Is the femoral intramedullary alignment already actual in total knee arthroplasty?

Abstract

Clinical outcomes and overall alignment after total knee arthroplasty (TKA) depend on femoral component positioning in the sagittal and the coronal plane, making choice of the distal femoral cutting guide crucial. Currently, there is no consensus on the potential advantage of an extramedullary (EM) guide compared to an intramedullary (IM) guide in TKA. The IM guide is the most widely used system for making the distal femoral cut although evidence for its superiority over the EM guide is lacking. However, inaccuracies arising with the IM guide include location of the rod entry point in the coronal plane, femoral canal diameter, femoral bowing, and structural features of the rod. Furthermore, the invasive procedure is associated with increased risk of postoperative blood loss, thromboembolic complications, and intraoperative fractures. While the EM guide has no such difficulties, its accuracy depends on the instruments used. Studies have reported results not inferior to the IM guide and a lower number of postoperative complications. Patient-specific instrumentation (PSI) and robotic and computer-assisted TKA have achieved excellent clinical and radiographic results and can overcome the problems inherent to the IM and the EM guide. Authors performed a systematic review of the literature and proposed a narrative review to summarize the characteristics of the IM and the EM guide and compare the advantages and disadvantages of each, as well as their limitations in comparison with new technologies. Authors also expressed their expert opinion.

Introduction

Two types of alignment guides for achieving accurate bone cuts are the intramedullary (IM) and the extramedullary (EM) guide. The femoral IM guide is widely applied in TKA owing to its ease of use. While studies have shown that a more accurate femoral bone cut may be obtained with the IM guide [7, 10, 11, 14], these studies are dated with limitations regarding inadequate technique and preoperative radiographic templating [1, 28, 59] and difficult-to-identify intraoperative landmarks with the EM guide [7, 11, 60]. Current studies have shown that IM guides don’t always ensure the expected component placement [33, 34]. Many factors can affect IM alignment accuracy, including location of the rod entry point in the coronal plane, femoral canal diameter, femoral bowing, and structural features of the rod [5, 16, 41, 52]. In addition, the invasive procedure is associated with increased risk of postoperative blood loss, thromboembolic complications, and intraoperative fractures [20, 21, 43]. Finally, patient-specific instruments (PSI) and robotic-assisted systems that provide highly accurate alignment are coming into wider use [35] though still not feasible in many settings [49]. For this narrative review we summarize recent evidence on two alignment systems for distal femoral bone cuts in TKA and compare the IM and EM guide with the most recent technologies. Despite the narrative nature of this review, the authors performed a systematic review of the literature. Two independent reviewers (P.Z. and A.T.) searched PubMed, EMBASE, and the Cochrane Library up to May 31, 2022. The electronic search strategy used was alignment AND (intramedullary OR extramedullary guide). In cases of disagreement, a third reviewer (P.A) made the final decision. The inclusion criteria were full-text basic science studies that compared the intramedullary with the extramedullary alignment and studies that analyze the main characteristics of one or both methods . Humans and cadaveric studies were included. Studies not written in English were not considered.

Intramedullary guide

The IM guide is the most widely used system for distal femoral cut in TKA despite lack of clear evidence for its superiority over an EM guide. The entry point for the femoral IM rod is a crucial factor for femoral component positioning and restoring the mechanical axis in TKA since it affects the extension gap and limb alignment in both the coronal (varus/valgus) and the sagittal (flexion/extension) plane [30, 48]. Even minor deviations in the insertion point can result in misalignment by several degrees [24, 46]. Errors can result from non-ideal entry point location, improper IM rod length, rod diameter, and rotation of the IM rod [46]. Identification of the ideal femoral entry point has been elusive probably due to high inter-individual anatomical variability, making it difficult to determine an entry point that fits all patients. Reed and Gollish [52] reported that the ideal entry point should be about 6.6 mm medial to the center of the notch, on average. Novotny et al. [46] suggested that the optimal entry point for the IM rod should be determined as a ratio. Wangroongsub and Cherdtaweesup [62] concluded that the entry point should be 1.5 ±2.01 mm medial and 12 ±2.72 mm superior to the top of the femoral intercondylar notch. In their 3D analysis of the femoral entry point, Jianlin Xiao et al. [64] argued that using a point 2.94 ±1.12 mm (range 0.79–4.91 mm) medial and 6.01 ±2.09 mm (range 2.49–9.51 mm) anterior to the deepest point of the intercondylar notch yields an error margin of less than 2°. Errors can also arise from incorrect intramedullary rod thickness and length: a rod that is too long can trigger insertion errors, leading to femoral bowing or other deformities of the distal femur, while a rod with a small radius may not achieve adequate fit within the canal. Nufio-Siebrecht [47] underscored that error can be minimized by increasing IM rod diameter and length. Novotny [46] recommended use of a rod 9 mm in diameter and 228.6 mm (9 inches ) in length with a long guiding rod to minimize chance of error. They argued that rotation can decrease valgus and impart flexion/extension to the anteroposterior cut. The effect for small degrees of malrotation is negligible: for each 10° of malrotation the effect is ~1° in the sagittal plane and ~0.3° in the frontal plane [46]. G. Maderbacher et al. [33] simulated by means of computer-aided engineering software the distal femoral cut using intramedullary alignment rods with the rod gradually aligned from 40° of external to 40° of internal rotation. They found that rotational changes of the distal femoral alignment guides affect both the coronal and the sagittal cutting plane. But because femoral bowing is very frequent in Asian populations, caution is warranted when evaluating data obtained from such populations. Furthermore, the negative results of the IM technique obtained in Asian patient samples should not be extended to Caucasian patient samples, in which lower limbs conformation is different. In their level II study, G. Maderbacher et al. [34] used the IM rod in 40 knees for distal femoral cut in TKA and found that only 25% (10/40) of the gamma angle measurements were within a range between 0° and 3° of flexion, which is the theoretical optimal range of flexion of the femoral component. They concluded that the appropriate sagittal femoral component alignment cannot be ensured with the use of IM alignment rods.

