CT based PSI blocks for osteotomies around the knee provide accurate results when intraoperative imaging is used

Abstract

Purpose

Correction osteotomies around the knee are common methods for the treatment of varus or valgus malalignment of the lower extremity. In recent years, patient specific instrumentation (PSI) guides were introduced in order to enhance the accuracy of these procedures. The purpose of this study was to determine the accuracy of CT based PSI guides for correction osteotomies around the knee of low volume osteotomy surgeons and to evaluate if CT based PSI blocks deliver a high degree of accuracy without using intraoperative fluoroscopy.

Methods

Two study arms with CT based PSI cutting blocks for osteotomies around the knee were conducted. Part one: A retrospective analysis of 19 osteotomies was made in order to evaluate the accuracy in the hands of a low volume surgeon on long-leg radiographs. Part two: A cadaveric study with 8 knees was performed for the purpose of analyzing the accuracy without using intraoperative fluoroscopy on pre- and postoperative CT scans. Hip-Knee-Ankle angle (HKA), lateral distal femoral angle (LDFA) and medial proximal tibial angle (MPTA) were analyzed. The mean absolute delta (∂) between the planned and postoperative parameters were calculated. The accuracy of both study arms were compared.

Results

Part one: The mean MPTA ∂, LDFA ∂ and HKA ∂ was 0.9°, 1.9° and 1.5°, respectively. Part two: The mean MPTA ∂ and LDFA ∂ was 3.5° and 2.2°, respectively. The mean ∂ of MPTA is significantly different between clinical patients with fluoroscopic control and cadaveric specimens without fluoroscopic control (P < 0.001). All surgeries were performed without complications such as a hinge fracture.

Conclusion

The clinical use of PSI guides for osteotomies around the knee in the hands of low volume surgeons is a safe procedure. The PSI guides deliver a reliable accuracy under fluoroscopic control whereas their non-use of intraoperative fluoroscopy leads to a lack of accuracy. The use of fluoroscopic control during PSI guided correction osteotomies is highly recommended.

Level of evidence

IV – Retrospective and experimental Study

Introduction

Correction osteotomies around the knee are common methods for the treatment of varus or valgus malalignment of the lower extremity. The choice of procedure depends on the location of the deformity [1, 17]. The intention is to reduce the one sided overload of the femoral-tibial joint [8]. Over or under correction can lead to persistent pain and a paradoxical joint line as well as biomechanical issues [23]. Various techniques have been described in the literature to achieve the desired correction [12, 19, 28]. Classic manual techniques use a standardized methodology with conventional instruments in order to determine the extent of the axis change [19]. However, those techniques reach their limit especially in beginners or low volume surgeons [18, 30, 32].

In the past decade, computer navigation was introduced to enhance the accuracy of correction osteotomies. An advantage is the real-time control of the alignment as well as the possibility to evaluate the correction more precisely in the sagittal plane. Certain publications focusing on this topic present more precise results for navigation assisted correction osteotomies compared to traditional manual methods [3, 12, 24, 26]. However, navigation assisted correction osteotomies do not show consistently high precision throughout the literature. Furthermore, the extended operating times as well as additional costs are reported drawbacks [30].

An alternative to the surgical procedures mentioned above are patient-specific instruments (PSI). These custom-made cut blocks allow three-dimensional correction of both the tibia and the femur. The majority of the cutting blocks are made using computer tomography (CT)-based 3D models of the patient’s individual anatomy. The CT includes images of the femoral head, the knee and the ankle in order to calculate not only the hip-knee-ankle angle (HKA) but also the medial proximal tibial angle (MPTA), the lateral distal femoral angle (LDFA) and posterior tibial slope (PTS). Based on these parameters, the plate position, the resection plane, the degree of correction and the screw positions are virtually planned preoperatively. Initial proof of concept studies showed an accuracy of less than 1° or 2° on average (range: ± 1°) [9, 21, 31]. In contrast to the use of PSI for total knee arthroplasty [20], their use for correction deformities is based on multiple fluoroscopy shots during the operation in order to secure their correct position.

This study aims to answer the following questions:

  1. 1.

    Part 1: Are CT based PSI guides for correction osteotomies around the knee in the hands of low volume osteotomy surgeons a safe procedure?

  2. 2.

    Part 2: Can CT based PSI blocks deliver a high degree of accuracy without using intraoperative fluoroscopy?

The primary hypothesis is that CT-based PSI guides for correction osteotomies around the knee of low volume osteotomy surgeons provide a high accuracy. The secondary hypothesis is that the high accuracy can be conducted without intraoperative fluoroscopic control.

