Free-floating medial meniscus implant kinematics do not change after simulation of medial open-wedge high tibial osteotomy and notchplasty

Abstract

Purpose

The purpose of this in-vitro study was to examine the kinematics of an artificial, free-floating medial meniscus replacement device under dynamic loading situations and different knee joint states.

Methods

A dynamic knee simulator was used to perform dynamic loading exercises on three neutrally aligned and three 10° valgus aligned (simulating a medial openwedge high tibial osteotomy - MOWHTO) left human cadaveric knee joints. The knee joints were tested in three states (intact, conventional notchplasty, extended notchplasty) while 11 randomised exercises were simulated (jump landing, squatting, tibial rotation and axial ground impacts at 10°, 30° and 60° knee joint flexion) to investigate the knee joint and implant kinematics by means of rigidly attached reflective marker sets and an according motion analysis.

Results

The maximum implant translation relative to the tibial plateau was < 13 mm and the maximum implant rotation was < 19° for all exercises. Both, the notchplasties and the valgus knee alignment did not affect the device kinematics.

Conclusions

The results of the present in-vitro study showed that the non-anchored free-floating device remains within the medial knee joint gap under challenging dynamic loading situations without indicating any luxation tendencies. This also provides initial benchtop evidence that the device offers suitable stability and kinematic behaviour to be considered a potential alternative to meniscus allograft transplantation in combination with an MOWHTO, potentially expanding the patient collective in the future.

Introduction

Although most meniscus tear patterns are indicated for arthroscopic repair, some require partial or even total meniscectomy [4, 22, 26]. The amount of resected meniscal tissue during such meniscectomy procedures is positively correlated to the risk for the development of knee joint osteoarthritis (OA) [19, 31, 32]. Therefore, artificial scaffold-based replacements [16, 48], three-dimensional (3D)-printed hydrogels [42], allografts [24, 36], autografts [21, 34] and synthetic prostheses including the NUsurface® meniscus implant (Active Implants LLC., Memphis, TN, USA) were developed to overcome the detrimental long-term impact of a meniscectomy on knee joint health [27, 35, 41]. Clinical studies have demonstrated that the NUsurface® device elicits clinical improvements compared to non-surgical treatments [2, 17, 28, 50]. The lateral wall of the non-anchored, free-floating and discoid-shaped device fits within the intercondylar space. Morphological changes of the intercondylar notch are influenced by age and sex differences [18]. Furthermore, the formation of notch osteophytes or a stenotic intercondylar notch is commonly observed in patients with OA [8, 12, 20, 30] and, therefore, occur in candidates that are indicated for the device. These are usually detected during pre-operative imaging and frequently make notchplasty, a widening procedure of the intercondylar notch, a necessary step within the implantation procedure with the aim to reduce the impingement of the device’s lateral wall by achieving a more exact femur-conforming fit [28]. However, due to the free-floating design of the device, an extensive notchplasty might lead to a general instability of the knee joint and, thus, also to device instability leading to an increased risk for luxation [33], compared to the generally low risk for device luxation [47]. Elsner et al. evaluated the long-term performance of the device during the simulation of 5 million gait-like cycles and found no device dislodgement from the joint space [10]. Therefore, the question arises as to whether there is a tissue resection limit to maintain both knee joint and device stability during the normal range of motion (ROM). Consequently, the first aim of this study was to determine the impact of two consecutive notchplasties with increasing tissue removal on the kinematics of the device during different dynamic activities.

Moreover, the importance of limb alignment in knee joint salvage procedures have been described [3, 6, 7, 14]. In the case of congenital or acquired lower limb deformities, the mechanical axis might be altered, which increases the risk for abnormal knee kinematics and concomitant premature joint degeneration [25, 45]. Therefore, operative correction of the mechanical axis prolongates the onset of knee joint OA [5]. The indications for a medial open-wedge high tibial osteotomy (MOWHTO) include varus deformity, mild degeneration, (partial) meniscectomy, knee joint instability and medial compartment overloading [7]. Takeuchi et al. reported an average correction angle of 10° valgus after an MOWHTO [44], leading to an opening of the medial joint space [49] and a significantly reduced tibiofemoral contact load in the medial compartment [1, 38]. Gelber et al. compared in their prospective study on two groups of 30 patients (< 65 years) the functional treatment outcome of a combined OWHTO and partial meniscectomy procedure with a combined OWHTO and partial meniscus replacement procedure and found no short-term patient satisfaction differences between the groups [14]. However, in the case of young (< 40 years), compliant patients who had undergone total meniscectomy and displayed clear pathological signs of mild degeneration states (grade 1 or 2) [6], a combined MOWHTO and meniscus allograft transplantation (MAT) procedure might be indicated [3, 6, 23]. Though currently not included as a use case in the device’s instructions for use, the present biomechanical benchtop study initially assesses the suitability of NUsurface® as an alternative to MAT, by evaluating the impact of an MOWHTO on the device stability: Following an MOWHTO in which the loading of the medial compartment is reduced, the freedom of movement of the femur-conforming device might increase, and with it, there might be a higher luxation risk for the device, particularly in combination with a notchplasty. Therefore, the second aim of this study was to investigate the impact of a 10° shift of knee alignment towards valgus in combination with two consecutive notchplasties on the kinematics of the device during different dynamic activities.

