Pivot shift intraoperative quantitative assessment using a smartphone accelerometer in ACL deficient knees

Abstract

Purpose

The Pivot Shift (PS) test is a complex clinical sign that assesses the internal rotation and anterior tibial translation, which occurs abnormally in ACL deficient-knees. Because of the high inter-observer variability, different devices have been designed to characterize this complex movement in quantitative variables. The objective of this pilot study is to validate the reproducibility of intraoperative quantitative assessment of the PS with a smartphone accelerometer.

Methods

Twelve ACL-injured knees were included and compared with the contralateral uninjured side. The PS was measured by two independent observers utilizing a smartphone accelerometer and graded according to the IKDC classification. Measurements were taken preoperatively, intraoperatively and postoperatively. Intraoperative readings were taken during each stage of reconstruction or repair of meniscoligamentous lesions including meniscal lesions, ramp lesions, ACL reconstruction and lateral tenodesis. Reproducibility of the measurements were evaluated according to an intraclass correlation coefficient (ICC).

Results

The intra-observer reliability was good for the first examiner and excellent for the second examiner, with the ICC 0.89 [0.67, 0.98] p < 0,001 and ICC 0.97 [0.91, 1.0] p < 0,001 respectively. The inter-observer reliability was excellent between the two observers with the ICC 0.99 [0.97, 1.0] p < 0,001. The mean tibial acceleration measured 3.45 m.s2 (SD = 1.71) preoperatively on the injured knees and 1.03 m.s2 (SD = 0.36) on the healthy knees, demonstrating a significant difference following univariate analysis p < 0.001. Postoperatively, no significant difference was observed between healthy and reconstructed knees The magnitudes of tibial acceleration values were correlated with the PS IKDC grade.

Conclusion

The smartphone accelerometer is a reproducible device to quantitatively assess the internal rotation and anterior tibial translation during ACL reconstruction surgery. The measurements are influenced by the different surgical steps. Other larger cohort studies are needed to evaluate the specific impact of each step of the ACL reconstruction and meniscal repair on this measurement. An external validation using other technologies are needed to validate the reliability of this device to assess the PS test.

Level of evidence

Level IV, case series, pilot study.

Introduction

The Pivot Shift (PS) test is a complex clinical sign that assesses the internal rotation and anterior tibial translation, which occurs abnormally in ACL deficient-knees [5]. In the presence of an ACL injury, the PS test has a specificity ranging from 0 to 97 to 0–99 and sensitivity ranging from 0 to 18 to 0–48 [23]. The pathological nature of this test is based on a subjective evaluation in 3 grades: grade 1 = glide guard, 2: clunk and grade 3: gross instability [9]. The exact pathophysiology of the PS remains poorly understood and incompletely accepted [30]. A multifactorial background involving multiple stabilizing anatomical structures of the knee seems to be the most likely hypothesis [17]. The PS test is correlated to several factors including the anterolateral complex [26], meniscal lesions (posterolateral and posteromedial [16] specifically), iliotibial band or Kaplan Fibers [3] and the bone morphology of the tibial plateau [25]. Because of the high inter-observer variability, different devices have been designed to characterize this complex movement in quantitative variables [12]. Multiple studies have investigated the use of accelerometers [14, 22]. Accelerometers using smartphone technologies have the advantages of being easy to use and reliable [2, 7, 22]. However, accelerometers have never been used intra-operatively as part of clinical practice.

The main objective of this pilot study is to validate the reproducibility of intraoperative quantitative assessment of the PS test with a smartphone accelerometer. The secondary objective was to identify potential parameters influencing the PS during ACL reconstruction surgery.

The hypothesis of this work was that the use of a smartphone accelerometer intra-operatively is a reproducible and reliable method to measure the PS, and the measurements could be influenced by the different surgical steps.

Methods

All patients provided written consent, and the study was approved by our hospital’s ethics committee. (CPP N°2020-A00910033). A prospective analysis was performed on 12 patients who underwent ACL reconstructions between May 2021 and February 2022.