Material and methods

Part one

Material

Between January 2018 and August 2019 there were 30 osteotomies around the knee performed in 29 patients using PSI guides with the Activmotion plate (Newclip Technics, Haute-Goulaine, France) by one low volume osteotomy surgeon. The surgeon had performed less than 10 procedures per year before 2018. Inclusion criteria for this radiological analysis were the presence of standard antero-posterior (AP), lateral and full weight bearing long leg radiographs. Only patients with OW-HTO, CW-DFO or OW-DFO were included. Patients with derotating osteotomies were excluded.

Surgical technique

OW-HTO, CW-DFO and OW-DFO models were used to virtually position the manufacturers specific plates. Within the 3D planning process, the plate position, the resection plane, the degree of correction as well as the screw positions were virtually planned (Fig. 1). Intraoperatively, the PSI guide was fixed with two or three K-wires under fluoroscopy control in order to control the correct positioning of the PSI guide (Figs. 2 and 3). The oblique k-wire served as a hinge protection. The medial collateral ligament and the pes anserinus are protected with retractors and were not harmed. After confirmation of the planned position, the holes for the plate were pre-drilled. With this step, the plate position was pre-set prior to the osteotomy. Changing the opening or closing angle was therefore no longer possible. In the next step, the osteotomy was performed using the guided slot of the PSI leaving the oblique K-wire in place in order to secure the hinge. Subsequently, the cutting block is divided and the guided biplanar cut is performed (Fig. 3C). The osteotomy was then successively opened respectively closed until the screw holes of the plate fitted with the pre-drilled holes of the bone. In the final step, the plate was fixed. Only the position of the PSI cutting guide and screw length were controlled with fluoroscopy.

Fig. 1
figure 1

Within the 3D planning process, the plate position, the resection plane, the degree of correction, as well as the screw positions were virtually planned

Fig. 2
figure 2

Intraoperatively, the PSI guide was fixed with two K-wires (A + B) under fluoroscopy control (C) in order to control the correct positioning of the PSI guide

Fig. 3
figure 3

Intraoperatively, the PSI guide was fixed with three K-wires (A + B) in order to control the correct positioning of the PSI guide. After performing the osteotomy, the cutting guide is divided and the biplanar cut can be performed (C)

Part 2

Material

In this cadaver study, both medial OW-HTO and lateral OW-DFO were performed on eight long leg cadaveric specimens (four right, four left/ Science Care, Phoenix, Arizona, USA). The BMI of the 6 male and 2 female specimens was lower than 28 kg/m2. The specimens had an age between 58 and 72 years. A CT according to the company’s protocol was performed. Therefore, images of the femoral head, the knee and the ankle were included. For acquisition hip, knee and ankle were scanned with 2.5 mm and for the reconstruction the knee was scanned with 0.625 mm slice thickness. For the OW-HTO an arbitrary 5° valgus correction was planned. For OW-DFO an arbitrary 5° varus correction was planned. The same plate as in part one was used for both the HTO and the DFO.

Surgical technique

Soft tissue was removed from the bone at the attachment points for the PSI guide. There was no further tissue removed or released in order to mimic an in-vivo situation to the highest degree (Fig. 4). The 3D printed cutting blocks were adapted to the individual anatomy of the specimen without fluoroscopy and only with respect to the individual plan. For orientation the distance to the joint gap was measured and compared to the preoperative plan (Figs. 2 and 3). The cutting guide was placed without fluoroscopic control. The rest of the surgery was performed as described in the clinical part (part one) of the study.

Fig. 4
figure 4

3D printed cutting guide fixed with two K-wires. The soft tissue was removed from the bone at the attachment points. No further tissue was removed or released in order to simulate an in-vivo situation to the highest degree

Parameters

All included patients from study arm part one received standard antero-posterior (AP), lateral and long leg radiographs postoperatively 5 days after surgery and full weight bearing long leg radiographs after 3 months. The accuracy of the procedure was defined as the delta (∂) between planned correction and postoperative alignment (∂ HKA, ∂ MPTA, ∂ LDFA / Fig. 5) measured on full weight bearing long leg radiographs at 3 months follow-up. After the surgery, the total radiation exposure due to intraoperative fluoroscopy was recorded in cGy*cm2. For study arm part two the accuracy of the procedure was defined as the ∂ between planned correction and postoperative CT scans (∂ MPTA, ∂ LDFA). All measurements were performed by one orthopedic senior consultant, one orthopedic resident and one experienced radiologist.