Extramedullary guide

Early studies from the late 1980s and early 1990s on the EM guide reported inferior results compared to the IM guide in which the anterosuperior iliac spine was used as an intraoperative reference for the EM guide [6, 7, 11]. More recent studies with modern EM guidance systems have obtained better results. Baldini and Adravanti [1] developed a set of EM instruments calibrated with preoperative templating to locate the mechanical axis and perform the distal femoral cut without violating the femoral canal. They found that the EM technique was more accurate than the IM technique in the sagittal plane. But because their study included only varus osteoarthritic knees, the conclusions may not be extended to all types of deformity. Matsumoto et al. [37] used a similar EM system and reported femoral component coronal alignment within 0° ±3° of the mechanical axis in 98% of patients. In their level I study, Woon-hwa Jung et al. [19] used an EM instrument to measure interfemoral head center distance for distal femoral bone cut in TKA and compared the results with data from a second group in which an IM guide was used. The overall lower limb alignment was within 0 ±3° of the mechanical axis in all EM group patients. The gamma angle was 2.5 ±1.0° in the IM and 2.3 ±1.7° in the EM group, but no statistically significant differences between the EM and the IM group were found in the coronal and the sagittal alignment of the femoral component. The EM guide is free from complications related to the violation of the femoral canal but much of its precision depends on the characteristics of the instruments. Several authors reported non-inferiority of the extramedullary guide compared to the intramedullary guide with a lower complication rate. The EM guide offers an excellent alternative when use of an IM rod is difficult or impossible. Studies [1, 4] have suggested using the EM guide for bilateral, simultaneous TKA, major femoral extraarticular deformities, ipsilateral long-stemmed hip arthroplasty, and when hardware (distal femoral plates and screws, IM nails) is present.

Intramedullary versus extramedullary guides

There is no consensus on whether the EM femoral guide is superior to the IM guide in TKA. One of the most invasive parts of TKA is violation of the IM femoral canal after use of IM instruments. Application of an IM guide for the femur can produce a variety of complications including blood loss, postoperative hypoxia, intraoperative fractures, and fat embolism [21, 27, 29]. In contrast, the EM method is associated with less morbidity and blood loss because the bone marrow is not invaded. However, identifying the anterosuperior iliac spine as an intraoperative landmark to locate the femoral head center for the EM guide is difficult because of the large soft tissue cover, tourniquets, and the anatomical variability of the proximal femur. In their metanalysis, Q. Tang et al. [58] found no statistically significant difference between the EM- and the IM-guided femoral cut for lower limb coronal alignment, coronal alignment of the femoral component, sagittal alignment of the femoral component, and operating time, but blood loss was less with use of the EM guide. Baldini and Adravanti [1] found that femoral component sagittal alignment within 0 ±2° (gamma angle) was achieved more often in the EM than in the IM group (90% versus 78% respectively). In their study investigating the influence of the IM and the EM guide on distal femoral cut in 6726 TKAs, Meding et al. [40] found no statistical difference in postoperative clinical scores or in survivorship at 15 years of follow-up; however, mean tibiofemoral (overall) anatomical alignment was statistically more accurate in the IM group.