Methods

Specimen preparation

Following IRB approval (IRB: 34/19), six fresh-frozen left human knee joints (2 female, 4 male; age: 60.8 ± 7.5 years, BMI: 26.5 ± 3.9) were obtained from an official tissue bank and their soft tissues were resected, leaving the capsuloligamentous structures intact (Fig. 1). Steel cables were firmly connected to bi-cortically anchored pins which were screwed in at the centre of the original muscle insertion sites (Fig. 2-B) to allow simulation of the seven major flexor and extensor knee muscles by utilizing seven pneumatic actuators (Fig. 2-A). Following these initial preparation steps, the knee joints were mounted in an upright position in an Oxford-rig dynamic knee simulator to enhance the further preparation process. Subsequently, two experienced surgeons (SH and VC) performed the arthroscopic NUsurface® device implantation, while starting with a subtotal medial meniscectomy leaving a 2–4 mm meniscal rim intact to provide device confinement. To allow for proper device sizing, a preceding 3D geometrical analysis, based on magnetic resonance imaging (MRI), was performed. The device of the correct size was then inserted into the medial compartment using the designated arthroscopic instruments. The knee joints were tested in three states (Fig. 2): native, conventional notchplasty (Fig. 3-A) and extended notchplasty (Fig. 3-B). Following native testing, the surgeons (SH and VC) performed the consecutive arthroscopic notchplasties, starting with the conventional notchplasty. Following testing, 3D MRI scans of each knee were performed with the device in-situ to evaluate the resected tissue amount during the extended notchplasty. For kinematical evaluation of the knee joint and the relative device kinematics, passive reflective rigid-body-markers were attached to the antero-medial aspect of the device (Fig. 2-B) and to the bony landmarks at the femur and tibia (Fig. 2-A). The small device marker was self-developed, in-house 3D-printed and equipped with reflective foil (Scotchlite 7610, 3 M, Minnesota, USA) (Fig. 2, B). The knee joints were kept moist throughout the preparation and testing process using saline solution.

Fig. 1
figure 1

MRI scans in the frontal view (A–P): A) intact left knee joint and B) resected, with capsuloligamentous structures intact in the left knee joint (white tissue capsule), applied notchplasty (green arrow) with muscle insertion sites (blue arrow, hamstring muscles) and device left in place (red dashed line, black shadow)

Fig. 2
figure 2

A Dynamic knee simulator with embedded specimen, muscle simulation via steel cables and passive reflective rigid marker bodies at the femur, tibia and device. B View at the meniscus replacement device, inserted in the medial joint compartment and an attached passive reflective rigid marker body that allows for motion analysis

Fig. 3
figure 3

Schematic illustration of the two different notchplasty states applied under arthroscopic surgery in the medial compartment: A) Conventional tissue resection (Notch) and B) extended tissue removal (xNotch) in the intercondylar notch

Biomechanical testing

An Oxford-rig knee simulator, which allowed physiological replication of the whole leg, including the hip joint with an adjustable caput-collum-diaphyseal (CCD) angle, femur, tibia and ankle joint, was used to perform dynamic exercises on cadaveric knee joints [37, 39] while providing all physiological degrees of freedom. Prior to testing, a pilot trial on one additional cadaveric knee joint confirmed that the removal of skin, subcutaneous fat and muscles did not affect the outcome measures.