The inclusion criteria were defined as patients aged between 16 and 45 years old undergoing primary ACLR without associated collateral ligament (≤ MRI grade 2) surgery. All patients had clinical anterior instability and complete ACL tear on MRI.

Exclusion criteria were defined as history of ipsilateral knee surgery, an associated bone or cartilage procedure, tibial spine avulsion, pathological hyperlaxity (Beighton score > 3), previous history of contralateral ACLR, multi-ligament injury confirmed by MRI, chronic inflammatory joint disease.

All patients provided written consent, and the study was approved by our hospital’s ethics committee. (CPP N°2020-A00910033).

Experimental protocol

All patients were operated by two senior surgeons using two different ACLR techniques with different types of grafts and a Lateral Extra-articular Tenodesis (LET). The choice of surgical techniques was based on individual surgeons’ preferences.

Surgical ACLR technique

  • Quadriceps tendon + LET using the modified Lemaire technique (QT + LET group) [10]

  • Combined ACL and Anterolateral Ligament Reconstruction (DT3I2 group) [24]

Meniscal repair techniques

  • For posterior and middle segments, all-inside meniscal repair techniques were used (AIR +®, Stryker, Mahwah NJ, USA).

  • For the anterior segment, an outside-in suture technique was used:

  • For ramp lesions, a single posteromedial approach was used with a 25 curved suture hook device (SutureLasso; Arthrex, Naples, FL) loaded with a No. 1 absorbable monofilament suture (PDS; Ethicon, Somerville, NJ) [28].

Data collection

Two senior orthopedic surgeons were blinded to the accurate diagnosis for the pre-operative measurements (ie information regarding meniscal status was withheld). They independently performed the PS test on the patient under anesthesia, firstly on the healthy limb and then on the limb to be operated on. All pre-operative measurements were repeated three times by each examiner. For a right knee, the examiner placed themselves on the right side of the patient who was in the supine position. The left hand then held the foot and elevated it from the plane of the bed (about 20° of hip flexion and abduction). The limb was placed in slight internal rotation, while the right hand positioned the knee in slight flexion (20°), placed in the proximal third of the leg, applying a valgus stress. Large forces and extreme rotations were avoided. The dynamic phase of the test consisted of gradually flexing the knee in a passive manner. The abrupt subluxation then reduction of the lateral tibial plateau under the femoral condyles occurred between 30° and 60° of flexion. This reduction was described according to the international IKDC classification in 3 stages, grade 1 = glide guard 2: clunk, and grade 3: gross instability, and the assessment was made according to the examiner’s perception of the contralateral healthy side [19]. The technique was standardized in order to be reproducible and to limit the variability of the results [8].

As per the CLIN (committee for the control of nosocomial infections)protocol which is defined by the institution, the injured limb was prepared utilizing the standard antiseptic techniques. Next, the smartphone was placed in a sterile bag and sealed airtight to maintain sterility (Fig. 1).

Fig. 1
figure 1

The smartphone was placed in a sterile bag and sealed airtight to maintain sterility and placed on the Gerdy’s tubercule

The PS test procedure was then repeated, three times during each per-operative stage (ACL reconstruction, and following LET or meniscal repair), and post-operatively, by the same two orthopedic surgeons. The PS test was also evaluated clinically according to the IKDC classification.

The smartphone (Pro Mate 20, Huawei, Shenzhen, China) with the Sensor Kinetics Pro application (INNOVENTIONS Inc., Houston, TX, USA) was attached to Gerdy’s tubercle with adhesive strips. The Y axis of the phone was parallel to the axis of the tibia. This protocol, already detailed in a previous study [29], allows the analysis of anteroposterior translation and acceleration during the reduction of the lateral compartment of the knee in the PS test.