General concept of alignment

Proper prosthetic placement and overall limb alignment are crucial for successful TKA implant survivorship and prevention of early failure [12, 53, 56]. Among different types of alignment described in the literature [25], there is broad consensus that coronal and sagittal component mechanical alignment should be within ±3° of neutral alignment to best ensure implant survival [12, 34, 53, 54, 56]. However, not all studies supported a neutral mechanical axis as gold standard [32]. Parrate [50] performed a retrospective study on 398 TKAs, he considered as outliers 106 knees with post-operative mechanical alignment outside 0 ° ± 3 ° and found no difference in survivorship beetween 106 outliers and 292 weel-aligned knees at 15 years. Other authors [39, 61] reported a non-inferiority of clinical results if the pre-operative deformity is preserved or partially preserved, without however reaching the range -3 / +3°. In addition, the current expansion of the kinematic alignment, and the excellent outcomes reported so far, further cast doubt on the need to achieve an alignment within -3 / + 3 ° from the mechanical plane as a gold standard. Consistent with the new concepts of alignment on the coronal and the sagittal plane, use of an EM guide correlates much closer with the concepts of knee kinematics and better individualized alignment. When using the IM guide on the coronal plane, the degree of valgus is limited to what that the guide shows (generally 0 to 9°). Differently, taking into account condylar wear and observing the rule of re-establishing the knee’s pre-arthritic condition, the EM guide overcomes these limitations and constraints. In addition, the sagittal plane maintains its anatomy by leaning on the anterior cortex and thus follows femur anatomy.

Practicing a robotic assisted knee prosthetic replacement allows to bypass the problem of femoral alignment on the coronal plane, therefore of the use of a intra or extra medullary guide, since the distal femoral cut is carried out according to a preoperative computed tomography scan of the knee and consequently on a three-dimensional planning. There is no need to violate the medullary canal or to use further anatomical landmarks. However, the robotic assisted method remains highly expensive and not accessible to all surgeons. This makes the dilemma of which technique is best to optimize the distal femoral cut still current.

Emily L. Hampp, et al. [15] compared the precision and accuracy of bone cuts and positioning of components on 6 cadavers (12 knees), six knees performed with standard technique and 6 knees with robot assisted technique, demonstrating greater accuracy and precision of the robot-assisted technique compared to the manual one, especially for the management of the varus-valgus alignment on the coronal plane. The error on the distal femoral cut showed a median of 0.5 ° and 2.6 ° respectively for the assisted and manual robotic cut, without a substantial difference in the management of flexion-extension.

James D. Sires et al. [57] performed 45 TKA knee arthroplasty and analysed bone cuts, made using the MAKO Total Knee system (Stryker) and saw that for the distal femoral cut the maximum error was 1.20mm, 1.00mm, 1.80mm with an average error of 0.32mm , 0.28mm, 0.46mm for depth, valgus and flexion respectively. This precision allowed them to obtain an overall alignment within 3 ° of the preoperative planning in 100% of cases. KA also depends on the anatomical axis and therefore on the correct positioning of the intramedullary landmark. The advantage of the kinematic technique lies in being able to predict the thickness of the resections, the measurement of the thickness of the bone cuts allowing to perform an important check and consequently to carry out the necessary corrections or recut. This technique allows the patient's knee to be brought back to a pre-osteoarthritis condition without affecting the soft tissues. It is known that KA reduces the recovery time, improves functional outcomes, reduces residual symptoms after TKA, increases patient satisfaction with good survivorship at a short term follow-up. Howell, Shelton and Hull [18] have done a 10-year follow-up of 191 cases with a survival rate of 98.8%.

The robot-assisted technique allows to pursue a kinematic or mechanical alignment, being able to refer with extreme precision to the anatomical axes. In a recent systematic review and meta-analysis [26] the robot-assisted knee arthroplasty showed more accurate component positioning and placement within target zones, however, this procedure is more expensive and requires more operating time and there is no clear superiority over the conventional technique in terms of clinical scores, satisfaction and implant survivorship. The literature is limited with regard to the accuracy of bone cuts with robotic system.

In these terms we can say that the robotic technique is highly accurate and can satisfy both kinematic and mechanical alignment with extreme precision. However, it is not yet clear whether this advantage can justify the increase in economic expenditure.