The specimens were divided into two groups to study the effect of a simulated MOWHTO procedure on device stability: under neutral alignment (n = 3) and under 10° alignment shift towards valgus (n = 3) representing the post-MOWHTO state [44]. The alignment shift was achieved by altering the CCD angle by 10°. For each condition, a total of 11 impact exercises were randomly simulated: jump landing, starting at 10° to a flexion angle of 50° at a velocity of 180°/s and back to the initial 10° starting position at a velocity of 120°/s (10°-50°-10°); squatting (10°-70°-10°) at a flexion velocity of 5°/s, axial ground impacts (50 mm) and internal/external tibial impact rotation (± 10°). The axial impacts and tibial rotation were applied at 10°, 30° and 60° knee flexion angles. During the axial impact simulations, the mechanical hip joint remained firmly fixed and ground impacts were applied directly through the simulated mechanical ankle joint. Both, deep knee flexions, as simulated during squatting, and internal/external impacts, which are injury mechanism of meniscus tears, were applied [13, 29].

Data acquisition

Following successful calibration in accordance with the manufacturer’s guidelines and a spatial mean error of 0.13 ± 0.01 mm, a kinematic tracking system (Prime13, OptiTrack, Oregon, USA) with seven cameras was used for motion tracking. To ensure measurement repeatability and accuracy, the calibration procedure was repeated, when the test setup or the specimens were manipulated. The motion sequences were recorded in a specific developer tool (Motive 4.0, OptiTrack) (Fig. 4-B). Export files with kinematic parameters (position, rotation) of each rigid body [15] were generated from the sequences and implemented in a customised analysis protocol (Matlab, The MathWorks, Massachusetts, USA) to determine the maximum spatial device translation and rotation (Fig. 4-A) in relation to the underlying tibial plateau. The description and measurement of six motion components of the rigid bodies were based on Grood and Suntay [15]. The rigid bodies were assumed to be non-deformable. Three translations along the three spatial axes X (medial-lateral), Y (anterior-posterior) and Z (cranial-caudal) as well as the rotation W around the corresponding Z axes (rotation of the device in axial tibial rotation) were kinematically evaluated. Therefore, two parameters were assessed to quantify the device kinematics and its stability consecutively rated: 1) absolute translation of the device in relation to the tibial plateau (device displacement, vectorial addition of X and Y) and 2) rotation of the device around its axis (device rotation, rotation around Z).

Fig. 4
figure 4

Illustrated frontal view on the knee joint: Coordinate system and the respective three translational axes as well as the rotation W around the corresponding Z axes

Statistics

Descriptive statistics were applied to present the device kinematics under particular consideration of the limb alignment and the notchplasty states (Figs. 5 and 6). All data is presented in Figs. 5 and 6 where each data point shows the maximum value of the translation or rotation under each test condition. The two groups examined are colour-coded and contrasted in the figures between neutral alignment (grey) and valgus alignment (black).

Fig. 5
figure 5

Maximum absolute translation in mm of the device during the four different exercises and three different knee states: intact knee joint (inserted device, no notchplasty), conventional notchplasty and extended notchplasty. Coloured distinction between the neutral leg axis position (grey) and the valgus alignment shift (black)

Fig. 6
figure 6

Maximum absolute rotation in ° of the device during the four different exercises and three different knee states: intact knee joint (inserted device, no notchplasty), conventional notchplasty and extended notchplasty. Coloured distinction between the neutral leg axis position (grey) and the valgus alignment shift (black)

Results

Neutral alignment

In neutral knee alignment, for both, jump landing and squatting the device translation was < 11 mm (Fig. 6) and rotation was < 19° (Fig. 5). Axial ground impact simulation resulted in maximum 6 mm translation and maximum 15° rotation under all three investigated flexion angles (10°, 30°, 60°). The consecutive tissue removal during the notchplasty preparation had no effect on device kinematics during the jump landing, squatting simulation or axial ground impact simulations. During application of 10° external and internal tibial rotation, the device translated < 4 mm (Fig. 5), while rotating < 9° (Fig. 6) under all three knee flexion states with no influence of the notchplasty state on the device kinematics.

Valgus alignment

During jump landing and squatting simulation with a 10° valgus shift simulating the average post MOWHTO-knee alignment, a maximum device translation of 9 mm (Fig. 5) and rotation of 10° (Fig. 6) were observed. Additionally, under MOWHTO knee alignment, the device kinematics were not altered after tissue resection due to notchplasty preparation during jump landing or squatting. Axial impact simulation resulted in 13 mm translation and 14° rotation, with a slight tendency for a translation increase after notchplasty tissue removal at the 10° knee flexion state. An increased device rotation was observed for progressing notchplasties in all knee flexion states under axial impact simulation. The simulation of internal and external tibial rotation resulted in a maximum device translation of 9 mm and 13° rotation. The device was less stable when applying external tibial rotation (< 9 mm; 13°) compared to internal rotation (< 6 mm; 7°).