The Sensor Kinetics Pro application was downloaded to the smartphone. The application measured the acceleration of the movement of the tibia relative to the femur along the X, Y and Z axes of the accelerometer quantitatively in m.s2 (Fig. 2) In this study, the Y-axis represented the acceleration of the longitudinal axis of the tibia, the X-axis the acceleration in the anteroposterior plane relative to the longitudinal axis, and the Z-axis the acceleration in the mediolateral plane. An evolving graph of the 3 axes was displayed directly on the screen during all the acquisition and the different manipulations (Fig. 3). The measurements were recorded in an .acc graphics files and renamed by a number to guarantee the anonymity of the patient. This data allowed us to analyze the changes, induced by the surgical procedure, on the knee kinetics and instability.

Fig. 2
figure 2

Representation of the different axes of the smartphone accelerometer in space

Fig. 3
figure 3

Evolution of acceleration during the PS test on a pathological knee with an anterior cruciate rupture. The Y axis (green) showed the flexion-extension movement induced by the maneuver. The subluxation and abrupt reduction of the anterior compartment of the tibia was highlighted by the abrupt deceleration with point A corresponding to the maximum of the subluxation and point B the reduction. The difference between point A and point B was called the alpha value

Statistical analysis

According to Kocher et al. [11], a change in acceleration of 1.6 m.s2 was significant (ruptured ACL 4.3 +/− 1.2 m.s2, healthy knee 2.7 +/− 0.7 m.s2). According to the original study [29], the Y-axis showed statistically the best inter- and intra-observer reproducibility, and according to the authors, best represented the direction of tibial acceleration during the PS.

The analyses were performed using Python (3.10.4) from the Company software. Intra- and inter-observer reliability was evaluated via intra-class correlation coefficient (ICC) analysis, in absolute agreement two-way mixed-effect model with single fixed rates. Reliability was defined as poor (ICC < 0.5), moderate (0.5 < ICC < 0.75), good (0.75 < ICC < 0.90) and excellent (ICC < 0.90). The standard error of measurements (SEM) were also reported. Abnormal distribution of continuous variables was verified using the Shapiro Wilk test. Data was expressed as means ± standard deviation (SD). The Spearman correlation was used to correlate the anterior tibial translation with the IKDC ranking grade. One-way analysis of variance was performed to determine whether the range of acceleration was significantly different among the clinical grading categories (0, 1, 2, or 3).

Results

After clinical examination of 12 knees were finally included in this study,

The mean measurements (of both examiners) of PS test during the different steps of the surgery are summarized in the Table 1. The intra-observer reliability was good for the first examiner and excellent for the second examiner, with the ICC 0.89 [0.67, 0.98] p < 0,001 and ICC 0.97 [0.91, 1.0] p < 0,001 respectively.

Table 1 Average measurements (of both examiners) and standard deviation of pivot shift during the different stages of intervention

The inter-observer reliability was excellent between the two observers with the ICC 0.99 [0.97, 1.0] p < 0,001.

The mean value of the tibial acceleration during PS test for the ACL-injured knees was 3.45 m.s2 (SD = 1.71), which was significantly higher than that of normal knees (1.03 m.s2; SD = 0.36, p < 0.001).

At the end of the 12 interventions performed, all PS tests were graded 0 in accordance with the IKDC classification.

The correlation between the IKDC grade of PS and the range of value of tibial acceleration was 0,927, p 10− 5. A forest plot was shown in Fig. 4.

Fig. 4
figure 4

Forest plot of the mean values of tibial acceleration in Y axis according to the IKDC grade of PS in X axis

Discussion

The main finding of the present study is that the use of a smartphone accelerometer intra-operatively to quantitatively characterize the subjective PS during ACL reconstruction surgery is validated with excellent interclass correlation coefficients between the two surgeons and good or excellent intraclass correlation. The results of the present study showed an evolution of the quantitative assessment of the PS test performed intra-operatively. The results presented in this study also considered the involvement of the meniscoligamentous injuries in the magnitude of the PS test. No single stage of the procedure demonstrated a significant reduction in the acceleration of the tibia on the femur between meniscal repair, ACL reconstruction, ramp repair and lateral tenodesis. We assumed that the population size was too small to show a statistical significance and that a larger trial would be required to answer this question.