New trends

Patient-specific instrumentation (PSI), robotic and computer-assisted TKA were introduced with the aim to improve alignment without violating the femoral canal. PSI entails the use of customized cutting blocks generated from a pre-operative three-dimensional model created using computed tomography (CT) or magnetic resonance imaging (MRI). The aim is to facilitate prosthesis implant by analysis of 3-D preoperative planning of TKA and simulation of the bone cuts. During this phase coronal orientation and depth of bone resection are planned. Customized cutting guides that fit on the patient’s anatomy are then manufactured. However, there is no consensus regarding the accuracy and the reliability of PSI and many studies have reported controversial results. It cannot be considered a gold standard in TKA, and so cannot be recommended for routine use. No substantial improvement in component alignment, surgical time, blood loss or functional outcomes has been reported. Mattei et al. [38] suggested the use of PSI in particular situations such as post-traumatic osteoarthritis with severe deformities (e.g., femoral or tibial fractures healed with malalignment), in which preoperative planning may be difficult and use of an IM rod difficult or impossible. Computer-assisted surgery (CAS) for TKA seems to provide more precise component placement in coronal, sagittal and rotational alignment, more accurate bone cuts, and better restoration of coronal limb alignment [55, 56]. Several meta-analysis studies [13, 17, 36, 51] found that CAS improves mechanical axis and implant survivorship but no clinical evidence of improvement in functional outcome. In a meta-analysis of 29 studies comparing CAS to conventional technique, Mason et al. [36] reported that femoral varus/valgus alignment was within 2° of the femoral mechanical axis in 90.4% of the CAS group patients versus 65.9% in the conventional group. Nam et al. [44] used a portable accelerometer-based surgical navigation system to record the hip center of rotation and femoral mechanical axis for the distal femoral cut in 48 patients undergoing TKA. They reported that 95.8% of the femoral components were placed within 90° ±2° of the femoral mechanical axis in the coronal plane and that overall lower extremity alignment was within 3° of neutral to the mechanical axis in 93.8% of cases. In their retrospective study, Biazzo A. et al. [3] compared the radiological results obtained from three groups of patients undergoing TKA via three different methods (n=97 patients available for radiological assessment). At a medium follow-up of 15 years, there was no statistically difference in the mean hip-knee-ankle (HKA) angle and the mean frontal femoral component angle (FFC) between the patients who underwent TKA via a computer-based alignment system and those in which a total IM alignment system and a total EM alignment system were used. Importantly, however, they did not consider the sagittal plane. In their systematic overview of meta-analyses, Kort et al. [26] reported that robot-assisted knee arthroplasty allows for more accurate component positioning and placement within target zones, but the results were inconclusive about whether it improves clinical scores, patient satisfaction or reduces complications and revision rates. In their review of clinical and radiological results of the MAKO CT-based robotic-assisted system for TKA, Batailler et al. [2] found lower postoperative pain scores, shorter time to hospital discharge, better functional scores, and better implant positioning with CT-based robotic-assisted TKA than with conventional TKA. Other studies reported similar functional results for both surgical techniques at 6 months and 1 year of follow up [23, 45]. Robotic-assisted TKA seems to be less invasive than conventional TKA, thus allowing better soft tissue protection and exposure, without robot-related complications. Several studies reported an increase in infection rate probably due to longer operating times at the beginning of the learning curve and fracture at the pin insertion site or pin breakage [31, 63]. Kayani et al. [22] found that seven robotic cases are necessary to gain an improvement in the operating time of robotic arm-assisted TKA and of surgical team stress levels without any learning curve effect on accuracy of implant position and limb alignment. Cool et al., Cotter et al., and Mont et al. reported that 90-day episode-of-care (EOC) costs were lower after robotic-assisted TKA [8, 9, 42].

Conclusions

There is no clear superiority of one guide over another (IM vs EM). The choice of either alignment system should be determined by the patient's anatomy and the surgeon’s confidence and experience. Orthopedic societies should devote more effort to the development of more precise EM guides. In our opinion, surgeons not familiar with navigation systems or robotic-assisted TKA should consider using the EM guide for the distal femoral cut. This technique has been proven safe, fast, and accurate for reproducing a cut in both the coronal and the sagittal planes. Moreover, overall results are not inferior to the IM guide with fewer post-operative side effects. Use of robot-assisted systems and navigation obviate the distinction between the EM and the IM guide. Numerous studies have reported better results without added complications and with lower 90-day EOC costs.

Abbreviations

TKA:

Total knee arthroplasty

EM:

Extramedullary

IM:

Intramedullary

PSI:

Patient-specific instrumentation

CT:

Computed tomography

MRI:

Magnetic resonance imaging

References

  1. Baldini A, Adravanti P (2008) Less invasive TKA: extramedullary femoral reference without navigation. Clin Orthop Relat Res 466:2694–2700. https://doi.org/10.1007/s11999-008-0435-9

    Article  PubMed  PubMed Central  Google Scholar 

  2. Batailler C, Fernandez A, Swan J et al. (2021) MAKO CT-based robotic arm-assisted system is a reliable procedure for total knee arthroplasty: a systematic review. Knee Surg Sports Traumatol Arthrosc 29:3585–3598. https://doi.org/10.1007/s00167-020-06283-z