Discussion

The most important finding of this study is that the non-anchored, free-floating device remained in its intended position without luxation under the investigated conditions. The extent of tissue removal during the consecutive notchplasty preparations had no adverse effect on the device. This was also observed in the knee joints with consecutive notchplasties that were investigated following a 10° alignment shift towards valgus, which is similar to a common correction angle after an MOWHTO procedure. To the best of the authors’ knowledge, this is the first study which investigates the influence of two consecutive notchplasty states and different leg alignments on the kinematics of a total meniscus replacement device under dynamic loading simulations.

The results of the present study are in agreement with the range of clinically measured translations [9]. In an open MRI study, translations of up to 10.4 mm of the device during 0° and 120° knee flexion and combined weight-bearing conditions were measured [3], indicating similar translation levels. Furthermore, the device kinematics were close to translation ranges of native medial menisci, where translations of up to 8 mm in a non-weight-bearing setup were measured [46]. The reason for the slightly increased translation of the device will be the prevalence of the high physiological contact forces in our setup. Tissue removal during the notchplasty may play a key role in the assessment of the kinematic device behaviour. Therefore, two expert surgeons who were familiar with the implantation technique performed the notchplasty, comparable to the clinical situation. This notchplasty state was described as the conventional state. Notch stenosis and the formation of notch osteophytes are common corollaries, indicators and even predictors of OA in the knee joint [8, 12, 20, 30]. Therefore, notch stenosis and/or notch osteophytes are expected to occur regularly in suitable candidates for the device. To function as intended, the lateral wall of the device requires adequate room between the intercondylar notch tibial spine and cruciate ligaments. When notch osteophytes articulate against the device, they may accelerate wear formation. A patient’s individual joint condition, such as the location and size of notch osteophytes or the degree of notch stenosis, requires clinicians to make a decision regarding the required degree of notchplasty on a case-by-case basis. This may result in the need for a more extensive notchplasty and concomitantly more tissue removal compared to the conventional state, particularly in the case of a stenotic intracondylar notch. Therefore, two notchplasty states with increasing tissue removal were investigated in the present study to assess their impact on the device kinematics. Furthermore, Gelber et al. and Song et al. have described a positive effect of a combined notchplasty and OWHTO procedure that led to an improved flexion-extension rage recovery which was maintained over time [14, 40], which might be a potential positive side-effect of the notchplasty intervention. The simulated valgus shift in alignment after the MOWHTO induced a joint gap widening on the medial compartment which resulted in a slight translational and rotational increase of the device when compared to neutral leg alignment. This could be mainly attributed to the joint gap widening, and to the concurrently reduced contact pressure on the medial compartment, which appears to move the NUsurface® device more compared to the normal leg alignment conditions [43]. The used knee joint simulator is able to perform dynamic loading situations on cadaveric knee joints that result in physiological tibiofemoral contact pressures [39]. Therefore, the contact pressure between the femur and tibia as well as the resulting contact transmission of the device in the present study could be assumed to be adequate.

This study contributes to existing knowledge of the in-situ biomechanics of the device. Combined knee joint axial loading and internal or external rotation leads to extensive meniscus rotation and according shear stresses in the meniscal body. These shear stresses are likely to result in a meniscal tear and, thus, represent one of the most common injury mechanisms [13, 29]. Similarly, an extensive device rotation would be also likely to result in a device luxation. In this case, the luxation would occur due to a twisting movement and related impingement of the intrinsic geometrical fit of the surface-conforming device. However, critical device behaviour in terms of luxation was not evident from the rotation results of the present study.