Numerous studies have looked at the relationship between PS and the patient’s meniscal status. The study by Jacquet et al. [10] presented risk factors for PS during clinical follow-up and highlighted the protective effect of meniscal repair, the deleterious effect of meniscectomy, therapeutic abstention or the appearance of a new meniscal lesion. Moreover, the presence of a high-grade PS was a predictive factor for the development of new meniscal injuries. The study by Mouton et al. [18] showed a greater PS in patients with a ramp lesion. The relationship between lateral tenodesis and PS has been more controversial in the literature. The ACL Consensus group [6] recommended that a high-grade preoperative PS could be an indication for an additional anterolateral procedure in the setting of ACL reconstruction. In contrast, Jacquet et al . [10] found that the addition of a lateral tenodesis at the time of the operation did not influence PS. However, in a recent study, Firth et al . [4] showed that the addition of LET reduces the odds of postoperative asymmetric PS by 46%. The results of this study showed that meniscal repairs, especially ramp and external meniscal injury repairs, decreased the PS acceleration and in most cases, lateral tenodesis had a little impact, even though the results were not significant.

This study had several limitations. First of all, the PS test is dependent on the surgeon’s experience, and grading it according to the IKDC classification is subjective. Berruto et al. [2] reported a learning curve, when measuring the tibia acceleration (a 50% specificity at the beginning to 90% specificity at the end of the study) attributed to their device (KiRA), but it could be attributed to the performance of the PS test [20]. Therefore, in this present study, the whole technique was well documented and described [21]. The same technique was taught and performed [8, 19] in order to limit this bias, but could not provide a total emancipation.

Another potential limitation may be related to the choice of the studied parameter. The acceleration of the longitudinal translation of the tibia during the maneuver was considered the key to quantify the extent of the PS, whereas the smartphone’s application could measure multiple other parameters which may contribute to the PS. The acceleration of the longitudinal translation of the tibia seemed to be the most reliable [29], whereas the rotational component lacked precision. Indeed, Vaidya et al. showed an AUC value of 0,98 and a specificity of 100% for the Y-axis with a smartphone accelerometer, better than other commercial accelerometers [2, 8, 22], and similar to electromagnetic devices [27]. A cadaveric study highlighted that the anterior tibial lateral compartment translation of 6-7 mm is necessary for a grade 1 PS, 15 mm for a grade 2, and over 20–25 mm for a grade 3 [1]. The use of this parameter is supported with strong evidence in literature [2, 7, 22, 29]. Berruto et al. [2] found a positive acceleration difference in favor of the pathologic knee among 100 patients and a reliable use of KiRA, another triaxial accelerometer, with mean reference values for every grade of the PS test.

A recent study [22] did not support the use of the triaxial accelerometer alone to assign a numerical value to the pivot-shift phenomenon. At first, parameters other than acceleration range were evaluated and appeared unable to show correlation with clinical examination. Moreover, the direction of the acceleration range was unspecified, although Vaidya et al. [29] showed statistical difference between the longitudinal (Y axis) and the others. However, the PS is still described as a translation corresponding to anterior subluxation of the lateral tibial plateau, which was measured, and a rotation of the tibia relative to the femur [15] which is fully occulted in biomechanical studies. Further studies should be conducted to use the included gyrometer in order to better understand the rotational laxity.

This study also presents some strengths. At first, to the best of our knowledge, no other study provided an intraoperative quantitative assessment of the PS. Most of the devices on the market are either invasive, bulky or expensive, such as electromagnetic devices [13]. The most commonly used and studied system, the KiRA, cannot be used intraoperatively. Smartphones provided a cheaper solution within everyone’s reach. A protocol is proposed to maintain sterility for it use intraoperatively. The use of this tool could help in the decision to perform a tenodesis if the measured rotational instability is not fully corrected by meniscal repairs and ACLR.