    Article  PubMed  Google Scholar 

  3. Biazzo A, Manzotti A, Confalonieri N (2019) Computer-assisted versus intramedullary and extramedullary alignment system in total knee replacement: Long term follow-up. J Clin Orthopaedic Trauma 10:555–559. https://doi.org/10.1016/j.jcot.2018.07.005

    Article  Google Scholar 

  4. Bonanzinga T, Tanzi P, Pia Neri M et al. (2018) Evaluation of blood loss and implant alignment after total knee arthroplasty with inertial based extramedullary femoral cutting guide. Joints 06:161–166. https://doi.org/10.1055/s-0038-1673404

    Article  Google Scholar 

  5. Bono OJ, Olcott CW, Carangelo R et al. (2020) Femoral intramedullary alignment in total knee arthroplasty: indications, results, pitfalls, alternatives, and controversies. J Knee Surg 33:012–014. https://doi.org/10.1055/s-0038-1676070

    Article  Google Scholar 

  6. Brys DA, Lombardi AV, Mallory TH, Vaughn BK (1991) A comparison of intramedullary and extramedullary alignment systems for tibial component placement in total knee arthroplasty. Clin Orthop Relat Res:175–179

  7. Cates HE, Ritter MA, Keating EM, Faris PM (1993) Intramedullary versus extramedullary femoral alignment systems in total knee replacement. Clin Orthop Relat Res:32–39

  8. Cool CL, Jacofsky DJ, Seeger KA et al. (2019) A 90-day episode-of-care cost analysis of robotic-arm assisted total knee arthroplasty. J Comparative Effective Res 8:327–336. https://doi.org/10.2217/cer-2018-0136

    Article  Google Scholar 

  9. Cotter EJ, Wang J, Illgen RL (2022) Comparative cost analysis of robotic-assisted and jig-based manual primary total knee arthroplasty. J Knee Surg 35:176–184. https://doi.org/10.1055/s-0040-1713895

    Article  PubMed  Google Scholar 

  10. Dennis DA, Channer M, Susman MH, Stringer EA (1993) Intramedullary versus extramedullary tibial alignment systems in total knee arthroplasty. J Arthroplasty 8:43–47. https://doi.org/10.1016/S0883-5403(06)80106-3

    Article  CAS  PubMed  Google Scholar 

  11. Engh GA, Petersen TL (1990) Comparative experience with intramedullary and extramedullary alignment in total knee arthroplasty. J Arthroplasty 5:1–8. https://doi.org/10.1016/S0883-5403(06)80002-1

    Article  CAS  PubMed  Google Scholar 

  12. Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty. J Arthroplasty 24:39–43. https://doi.org/10.1016/j.arth.2009.04.034

    Article  PubMed  Google Scholar 

  13. Fu Y, Wang M, Liu Y, Fu Q (2012) Alignment outcomes in navigated total knee arthroplasty: a meta-analysis. Knee Surg Sports Traumatol Arthrosc 20:1075–1082. https://doi.org/10.1007/s00167-011-1695-6

    Article  PubMed  Google Scholar 

  14. Gromov K, Korchi M, Thomsen MG et al. (2014) What is the optimal alignment of the tibial and femoral components in knee arthroplasty?: An overview of the literature. Acta Orthop 85:480–487. https://doi.org/10.3109/17453674.2014.940573

    Article  PubMed  PubMed Central  Google Scholar 

  15. Hampp E, Chughtai M, Scholl L et al. (2019) Robotic-arm assisted total knee arthroplasty demonstrated greater accuracy and precision to plan compared with manual techniques. J Knee Surg 32:239–250. https://doi.org/10.1055/s-0038-1641729

    Article  PubMed  Google Scholar 

  16. Haruta Y, Kawahara S, Tsuchimochi K et al. (2018) Deviation of femoral intramedullary alignment rod influences coronal and sagittal alignment during total knee arthroplasty. Knee 25:644–649. https://doi.org/10.1016/j.knee.2018.04.011

    Article  PubMed  Google Scholar 

  17. Hetaimish BM, Khan MM, Simunovic N et al. (2012) Meta-analysis of navigation vs conventional total knee arthroplasty. J Arthroplasty 27:1177–1182. https://doi.org/10.1016/j.arth.2011.12.028

    Article  PubMed  Google Scholar 

  18. Howell SM, Shelton TJ, Hull ML (2018) Implant survival and function ten years after kinematically aligned total knee arthroplasty. J Arthroplasty 33:3678–3684. https://doi.org/10.1016/j.arth.2018.07.020