Limitations

This in-vitro study has some limitations. First, the small sample size for each leg alignment group (n = 3) allowed only a qualitative and descriptive evaluation. Despite the low number of samples, a post-hoc power analysis indicated a power of 1 − β > 0.4 for the kinematic differences of the device between the intact and extended notchplasty states and 1 − β > 0.3 for the differences between the leg alignment states (G*Power 3.1 [11];; α error = 0.05, n = 3). Therefore, it is assumed that the results of the present study represent promising trends that are likely to be statistically underlined when using a sample size of n = 6 (G*Power 3.1; α error = 0.05; power (1 − β) = 0.6) per group. Moreover, it should be mentioned, that despite the low sample size number, we did not see any luxation of the device under the investigated dynamic loading simulations. It should be further noted that a few data points are lacking in the results. The reason for this might be that the optical evaluation led to falsified data due to distortions at the experimental setup and, therefore, it was decided to exclude the respective measurements. Again, also for these simulations where data points had to be removed, the device remained within the joint gap without signs of luxation. Second, the investigated dynamic simulations do not reflect daily activities. However, the simulations and particularly the drop jump and rotation impact sequences represented challenging tasks to provoke a device failure compared to, for example, level walking simulation that includes lower flexion speeds and a reduced total ROM. Third, the MOWHTO was not surgically performed in vitro but simulated by adjusting the leg alignment by an altered CCD angle that resulted in a 10° valgus shift of the knee joint angle. Due to the combination of a slightly compromised stability of the knee joint, which was induced by soft tissue removal, and the high mechanical loads that result in realistic tibiofemoral contact pressures [39] that are applied to the knee joints could have led to lateral hinge fractures after MOWHTO surgery. In addition, it would not have been possible to clearly determine the cause of the obtained results, because the surgical interventions of the MOWHTO and the notchplasty could potentially influence each other. Particularly in contracted knees with a stiff ligamentous situation, it would have been necessary to perform a partial release of the medial collateral ligament to obtain the desired valgisation of the joint line [38], thus also affecting the general kinematics of the knee joint. For these two reasons it was rather decided to perform a kinematic alignment shift instead of a MOWHTO intervention. Future studies are required to prove the device stability under individual patient morphologies to improve the patient outcome by optimising surgical interventions.

Conclusions

The results of the present in-vitro study show that the non-anchored free-floating device remains within the medial knee joint gap under challenging dynamic loading situations without indicating any luxation tendencies, thereby also providing initial benchtop evidence that the device offers suitable stability and kinematic behaviour to be considered a potential alternative to MAT in combination with MOWHTO, potentially expanding the patient collective in the future.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

3D:

Three dimensional

BMI:

Body Mass Index

CCD :

Caput-Collum-Diaphyseal (angle)

IRB:

Institutional Review Board

MAT:

Meniscal Allograft Transplantation

MOWHTO:

Medial Open-Wedge High Tibial Osteotomy

MRI:

Magnetic Resonance Imaging

OA:

Osteoarthritis

OWHTO:

Open Wedge High Tibial Osteotomy

ROM:

Range of Motion

References

  1. Agneskirchner JD, Hurschler C, Wrann CD, Lobenhoffer P (2007) The effects of valgus medial opening wedge high tibial osteotomy on articular cartilage pressure of the knee: a biomechanical study. Arthroscopy 23:852–861

    Article  PubMed  Google Scholar 

  2. Alley R, Alfred R, Edelson R, Greenleaf J, Gersoff W, Gomoll A et al (2020) Clinical results of the NUsurface® Meniscus implant versus non-surgical controls at 24 months: data from a pooled cohort of a randomized controlled study and single arm study. Orthopaedic J Sports Med 8(7 suppl6):2325967120S00367

    Google Scholar 

  3. Amendola A (2007) Knee osteotomy and meniscal transplantation: indications, technical considerations, and results. Sports Med Arthrosc Rev 15:32–38

    Article  PubMed  Google Scholar 

  4. Anetzberger H, Birkenmaier C, Lorenz S (2014) Meniscectomy: indications, procedure, outcomes, and rehabilitation. Orthop Res Rev 6:1–9

    Google Scholar 

  5. Birmingham TB, Moyer R, Leitch K, Chesworth B, Bryant D, Willits K et al (2017) Changes in biomechanical risk factors for knee osteoarthritis and their association with 5-year clinically important improvement after limb realignment surgery. Osteoarthr Cartil 25:1999–2006

    Article  CAS  Google Scholar 

  6. Bonasia DE, Amendola A (2010) Combined medial meniscal transplantation and high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 18:870–873

    Article  PubMed  Google Scholar 

  7. Bonasia DE, Dettoni F, Sito G, Blonna D, Marmotti A, Bruzzone M et al (2014) Medial opening wedge high tibial osteotomy for medial compartment overload/arthritis in the varus knee: prognostic factors. Am J Sports Med 42:690–698

    Article  PubMed  Google Scholar 

  8. Chen C, Ma Y, Geng B, Tan X, Zhang B, Jayswal CK et al (2016) Intercondylar notch stenosis of knee osteoarthritis and relationship between stenosis and osteoarthritis complicated with anterior cruciate ligament injury: a study in MRI. Medicine (Baltimore) 95:e3439