Conclusion

The smartphone accelerometer is a reproducible device to quantitatively assess the internal rotation and anterior tibial translation during ACL reconstruction surgery. The measurements are influenced by the different surgical steps. Other larger cohort studies are needed to evaluate the specific impact of each step of the ACL reconstruction and meniscal repair on this measurement. An external validation using other technologies are needed to validate the reliability of this device to assess the PS test.

Availability of data and materials

Yes

References

  1. Bedi A, Musahl V, Lane C, Citak M, Warren RF, Pearle AD (2010) Lateral compartment translation predicts the grade of PS: a cadaveric and clinical analysis. Knee Surg Sports Traumatol Arthrosc 18:1269–1276

    Article  Google Scholar 

  2. Berruto M, Uboldi F, Gala L, Marelli B, Albisetti W (2013) Is triaxial accelerometer reliable in the evaluation and grading of knee pivot-shift phenomenon? Knee Surg Sports Traumatol Arthrosc 21:981–985

    Article  CAS  Google Scholar 

  3. DePhillipo NN, Moatshe G, Brady A, Chahla J, Aman ZS, Dornan GJ, Nakama GY, Engebretsen L, LaPrade RF (2018) Effect of Meniscocapsular and Meniscotibial Lesions in ACL-Deficient and ACL-Reconstructed Knees: A Biomechanical Study. Am J Sports Med 46:2422–2431

    Article  Google Scholar 

  4. Firth AD, Bryant DM, Litchfield R, McCormack RG, Heard M, MacDonald PB, Spalding T, PCM V, Peterson D, Bardana D, Rezansoff A, STABILITY Study Group, AMJ G, Willits K, Birmingham T, Hewison C, Wanlin S, Pinto R, Martindale A, O’Neill L, Jennings M, Daniluk M, Boyer D, Zomar M, Moon K, Moon R, Fan B, Mohan B, Buchko GM, Hiemstra LA, Kerslake S, Tynedal J, Stranges G, Mcrae S, Gullett L, Brown H, Legary A, Longo A, Christian M, Ferguson C, Mohtadi N, Barber R, Chan D, Campbell C, Garven A, Pulsifer K, Mayer M, Simunovic N, Duong A, Robinson D, Levy D, Skelly M, Shanmugaraj A, Howells F, Tough M, Thompson P, Metcalfe A, Asplin L, Dube A, Clarkson L, Brown J, Bolsover A, Bradshaw C, Belgrove L, Milan F, Turner S, Verdugo S, Lowe J, Dunne D, McGowan K, Suddens C-M, Declerq G, Vuylsteke K, Van Haver M (2022) Predictors of Graft Failure in Young Active Patients Undergoing Hamstring Autograft Anterior Cruciate Ligament Reconstruction With or Without a Lateral Extra-articular Tenodesis: The Stability Experience. Am J Sports Med 50:384–395

    Article  Google Scholar 

  5. Galway HR, MacIntosh DL (1980) The lateral PS: a symptom and sign of anterior cruciate ligament insufficiency. Clin Orthop Relat Res 147:45–50

    Article  Google Scholar 

  6. Getgood A, Brown C, Lording T, Amis A, Claes S, Geeslin A, Musahl V, ALC Consensus Group (2019) The anterolateral complex of the knee: results from the International ALC Consensus Group Meeting. Knee Surg Sports Traumatol Arthrosc 27:166–176

    Article  Google Scholar 

  7. Hardy A, Casabianca L, Hardy E, Grimaud O, Meyer A (2017) Combined reconstruction of the anterior cruciate ligament associated with anterolateral tenodesis effectively controls the acceleration of the tibia during the PS. Knee Surg Sports Traumatol Arthrosc 25:1117–1124

    Article  Google Scholar 

  8. Hoshino Y, Araujo P, Ahlden M, Moore CG, Kuroda R, Zaffagnini S, Karlsson J, Fu FH, Musahl V (2012) Standardized PS test improves measurement accuracy. Knee Surg Sports Traumatol Arthrosc 20:732–736