    Article  PubMed  Google Scholar 

  19. Jung W, Chun C, Lee J et al. (2013) The accuracy of the extramedullary and intramedullary femoral alignment system in total knee arthroplasty for varus osteoarthritic knee. Knee Surg Sports Traumatol Arthrosc 21:629–635. https://doi.org/10.1007/s00167-012-1994-6

    Article  PubMed  Google Scholar 

  20. Kalairajah Y, Simpson D, Cossey AJ et al. (2005) Blood loss after total knee replacement: effects of computer-assisted surgery. J Bone Joint Surger Brit 87-B:1480–1482. https://doi.org/10.1302/0301-620X.87B11.16474

    Article  Google Scholar 

  21. Kandel L, Vasili C, Kirsh G (2010) Extramedullary femoral alignment instrumentation reduces blood loss after uncemented total knee arthroplasty. J Knee Surg 19:256–258. https://doi.org/10.1055/s-0030-1248115

    Article  Google Scholar 

  22. Kayani B, Konan S, Huq SS et al. (2019) Robotic-arm assisted total knee arthroplasty has a learning curve of seven cases for integration into the surgical workflow but no learning curve effect for accuracy of implant positioning. Knee Surg Sports Traumatol Arthrosc 27:1132–1141. https://doi.org/10.1007/s00167-018-5138-5

    Article  PubMed  Google Scholar 

  23. Khlopas A, Sodhi N, Hozack WJ et al. (2020) Patient-reported functional and satisfaction outcomes after robotic-arm-assisted total knee arthroplasty: early results of a prospective multicenter investigation. J Knee Surg 33:685–690. https://doi.org/10.1055/s-0039-1684014

    Article  PubMed  Google Scholar 

  24. Kim J-G, Kasat NS, Bae J-H et al. (2012) The radiological parameters correlated with the alignment of the femoral component after Oxford Phase 3 unicompartmental knee replacement. J Bone Joint Surger Brit 94-B:1499–1505. https://doi.org/10.1302/0301-620X.94B11.29217

    Article  Google Scholar 

  25. Kim Y-H, Park J-W, Kim J-S, Park S-D (2014) The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesis. Int Orthopaedic (SICOT) 38:379–385. https://doi.org/10.1007/s00264-013-2097-9

    Article  Google Scholar 

  26. Kort N, Stirling P, Pilot P, Müller JH (2022) Robot-assisted knee arthroplasty improves component positioning and alignment, but results are inconclusive on whether it improves clinical scores or reduces complications and revisions: a systematic overview of meta-analyses. Knee Surg Sports Traumatol Arthrosc 30:2639–2653. https://doi.org/10.1007/s00167-021-06472-4

    Article  PubMed  Google Scholar 

  27. Kumar N, Saleh J, Gardiner E et al. (2000) Plugging the intramedullary canal of the femur in total knee arthroplasty. J Arthroplasty 15:947–949. https://doi.org/10.1054/arth.2000.8592

    Article  CAS  PubMed  Google Scholar 

  28. Laskin RS (1984) Alignment of total knee components. Orthopedics 7:62–72. https://doi.org/10.3928/0147-7447-19840101-09

    Article  CAS  PubMed  Google Scholar 

  29. Laskin RS (2005) Minimally invasive total knee arthroplasty: the results justify its use. Clin Orthop Relat Res 440:54–59. https://doi.org/10.1097/01.blo.0000186562.08685.a2

    Article  PubMed  Google Scholar 

  30. Lombardi AV, Berend KR, Ng VY (2011) Neutral mechanical alignment: a requirement for successful TKA: affirms. Orthopedics 34. https://doi.org/10.3928/01477447-20110714-40

  31. Lonner JH, Kerr GJ (2019) Low rate of iatrogenic complications during unicompartmental knee arthroplasty with two semiautonomous robotic systems. Knee 26:745–749. https://doi.org/10.1016/j.knee.2019.02.005

    Article  PubMed  Google Scholar 

  32. Lording T, Lustig S, Neyret P (2016) Coronal alignment after total knee arthroplasty. EFORT Open Rev 1:12–17. https://doi.org/10.1302/2058-5241.1.000002

    Article  PubMed  PubMed Central  Google Scholar 

  33. Maderbacher G, Matussek J, Keshmiri A et al. (2018) Rotation of intramedullary alignment rods affects distal femoral cutting plane in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 26:3311–3316. https://doi.org/10.1007/s00167-018-4875-9

    Article  PubMed  Google Scholar 

  34. Maderbacher G, Schaumburger J, Baier C et al. (2016) Appropriate sagittal femoral component alignment cannot be ensured by intramedullary alignment rods. Knee Surg Sports Traumatol Arthrosc 24:2453–2460. https://doi.org/10.1007/s00167-015-3541-8