Fig. 5
figure 5

The accuracy of the procedure was defined as the delta (∂) between planned correction and postoperative alignment measured on full weight bearing long leg radiographs (∂ HKA A + C, ∂ MPTA B + D, ∂ LDFA B + D)

Statistical analysis

The statistical analysis was performed with GraphPad Prism 7 (GraphPad Inc., San Diego, CA, USA). To calculate differences between the mean values, the student´s t-test was used. To calculate differences between the medians, the Wilcoxon-u-test was used. The level of significance was set to 0.05. To determine the inter observer reliability (IOR), the interclass correlation coefficient (ICC) was calculated. A post hoc analysis was made computing achieved power. The effect side was calculated with the means and sample size of ∂ MPTA and ∂ LDFA. Alpha was set to 0.05. The post hoc power for ∂ MPTA and ∂ LDFA was 0.99 and 0.08, respectively.

Results

Part 1: After in- and exclusion criteria: Three patients were excluded due to missing long leg x-rays postoperatively, while two patients with derotating osteotomies were excluded as well. In total, 19 PSI based osteotomies were included into this retrospective radiological analysis (Table 1). 11 extremities were treated with an OW-HTO, while 4 extremities were treated with a CW-DFO and 2 extremities received an OW-DFO. 2 extremities received a double-level osteotomy (OW-HTO + OW-DFO). The mean age was 42.6 years (SD ± 13.7), there were 13 male and 6 female patients. The mean BMI was 27.5 kg/m2 (SD ± 4.1).

Table 1 Fluoroscopic data of all patients with the preoperative and postoperative alignment as well as the planned parameter and the absolute deviation between the postoperative radiological outcome and the planned values

The mean planned MPTA and LDFA were 89.31° (SD ± 1.3°) and 89° (SD ± 1.19°), respectively. The mean postoperative MPTA and LDFA was 89.87° (SD ± 1.51°) and 89.28° (SD ± 2.42°), respectively. The mean ∂ between planned HKA and postoperative HKA was 1.45° ± 1.16°, the mean absolute ∂ between planned MPTA and postoperative MTPA was 0.86° ± 0.6° and the mean absolute ∂ between planned LDFA and postoperative LDFA was 1.98° ± 1.33°. The mean radiation exposure per procedure was 33.28° ± 24.89 cGy*cm2. All findings are summarized in Table 1. The ICC for part 1 is 0.962 (95% CI 0.915 – 0.994). All surgeries were performed without complications such as hinge fractures.

Part 2: All surgeries were performed without complications such as a hinge fracture. The mean ∂ between planned MPTA and postoperative MTPA was 3.47° ± 1.07° and the mean ∂ between planned LDFA and postoperative LDFA was 2.18° ± 1.9°. All findings are summarized in Table 2. The ICC for part 2 is 0.912 (95% CI 0.865 – 0.975).

Table 2 Fluoroscopic data of all specimens with the preoperative and postoperative alignment as well as the absolute deviation between the postoperative radiological outcome and the planned correction of 5° for the HTO and DFO

The mean ∂ MPTA was significantly different between clinical patients and cadaveric specimens (P < 0.001, 95% CI: -3.4 to -1.9 / Table 3). The mean ∂ LDFA was not significantly different between clinical patients and cadaveric specimens (p = 0,813, 95% CI: -2 to 1.6 / Table 3).

Table 3 Statistical analysis of the absolute delta between the mean and median of the planned alignment and the radiological outcome

Discussion

The most important finding of this study is the fact that the clinical use of PSI guides for correction osteotomies around the knee delivers a reliable accuracy under fluoroscopic control in the hands of low volume osteotomy surgeons. Thus, there is no accuracy for the learning curve. In contrast to that, omitting fluoroscopic control consequences in a lack of accuracy.

The clinical part of our study showed a high degree of accuracy in OW-HTO in the hand of a low volume osteotomy surgeon. Our results with a mean ∂ MPTA of 0.86° are comparable to the findings reported by Chaouche et al. who evaluated a high degree of accuracy for PSI guides in OW-HTO. They reported a mean ∂ HKA and ∂ MPTA of 1° ± 0.95° and 0.54° ± 0.63° in a cohort of a hundred patients. Furthermore, no complications specific to PSI were observed [5]. These values are comparable to recently published studies that showed no learning curve in terms of accuracy for PSI based osteotomies around the knee [15, 21].

However, publications on DFO osteotomies using PSI cutting blocks are rare as well [14, 27]. Two studies reported data concerning the accuracy of PSI guides in opening wedge distal femur varization. Jacquet et al. reported a significantly improved accuracy concerning the postoperative alignment compared to a conventional group (0.43° ± 0.50° vs 3.95° ± 1.64°) [14]. Elattar et al. [10] and Victor et al. [31] presented similar findings for OW-DFO and considered PSI guides as reliable tools for this procedure whereas Shi et al. reported a high degree of accuracy for CW-DFO in 33 knees as well.

