker W (2018) Patellofemoral dysbalance and genua valga: outcome after femoral varisation osteotomies. Arch Orthop Trauma Surg 138:19–25
Ekeland A, Nerhus TK, Dimmen S, Heir S (2016) Good functional results of distal femoral opening-wedge osteotomy of knees with lateral osteoarthritis. Knee Surg Sports Traumatol Arthrosc 24:1702–1709
Feng Z, Rho J, Han S, Ziv I (2000) Orientation and loading condition dependence of fracture toughness in cortical bone. Mater Sci Eng, C 11:41–46
Fliss B, Luethi M, Fuernstahl P, Christensen AM, Sibold K, Thali M et al (2019) CT-based sex estimation on human femora using statistical shape modeling. Am J Phys Anthropol 169:279–286
Flury A, Hoch A, Andronic O, Fritz B, Imhoff FB, Fucentese SF (2021) Increased femoral antetorsion correlates with higher degrees of lateral retropatellar cartilage degeneration, further accentuated in genu valgum. Knee Surg Sports Traumatol Arthrosc 29:1760–1768
Flury A, Jud L, Hoch A, Camenzind RS, Fucentese SF (2021) Linear influence of distal femur osteotomy on the Q-angle: one degree of varization alters the Q-angle by one degree. Knee Surg Sports Traumatol Arthrosc 29:540–545
Forkel P, Achtnich A, Metzlaff S, Zantop T, Petersen W (2015) Midterm results following medial closed wedge distal femoral osteotomy stabilized with a locking internal fixation device. Knee Surg Sports Traumatol Arthrosc 23:2061–2067
Frings J, Krause M, Akoto R, Wohlmuth P, Frosch KH (2018) Combined distal femoral osteotomy (DFO) in genu valgum leads to reliable patellar stabilization and an improvement in knee function. Knee Surg Sports Traumatol Arthrosc 26:3572–3581
Fucentese SF, Tscholl PM, Sutter R, Brucker PU, Meyer DC, Koch PP (2019) Bone autografting in medial open wedge high tibial osteotomy results in improved osseous gap healing on computed tomography, but no functional advantage: a prospective, randomised, controlled trial. Knee Surg Sports Traumatol Arthrosc 27:2951–2957
Fujita K, Sawaguchi T, Goshima K, Shigemoto K, Iwai S (2021) Influence of lateral hinge fractures on biplanar medial closing-wedge distal femoral osteotomy for valgus knee: a new classification of lateral hinge fracture. Arch Orthopaedic Trauma Surge. https://doi.org/10.1007/s00402-021-04212-4
He QF, Sun H, Shu LY, Zhu Y, Xie XT, Zhan Y et al (2018) Radiographic predictors for bone mineral loss: Cortical thickness and index of the distal femur. Bone Joint Res 7:468–475
Jacobi M, Wahl P, Bouaicha S, Jakob RP, Gautier E (2011) Distal femoral varus osteotomy: problems associated with the lateral open-wedge technique. Arch Orthop Trauma Surg 131:725–728
Kayabasi O, Ekici B (2007) The effects of static, dynamic and fatigue behavior on three-dimensional shape optimization of hip prosthesis by finite element method. Mater Des 28:2269–2277
Kim SC, Kim DH, Lee JI, Kim TW, Lee YS (2021) Comparative analysis of serial union patterns after opening-wedge high Tibial osteotomy with and without bone-void fillers. J Bone Joint Surg Am 103:1788–1797
Kim TW, Lee MC, Cho JH, Kim JS, Lee YS (2019) The ideal location of the lateral hinge in medial closing wedge osteotomy of the distal femur: analysis of soft tissue coverage and bone density. Am J Sports Med 47:2945–2951
Kosashvili Y, Safir O, Gross A, Morag G, Lakstein D, Backstein D (2010) Distal femoral varus osteotomy for lateral osteoarthritis of the knee: a minimum ten-year follow-up. Int Orthop 34:249–254
Lakes R, Nakamura S, Behiri J, Bonfield W (1990) Fracture mechanics of bone with short cracks. J Biomech 23:967–975
Lee O-S, Lee YS (2018) Diagnostic value of computed tomography and risk factors for lateral hinge fracture in the open wedge high tibial osteotomy. Arthroscopy 34:1032–1043
Lee S-S, Celik H, Lee D-H (2018) Predictive factors for and detection of lateral hinge fractures following open wedge high tibial osteotomy: plain radiography versus computed tomography. Arthroscopy 34:3073–3079
Liska F, Voss A, Imhoff FB, Willinger L, Imhoff AB (2018) Nonunion and delayed union in lateral open wedge distal femoral osteotomies—a legitimate concern? Int Orthop 42:9–15
Lüthi M, Gerig T, Jud C, Vetter T (2017) Gaussian process morphable models. IEEE Trans Pattern Anal Mach Intell 40:1860–1873
Nakamura R, Komatsu N, Fujita K, Kuroda K, Takahashi M, Omi R et al (2017) Appropriate hinge position for prevention of unstable lateral hinge fracture in open wedge high tibial osteotomy. Bone Joint J 99:1313–1318
Nalla RK, Kinney JH, Ritchie RO (2003) Mechanistic fracture criteria for the failure of human cortical bone. Nat Mater 2:164–168
Nha KW, Ha Y, Oh S, Nikumbha VP, Kwon SK, Shin WJ et al (2018) Surgical treatment with closing-wedge distal femoral osteotomy for recurrent patellar dislocation with genu Valgum. Am J Sports Med 46:1632–1640