    Article  Google Scholar 

  9. Irrgang JJ, Anderson AF, Boland AL, Harner CD, Kurosaka M, Neyret P, Richmond JC, Shelborne KD (2001) Development and validation of the international knee documentation committee subjective knee form. Am J Sports Med 29:600–613

    Article  CAS  Google Scholar 

  10. Jacquet C, Pioger C, Seil R, Khakha R, Parratte S, Steltzlen C, Argenson J-N, Pujol N, Ollivier M (2021) Incidence and Risk Factors for Residual High-Grade PS After ACL Reconstruction With or Without a Lateral Extra-articular Tenodesis. Orthop J Sports Med 9:232596712110035

    Article  Google Scholar 

  11. Kocher MS, Steadman JR, Briggs KK, Sterett WI, Hawkins RJ (2004) Relationships between objective assessment of ligament stability and subjective assessment of symptoms and function after anterior cruciate ligament reconstruction. Am J Sports Med 32:629–634

    Article  Google Scholar 

  12. Lian J, Novaretti JV, Sheean AJ, Patel NK, Whaley S, Popchak A, Musahl V (2019) Static Lateral Tibial Plateau Subluxation Predicts High-Grade Rotatory Knee Laxity in Anterior Cruciate Ligament-Deficient Knees. Am J Sports Med 47:277–284

    Article  Google Scholar 

  13. Lopomo N, Zaffagnini S, Bignozzi S, Visani A, Marcacci M (2010) Pivot-shift test: analysis and quantification of knee laxity parameters using a navigation system. J Orthop Res 28:164–169

    Article  Google Scholar 

  14. Lopomo N, Zaffagnini S, Signorelli C, Bignozzi S, Giordano G, Marcheggiani Muccioli GM, Visani A (2012) An original clinical methodology for non-invasive assessment of pivot-shift test. Comput Methods Biomech Biomed Engin 15:1323–1328

    Article  Google Scholar 

  15. Magnussen RA, Reinke EK, Huston LJ, MOON Knee Group, Hewett TE, Spindler KP, Amendola A, Andrish JT, Brophy RH, Dunn WR, Flanigan DC, Jones MH, Kaeding CC, Marx RG, Matava MJ, Parker RD, Vidal AF, Wolcott ML, Wolf BR, Wright RW (2018) Effect of High-Grade Preoperative Knee Laxity on 6-Year Anterior Cruciate Ligament Reconstruction Outcomes. Am J Sports Med 46:2865–2872

    Article  Google Scholar 

  16. Mariani PP (2011) Posterior horn instability of the medial meniscus a sign of posterior meniscotibial ligament insufficiency. Knee Surg Sports Traumatol Arthrosc 19:1148–1153

    Article  CAS  Google Scholar 

  17. Matsumoto H (1990) Mechanism of the PS. J Bone Joint Surg Br 72:816–821

    Article  CAS  Google Scholar 

  18. Mouton C, Magosch A, Pape D, Hoffmann A, Nührenbörger C, Seil R (2020) Ramp lesions of the medial meniscus are associated with a higher grade of dynamic rotatory laxity in ACL-injured patients in comparison to patients with an isolated injury. Knee Surg Sports Traumatol Arthrosc 28:1023–1028

    Article  Google Scholar 

  19. Musahl V, Hoshino Y, Ahlden M, Araujo P, Irrgang JJ, Zaffagnini S, Karlsson J, Fu FH (2012) The PS: a global user guide. Knee Surg Sports Traumatol Arthrosc 20:724–731

    Article  Google Scholar 

  20. Naendrup J-H, Patel NK, Zlotnicki JP, Murphy CI, Debski RE, Musahl V (2019) Education and repetition improve success rate and quantitative measures of the PS test. Knee Surg Sports Traumatol Arthrosc 27:3418–3425