    Article  CAS  PubMed  Google Scholar 

  9. De Coninck T, Elsner JJ, Linder-Ganz E, Cromheecke M, Shemesh M, Huysse W et al (2014) In-vivo evaluation of the kinematic behavior of an artificial medial meniscus implant: a pilot study using open-MRI. Clin Biomech (Bristol, Avon) 29:898–905

    Article  PubMed  Google Scholar 

  10. Elsner JJ, Shemesh M, Shefy-Peleg A, Gabet Y, Zylberberg E, Linder-Ganz E (2015) Quantification of in vitro wear of a synthetic meniscus implant using gravimetric and micro-CT measurements. J Mech Behav Biomed Mater 49:310–320

    Article  CAS  PubMed  Google Scholar 

  11. Faul F, Erdfelder E, Lang AG, Buchner A (2007) G*power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods 39:175–191

    Article  PubMed  Google Scholar 

  12. Ferrari MB, Mannava S, DePhillipo N, Sanchez G, LaPrade RF (2017) Notchplasty for the arthroscopic treatment of limited knee extension. Arthrosc Tech 6:e517–e524

    Article  PubMed  PubMed Central  Google Scholar 

  13. Furumatsu T, Okazaki Y, Okazaki Y, Hino T, Kamatsuki Y, Masuda S et al (2019) Injury patterns of medial meniscus posterior root tears. Orthop Traumatol Surg Res 105:107–111

    Article  PubMed  Google Scholar 

  14. Gelber PE, Isart A, Erquicia JI, Pelfort X, Tey-Pons M, Monllau JC (2015) Partial meniscus substitution with a polyurethane scaffold does not improve outcome after an open-wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 23:334–339

    Article  PubMed  Google Scholar 

  15. Grood ES, Suntay WJ (1983) A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng 105:136–144

    Article  CAS  PubMed  Google Scholar 

  16. Gruchenberg K, Ignatius A, Friemert B, von Lubken F, Skaer N, Gellynck K et al (2015) In vivo performance of a novel silk fibroin scaffold for partial meniscal replacement in a sheep model. Knee Surg Sports Traumatol Arthrosc 23:2218–2229

    Article  PubMed  Google Scholar 

  17. Hershman E, McKeon B, Kaeding C, Edelson R, Greenleaf J, Gersoff W (2021) Superior improvements in knee pain and function with a novel synthetic medial Meniscus replacement implant compared to non-surgical Care in Subjects with knee pain following partial Meniscectomy: two-year results from two prospective US clinical trials. Orthopaedic J Sports Med 9(7 suppl4):2325967121S00206

    Google Scholar 

  18. Hirtler L, Kainberger F, Rohrich S (2022) The intercondylar fossa-a narrative review. Clin Anat 35:2–14

    Article  PubMed  Google Scholar 

  19. Hunter DJ, Zhang YQ, Niu JB, Tu X, Amin S, Clancy M et al (2006) The association of meniscal pathologic changes with cartilage loss in symptomatic knee osteoarthritis. Arthritis Rheum 54:795–801

    Article  CAS  PubMed  Google Scholar 

  20. Katsuragi J, Sasho T, Yamaguchi S, Sato Y, Watanabe A, Akagi R et al (2015) Hidden osteophyte formation on plain X-ray is the predictive factor for development of knee osteoarthritis after 48 months--data from the Osteoarthritis Initiative. Osteoarthr Cartil 23:383–390

    Article  CAS  Google Scholar 

  21. Kohn D, Wirth CJ, Reiss G, Plitz W, Maschek H, Erhardt W et al (1992) Medial meniscus replacement by a tendon autograft. Experiments in sheep. J Bone Joint Surg (Br) 74:910–917

    Article  CAS  PubMed  Google Scholar 

  22. Kopf S, Beaufils P, Hirschmann MT, Rotigliano N, Ollivier M, Pereira H et al (2020) Management of traumatic meniscus tears: the 2019 ESSKA meniscus consensus. Knee Surg Sports Traumatol Arthrosc 28:1177–1194

    Article  PubMed  PubMed Central  Google Scholar 

  23. Liu JN, Agarwalla A, Gomoll AH (2019) High Tibial osteotomy and medial Meniscus transplant. Clin Sports Med 38:401–416