Nha KW, Lee SH, Rhyu IJ, Kim HJ, Song JG, Han JH et al (2016) Safe zone for medial open-wedge supramalleolar osteotomy of the ankle: a cadaveric study. Foot Ankle Int 37:102–108
Ogawa H, Matsumoto K, Akiyama H (2017) The prevention of a lateral hinge fracture as a complication of a medial opening wedge high tibial osteotomy: a case control study. The Bone & Joint Journal 99:887–893
Pietsch M, Hochegger M, Winkler M, Sandriesser S, Freude T, Augat P (2019) Opening-wedge osteotomies of the distal femur: minor advantages for a biplanar compared to a uniplanar technique. Knee Surg Sports Traumatol Arthrosc 27:2375–2384
Raja Izaham RM, Abdul Kadir MR, Abdul Rashid AH, Hossain MG, Kamarul T (2012) Finite element analysis of Puddu and Tomofix plate fixation for open wedge high tibial osteotomy. Injury 43:898–902
Rupp MC, Winkler PW, Lutz PM, Irger M, Forkel P, Imhoff AB, et al. (2021) Dislocated hinge fractures are associated with malunion after lateral closing wedge distal femoral osteotomy. Knee Surg Sports Traumatol Arthrosc;https://doi.org/10.1007/s00167-021-06466-2
Saithna A, Kundra R, Getgood A, Spalding T (2014) Opening wedge distal femoral varus osteotomy for lateral compartment osteoarthritis in the valgus knee. Knee 21:172–175
Schileo E, Dall’ara E, Taddei F, Malandrino A, Schotkamp T, Baleani M et al (2008) An accurate estimation of bone density improves the accuracy of subject-specific finite element models. J Biomech 41:2483–2491
Scott C, Eaton M, Nutton R, Wade F, Evans S, Pankaj P (2017) Metal-backed versus all-polyethylene unicompartmental knee arthroplasty: proximal tibial strain in an experimentally validated finite element model. Bone & Joint Research 6:22–30
Stoffel K, Stachowiak G, Kuster M (2004) Open wedge high tibial osteotomy: biomechanical investigation of the modified Arthrex Osteotomy Plate (Puddu Plate) and the TomoFix Plate. Clin Biomech (Bristol, Avon) 19:944–950
Takeuchi R, Ishikawa H, Kumagai K, Yamaguchi Y, Chiba N, Akamatsu Y et al (2012) Fractures around the lateral cortical hinge after a medial opening-wedge high tibial osteotomy: a new classification of lateral hinge fracture. Arthroscopy 28:85–94
Weng P-W, Chen C-H, Luo C-A, Sun J-S, Tsuang Y-H, Cheng C-K et al (2017) The effects of tibia profile, distraction angle, and knee load on wedge instability and hinge fracture: a finite element study. Med Eng Phys 42:48–54
Winkler PW, Rupp MC, Lutz PM, Geyer S, Forkel P, Imhoff AB, et al. (2020) A hinge position distal to the adductor tubercle minimizes the risk of hinge fractures in lateral open wedge distal femoral osteotomy. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-020-06244-6
Wylie JD, Jones DL, Hartley MK, Kapron AL, Krych AJ, Aoki SK et al (2016) Distal femoral osteotomy for the valgus knee: medial closing wedge versus lateral opening wedge: a systematic review. Arthroscopy 32:2141–2147
Wylie JD, Maak TG (2016) Medial closing-wedge distal femoral osteotomy for genu Valgum with lateral compartment disease. Arthrosc Tech 5:e1357–e1366
Wylie JD, Scheiderer B, Obopilwe E, Baldino JB, Pavano C, Macken CJ et al (2018) The effect of lateral opening wedge distal femoral varus osteotomy on tibiofemoral contact mechanics through knee flexion. Am J Sports Med 46:3237–3244
Acknowledgements
Sandro Fucentese is a paid consultant for Medacta SA, Karl Storz, Zimmer Biomet and a member of the EKA-ESSKA Osteotomy Committee.
Funding
This research did not receive any specifc grant from funding agencies in the public, commercial, or not-for-proft sectors.
Author information
Authors and Affiliations
Contributions
MM: writing, revising, creating CAD models, performing FEA, data read-out and interpretation; FA: writing, revising, data interpretation; CZ: data read-out and interpretation; SMZ: writing, revising; LV: contributions to the conception, revising; JGS: contributions to the conception, revising; SFF: contributions to the conception, revising. All authors approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This study complies with ethical standards.
Consent for publication
Not applicable.
Competing interest
The remaining authors declare no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Meisterhans, M., Flury, A., Zindel, C. et al. Finite element analysis of medial closing and lateral opening wedge osteotomies of the distal femur in relation to hinge fractures. J EXP ORTOP 10, 33 (2023). https://doi.org/10.1186/s40634-023-00597-w
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s40634-023-00597-w
Keywords
- Distal femur osteotomies
- Medial closed wedge
- Lateral open wedge
- Finite element analysis
- Statistical shape model
- Hinge fracture risk
- Biomechanics