    Article  PubMed  Google Scholar 

  35. Mahoney O, Kinsey T, Sodhi N et al. (2022) Improved component placement accuracy with robotic-arm assisted total knee arthroplasty. J Knee Surg 35:337–344. https://doi.org/10.1055/s-0040-1715571

    Article  PubMed  Google Scholar 

  36. Mason JB, Fehring TK, Estok R et al. (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty 22:1097–1106. https://doi.org/10.1016/j.arth.2007.08.001

    Article  PubMed  Google Scholar 

  37. Matsumoto K, Mori N, Ogawa H, Akiyama H (2015) Accuracy of a novel extramedullary femoral alignment guide system in primary total knee arthroplasty. Arch Orthop Trauma Surg 135:1743–1748. https://doi.org/10.1007/s00402-015-2318-3

    Article  PubMed  Google Scholar 

  38. Mattei L, Pellegrino P, Calò M et al. (2016) Patient specific instrumentation in total knee arthroplasty: a state of the art. Ann Transl Med 4:126–126. https://doi.org/10.21037/atm.2016.03.33

    Article  PubMed  PubMed Central  Google Scholar 

  39. Matziolis G, Adam J, Perka C (2010) Varus malalignment has no influence on clinical outcome in midterm follow-up after total knee replacement. Arch Orthop Trauma Surg 130:1487–1491. https://doi.org/10.1007/s00402-010-1064-9

    Article  PubMed  Google Scholar 

  40. Meding JB, Berend ME, Ritter MA et al. (2011) Intramedullary vs extramedullary femoral alignment guides. J Arthroplasty 26:591–595. https://doi.org/10.1016/j.arth.2010.05.008

    Article  PubMed  Google Scholar 

  41. Mihalko WM, Boyle J, Clark LD, Krackow KA (2005) The variability of intramedullary alignment of the femoral component during total knee arthroplasty. J Arthroplasty 20:25–28. https://doi.org/10.1016/j.arth.2004.10.011

    Article  PubMed  Google Scholar 

  42. Mont MA, Cool C, Gregory D et al. (2021) Health care utilization and payer cost analysis of robotic arm assisted total knee arthroplasty at 30, 60, and 90 days. J Knee Surg 34:328–337. https://doi.org/10.1055/s-0039-1695741

    Article  PubMed  Google Scholar 

  43. Morawa LG, Manley MT, Edidin AA, Reilly DT (1996) Transesophageal echocardiographic monitored events during total knee arthroplasty. Clin Orthop Relat Res 331:192–198. https://doi.org/10.1097/00003086-199610000-00027

    Article  Google Scholar 

  44. Nam D, Nawabi DH, Cross MB et al. (2012) Accelerometer-based computer navigation for performing the distal femoral resection in total knee arthroplasty. J Arthroplasty 27:1717–1722. https://doi.org/10.1016/j.arth.2012.02.007

    Article  PubMed  Google Scholar 

  45. Naziri Q, Cusson BC, Chaudhri M et al. (2019) Making the transition from traditional to robotic-arm assisted TKA: What to expect? A single-surgeon comparative-analysis of the first-40 consecutive cases. J Orthop 16:364–368. https://doi.org/10.1016/j.jor.2019.03.010

    Article  PubMed  PubMed Central  Google Scholar 

  46. Novotny J, Gonzalez MH, Amirouche FML, Li YC (2001) Geometric analysis of potential error in using femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 16:641–647. https://doi.org/10.1054/arth.2001.23714

    Article  CAS  PubMed  Google Scholar 

  47. Nuño-Siebrecht N, Tanzer M, Bobyn JD (2000) Potential errors in axial alignment using intramedullary instrumentation for total knee arthroplasty. J Arthroplasty 15:228–230. https://doi.org/10.1016/S0883-5403(00)90352-8

    Article  PubMed  Google Scholar 

  48. O’Connor MI (2011) Implant survival, knee function, and pain relief after TKA: are there differences between men and women? Clin Orthop Relat Res 469:1846–1851. https://doi.org/10.1007/s11999-011-1782-5

    Article  PubMed  PubMed Central  Google Scholar 

  49. Pailhé R (2021) Total knee arthroplasty: Latest robotics implantation techniques. Orthop Traumatol Surg Res 107:102780. https://doi.org/10.1016/j.otsr.2020.102780

    Article  PubMed  Google Scholar 

  50. Parratte S, Pagnano MW, Trousdale RT, Berry DJ (2010) Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surger Am 92:2143–2149. https://doi.org/10.2106/JBJS.I.01398