    Article  CAS  PubMed  Google Scholar 

  24. Lubowitz JH, Verdonk PC, Reid JB 3rd, Verdonk R (2007) Meniscus allograft transplantation: a current concepts review. Knee Surg Sports Traumatol Arthrosc 15:476–492

    Article  PubMed  Google Scholar 

  25. Maderbacher G, Baier C, Springorum HR, Zeman F, Grifka J, Keshmiri A (2016) Lower limb anatomy and alignment affect natural Tibiofemoral knee kinematics: a cadaveric investigation. J Arthroplast 31:2038–2042

    Article  Google Scholar 

  26. McDermott I (2011) Meniscal tears, repairs and replacement: their relevance to osteoarthritis of the knee. Br J Sports Med 45:292–297

    Article  PubMed  Google Scholar 

  27. McDermott ID, Amis AA (2006) The consequences of meniscectomy. J Bone Joint Surg (Br) 88:1549–1556

    Article  CAS  PubMed  Google Scholar 

  28. McKeon BP, Zaslav KR, Alfred RH, Alley RM, Edelson RH, Gersoff WK et al (2020) Preliminary results from a US clinical trial of a novel synthetic polymer meniscal implant. Orthop J Sports Med 8:2325967120952414

    Article  PubMed  PubMed Central  Google Scholar 

  29. Nielsen AB, Yde J (1991) Epidemiology of acute knee injuries: a prospective hospital investigation. J Trauma 31:1644–1648

    Article  CAS  PubMed  Google Scholar 

  30. Okada K, Yamaguchi S, Sato Y, Enomoto T, Ogawa Y, Ohtori S et al (2019) Comparison of meniscal extrusion and osteophyte formation at the intercondylar notch as a predictive biomarker for incidence of knee osteoarthritis-data from the osteoarthritis initiative. J Orthop Sci 24:121–127

    Article  PubMed  Google Scholar 

  31. Papalia R, Del Buono A, Osti L, Denaro V, Maffulli N (2011) Meniscectomy as a risk factor for knee osteoarthritis: a systematic review. Br Med Bull 99:89–106

    Article  PubMed  Google Scholar 

  32. Petty CA, Lubowitz JH (2011) Does arthroscopic partial meniscectomy result in knee osteoarthritis? A systematic review with a minimum of 8 years' follow-up. Arthroscopy 27:419–424

    Article  PubMed  Google Scholar 

  33. Ranuccio F, Familiari F, Tedesco G, La Camera F, Gasparini G (2017) Effects of Notchplasty on anterior cruciate ligament reconstruction: a systematic review. Joints 5:173–179

    Article  PubMed  PubMed Central  Google Scholar 

  34. Ronnblad E, Rotzius P, Eriksson K (2022) Autologous semitendinosus tendon graft could function as a meniscal transplant. Knee Surg Sports Traumatol Arthrosc 30:1520–1526

    Article  PubMed  Google Scholar 

  35. Roos EM, Ostenberg A, Roos H, Ekdahl C, Lohmander LS (2001) Long-term outcome of meniscectomy: symptoms, function, and performance tests in patients with or without radiographic osteoarthritis compared to matched controls. Osteoarthr Cartil 9:316–324

    Article  CAS  Google Scholar 

  36. Rosso F, Bisicchia S, Bonasia DE, Amendola A (2015) Meniscal allograft transplantation: a systematic review. Am J Sports Med 43:998–1007

    Article  PubMed  Google Scholar 

  37. Schall F, Seitz AM, Hacker S, van Drongelen S, Wolf SI, Ignatius A et al (2019) German Society of Biomechanics (DGfB) young investigator award 2019: proof-of-concept of a novel knee joint simulator allowing rapid motions at physiological muscle and ground reaction forces. Front Bioeng Biotechnol 7:244

    Article  PubMed  PubMed Central  Google Scholar 

  38. Seitz AM, Nelitz M, Ignatius A, Durselen L (2019) Release of the medial collateral ligament is mandatory in medial open-wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 27:2917–2926

    Article  PubMed  Google Scholar 

  39. Seitz AM, Schall F, Hacker SP, van Drongelen S, Wolf S, Durselen L (2021) Forces at the anterior Meniscus attachments strongly increase under dynamic knee joint loading. Am J Sports Med 49:994–1004

    Article  PubMed  Google Scholar 

  40. Song JH, Bin SI, Kim JM, Lee BS, Choe JS, Cho HK (2021) Flexion contracture can be relieved by concurrent notchplasty in medial open wedge high tibial osteotomy. Orthop Traumatol Surg Res 107:103020