    Article  Google Scholar 

  21. Naendrup J-H, Zlotnicki JP, Murphy CI, Patel NK, Debski RE, Musahl V (2019) Influence of knee position and examiner-induced motion on the kinematics of the PS. J Exp Orthop 6:11

    Article  Google Scholar 

  22. Napier RJ, Feller JA, Devitt BM, McClelland JA, Webster KE, Thrush CSJ, Whitehead TS (2021) Is the KiRA Device Useful in Quantifying the PS in Anterior Cruciate Ligament-Deficient Knees? Orthop J Sports Med 9:2325967120977869

    Article  Google Scholar 

  23. Ostrowski JA (2006) Accuracy of 3 diagnostic tests for anterior cruciate ligament tears. J Athl Train 41:120–121

    Google Scholar 

  24. Saithna A, Thaunat M, Delaloye JR, Ouanezar H, Fayard JM, Sonnery-Cottet B (2018) Combined ACL and Anterolateral Ligament Reconstruction. JBJS Essent Surg Tech 8:e2

    Article  Google Scholar 

  25. Song G, Zhang H, Wang Q, Zhang J, Li Y, Feng H (2016) Risk Factors Associated With Grade 3 PS After Acute Anterior Cruciate Ligament Injuries. Am J Sports Med 44:362–369

    Article  Google Scholar 

  26. Song G-Y, Zhang H, Wu G, Zhang J, Liu X, Xue Z, Qian Y, Feng H (2017) Patients with high-grade pivot-shift phenomenon are associated with higher prevalence of anterolateral ligament injury after acute anterior cruciate ligament injuries. Knee Surg Sports Traumatol Arthrosc 25:1111–1116

    Article  Google Scholar 

  27. Tanaka T, Hoshino Y, Miyaji N, Ibaragi K, Nishida K, Nishizawa Y, Araki D, Kanzaki N, Matsushita T, Kuroda R (2018) The diagnostic reliability of the quantitative pivot-shift evaluation using an electromagnetic measurement system for anterior cruciate ligament deficiency was superior to those of the accelerometer and iPad image analysis. Knee Surg Sports Traumatol Arthrosc 26:2835–2840

    Article  Google Scholar 

  28. Thaunat M, Fayard JM, Guimaraes TM, Jan N, Murphy CG, Sonnery-Cottet B (2016) Classification and Surgical Repair of Ramp Lesions of the Medial Meniscus. Arthrosc Tech 5:e871–e875

    Article  Google Scholar 

  29. Vaidya RK, Yoo CW, Lee J, Han H-S, Lee MC, Ro DH (2020) Quantitative assessment of the PS test with smartphone accelerometer. Knee Surg Sports Traumatol Arthrosc 28:2494–2501

    Article  Google Scholar 

  30. Vaudreuil NJ, Rothrauff BB, de Sa D, Musahl V (2019) The PS: current experimental methodology and clinical utility for anterior cruciate ligament rupture and associated injury. Curr Rev Musculoskelet Med 12:41–49

    Article  Google Scholar 

Download references

Funding

No Funding was needed for this study.

Author information

Authors and Affiliations

Authors

Contributions

CJ, GS have made substantial contributions to the conception and design of this report. All authors have made substantial contributions to the acquisition of data, or analysis and interpretation of data and have been involved in drafting the manuscript or revising it critically. All authors have given final approval of the version to be published and agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

Corresponding author

Correspondence to Christophe Jacquet.

Ethics declarations

Ethics approval consent for publication

All patients provided written consent, and the study was approved by our hospital’s ethics committee. (CPP N°2020-A00910033).

Competing interests

The authors declare that they have no competing interests.

Only MO had given paid conferences for Stryker. All other authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Soudé, G., De Villeneuve Bargemon, JB., Khakha, R. et al. Pivot shift intraoperative quantitative assessment using a smartphone accelerometer in ACL deficient knees. J EXP ORTOP 10, 6 (2023). https://doi.org/10.1186/s40634-023-00570-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40634-023-00570-7

Keywords