    Article  Google Scholar 

  51. Rebal BA, Babatunde OM, Lee JH et al. (2014) Imageless computer navigation in total knee arthroplasty provides superior short term functional outcomes: a meta-analysis. J Arthroplasty 29:938–944. https://doi.org/10.1016/j.arth.2013.09.018

    Article  PubMed  Google Scholar 

  52. Reed SC, Gollish J (1997) The accuracy of femoral intramedullary guides in total knee arthroplasty. J Arthroplasty 12:677–682. https://doi.org/10.1016/S0883-5403(97)90141-8

    Article  CAS  PubMed  Google Scholar 

  53. Ritter MA, Davis KE, Meding JB et al. (2011) The effect of alignment and bmi on failure of total knee replacement. J Bone Joint Surg 93:1588–1596. https://doi.org/10.2106/JBJS.J.00772

    Article  PubMed  Google Scholar 

  54. Rivière C, Iranpour F, Auvinet E et al. (2017) Alignment options for total knee arthroplasty: A systematic review. Orthop Traumatol Surg Res 103:1047–1056. https://doi.org/10.1016/j.otsr.2017.07.010

    Article  PubMed  Google Scholar 

  55. Rosenberger RE, Hoser C, Quirbach S et al. (2008) Improved accuracy of component alignment with the implementation of image-free navigation in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 16:249–257. https://doi.org/10.1007/s00167-007-0420-y

    Article  PubMed  Google Scholar 

  56. Sikorski JM (2008) Alignment in total knee replacement. J Bone Joint Surger Brit 90-B:1121–1127. https://doi.org/10.1302/0301-620X.90B9.20793

    Article  Google Scholar 

  57. Sires JD, Craik JD, Wilson CJ (2021) Accuracy of bone resection in MAKO total knee robotic-assisted surgery. J Knee Surg 34:745–748. https://doi.org/10.1055/s-0039-1700570

    Article  PubMed  Google Scholar 

  58. Tang Q, Shang P, Zheng G et al. (2017) Extramedullary versus intramedullary femoral alignment technique in total knee arthroplasty: a meta-analysis of randomized controlled trials. J Orthop Surg Res 12:82. https://doi.org/10.1186/s13018-017-0582-3

    Article  PubMed  PubMed Central  Google Scholar 

  59. Tanzer M, Makhdom AM (2016) Preoperative planning in primary total knee arthroplasty. J Am Acad Orthopaedic Surg 24:220–230. https://doi.org/10.5435/JAAOS-D-14-00332

    Article  Google Scholar 

  60. Teter KE, Bregman D, Colwell CW (1995) The efficacy of intramedullary femoral alignment in total knee replacement. Clin Orthop Relat Res 321:117–121

    Google Scholar 

  61. Vanlommel L, Vanlommel J, Claes S, Bellemans J (2013) Slight undercorrection following total knee arthroplasty results in superior clinical outcomes in varus knees. Knee Surg Sports Traumatol Arthrosc 21:2325–2330. https://doi.org/10.1007/s00167-013-2481-4

    Article  PubMed  Google Scholar 

  62. Wangroongsub Y, Cherdtaweesup S (2009) Proper entry point for femoral intramedullary guide in total knee arthroplasty. J Med Assoc Thai 92(Suppl 6):S1–S5

    PubMed  Google Scholar 

  63. Wysocki RW, Sheinkop MB, Virkus WW, Della Valle CJ (2008) Femoral fracture through a previous pin site after computer-assisted total knee arthroplasty. J Arthroplasty 23:462–465. https://doi.org/10.1016/j.arth.2007.03.019

    Article  PubMed  Google Scholar 

  64. Xiao J, Wang C, Zhu L, et al. (2014) Improved method for planning intramedullary guiding rod entry point in total knee arthroplasty. Arch Orthop Trauma Surg 134:693–698. https://doi.org/10.1007/s00402-014-1943-6

Download references

Acknowledgements

NA.

Informed consent

NA.

Funding

No funding was requested for this study.

Author information

Authors and Affiliations

Authors

Contributions

AT and PA designed the protocol and paragraphs. AG and PZ collected the data. AG and PZ wrote the initial draft and edited the draft manuscript. AT and PA corrected the draft manuscript versions. All authors approved the submitted version.

Corresponding author

Correspondence to Andrea Tecame.

Ethics declarations

Ethics approval and consent to participate

Ethical approval was not required, as this is a narrative review (expert opinion) of the literature and does not involve humans or animal subjects.

Competing interests

AT, AG, PZ, and PA have nothing to disclose.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Tecame, A., Gambineri, A., Za, P. et al. Is the femoral intramedullary alignment already actual in total knee arthroplasty?. J EXP ORTOP 10, 16 (2023). https://doi.org/10.1186/s40634-022-00563-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40634-022-00563-y

Keywords