    Article  PubMed  Google Scholar 

  41. Stein T, Mehling AP, Welsch F, von Eisenhart-Rothe R, Jager A (2010) Long-term outcome after arthroscopic meniscal repair versus arthroscopic partial meniscectomy for traumatic meniscal tears. Am J Sports Med 38:1542–1548

    Article  PubMed  Google Scholar 

  42. Stocco E, Porzionato A, De Rose E, Barbon S, De Caro R, Macchi V (2022) Meniscus regeneration by 3D printing technologies: current advances and future perspectives. J Tissue Eng 13:20417314211065860

    Article  PubMed  PubMed Central  Google Scholar 

  43. Sukopp M, Schall F, Hacker SP, Ignatius A, Durselen L, Seitz AM (2021) Influence of menisci on Tibiofemoral contact mechanics in human knees: a systematic review. Front Bioeng Biotechnol 9:765596

    Article  PubMed  PubMed Central  Google Scholar 

  44. Takeuchi R, Ishikawa H, Aratake M, Bito H, Saito I, Kumagai K et al (2009) Medial opening wedge high tibial osteotomy with early full weight bearing. Arthroscopy 25:46–53

    Article  PubMed  Google Scholar 

  45. Tetsworth K, Paley D (1994) Malalignment and degenerative arthropathy. Orthop Clin North Am 25:367–377

    Article  CAS  PubMed  Google Scholar 

  46. Thompson WO, Thaete FL, Fu FH, Dye SF (1991) Tibial meniscal dynamics using three-dimensional reconstruction of magnetic resonance images. Am J Sports Med 19:210–215 discussion 215–216

    Article  CAS  PubMed  Google Scholar 

  47. Verhaeghe L, Boeren K (2018) A rare complication after synthetic Meniscus replacement. J Belg Soc Radiol 102:63

    Article  PubMed  PubMed Central  Google Scholar 

  48. Vrancken AC, Buma P, van Tienen TG (2013) Synthetic meniscus replacement: a review. Int Orthop 37:291–299

    Article  PubMed  Google Scholar 

  49. Willinger L, Foehr P, Achtnich A, Forkel P, Voss A, Liska F et al (2019) Effect of lower limb alignment in medial Meniscus-deficient knees on Tibiofemoral contact pressure. Orthop J Sports Med 7:2325967118824611

    Article  PubMed  PubMed Central  Google Scholar 

  50. Zaslav KR, Farr J, Alfred R, Alley RM, Dyle M, Gomoll AH et al (2022) Treatment of post-meniscectomy knee symptoms with medial meniscus replacement results in greater pain reduction and functional improvement than non-surgical care. Knee Surg Sports Traumatol Arthrosc 30:1325–1335

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to acknowledge Patrizia Horny from the Institute of Orthopaedic Research and Biomechanics Ulm for her design support. Furthermore, we would like to thank Kilian Stumpf (Experimental Cardiovascular Imaging, Ulm University Medical Centre) and Carsten Hackenbroch, MD (Diagnostic and Interventional Radiology, Ulm University Medical Centre) for MRI support.

Funding

Open Access funding enabled and organized by Projekt DEAL. This study was funded by Active Implants LLC.

Author information

Authors and Affiliations

Authors

Contributions

MS, JS and AMS performed the specimen preparation, whereas surgical interventions have been conducted by SH and VC. MS and JS performed the joint simulator tests and conceived and designed the data analysis/evaluation. MS and AMS drafted the manuscript. MSh, EP, ELG, AI and LD supported with expert knowledge and helped in drafting. LD, MSh and AMS conceived the study. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Andreas Martin Seitz.

Ethics declarations

Ethics approval and consent to participate

The study design was approved by the institutional review board of the University of Ulm under reference number 34/19.

No informed consent was required for this study.

Consent for publication

Not applicable.

Competing interests

Maoz Shemesh, Elena Pruech and Eran Linder-Ganz are present under employment at Active Implants LLC and hold shares of Active Implants LLC. Scott Hacker and Vincenzo Condello are consultants for Active Implants LLC and get compensation for time spent in the lab according to a service contract.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Sukopp, M., Shemesh, M., Pruech, E. et al. Free-floating medial meniscus implant kinematics do not change after simulation of medial open-wedge high tibial osteotomy and notchplasty. J EXP ORTOP 10, 13 (2023). https://doi.org/10.1186/s40634-023-00576-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40634-023-00576-1