One potential benefit of a PSI block for osteotomies around the knee is a guided placement of a K-wire in order to protect the lateral hinge during the procedure (Figs. 2 and 3). A hinge fracture incidence of about 30% is reported throughout the literature [7, 29]. A stable hinge is considered as a crucial part in osteotomies around the knee since the hinge is important for preservation of the correction as well as the bony consolidation of the osteotomy [16]. Practically, by intersecting the K-wire into the cutting plane at the planned hinge location, the cutting depth would be limited [6]. Further, the K-wire helps to preserve the lateral hinge during the opening of the osteotomy. Dessyn et al. [6] evaluated this effect in a biomechanical approach testing fresh frozen cadavers with and without a PSI guided K-wire for lateral hinge protection. Their biomechanical study revealed that the maximum load to breakage and the maximum permissible displacement were, respectively, 880% and 260% higher during the opening of the OWHTO by using K-wires compared to the non-K-wire control group. This confirms the mechanical advantage of using a K-wire for both stabilization and protecting the hinge during OW-HTO. Equal clinical findings were recently published by Gulagaci et al. who demonstrated that during OW-HTO a K-wire in the lateral hinge location reduced the occurrence of intraoperative lateral hinge fractures [11]. These findings are comparable to our results since there were no hinge fractures seen in both parts of our study.

Besides the clinical accuracy, the second research question was to evaluate the use of PSI cutting guides without intraoperative fluoroscopy like PSI blocks for TKA. We evaluated a mean dose area product (DAP) of about 33 cGy*cm2 and about 18 fluoroscopic images per patient within our clinical evaluation. The mean DAP of a standard thorax p.a. x-ray is 9 cGy*cm2 and of a lower spine x-ray is 140 cGy*cm2 [25]. To evaluate the actual radiation effect for the patient the effective dose has to be considered. However, this can be neglected for the knee joint due to the very low conversion factor to the effective dose (0,1 µSv/cGy*cm2) [13]. More important is the radiation dose reduction for the surgical team. Furthermore, no other study focusing on PSI guided osteotomies reported the intraoperative fluoroscopy radiation dose before. Thus, a comparison to current literature is not possible. However, Jacquet et al. reported a mean of 4.3 fluoroscopy images after a 9 cases learning curve for a PSI guided technique [15]. Further, Gulagaci et al. [11] reported a mean of 5.1 fluoroscopic images per procedure in a series of 60 knees using the same PSI guides. Pérez-Mañanes et al. report a mean of 8 fluoroscopic images for HTO and Arnal-Burró et al. a mean of 6 fluoroscopic images for DFO per procedure with do-it-yourself cutting guides [2, 22]. One explanation for these results might be differences concerning the level of experience of the surgeon by using these specific PSI blocks.

The results of our cadaveric approach reveal that omitting intraoperative fluoroscopy leads to a lack of accuracy. The results of our cadaveric analysis are significantly different compared to Donnez et al. [9] whose cadaveric study, with the exact same approach as ours, revealed an accuracy of a ∂ MPTA of 0.2° compared to a mean ∂ MPTA of 3.47° in our cadaveric study. This significant difference might be explainable due to the fact that Donnez et al. removed all soft tissues except for the patellae tendon insertion. In contrast to that, we only removed soft tissue from the bone at the attachment points for the PSI guide in order to mimic the in vivo situation as best as possible (Fig. 5). While referencing only with respect to the bony landmarks without fluoroscopy the soft tissue irritation is in our point of view the main issue and leads to the minor accuracy of the cutting guides. Further, the conflict of approach size and minimal invasive surgery cannot be controlled without objective information like x-ray data.

Several study limitations have to be mentioned. Part one of this study is a retrospective analysis without any patient outcome measurements in a small series of patients. Further, no control group was analyzed. The retrospective approach might introduce a potential selection bias. We used postoperative long leg radiographs in order to calculate the preoperative to postoperative ∂ of the HKA, MPTA and LDFA. However, Boonen et al. reported a good validity comparing measurements of long leg radiographs to 3D CT-scans [4]. In part two a small number of specimens were used in order to evaluate the effect of no fluoroscopic imaging. Moreover, not all specimens had a sufficient deformity to be considered as candidates for an osteotomy around the knee. In order to reduce a planning bias, OW-HTOs were planned with an arbitrary of 5° valgus correction and OW-DFOs with an arbitrary of 5° varus correction. However, lateral hinge fractures may not have occurred due to the correction of only 5°. Further, the amount of the correction has not yet been studied for the accuracy of PSI blocks. Nevertheless, the ∂ of accuracy compared to current literature and our clinical cohort is significantly different.

Conclusion

The clinical use of PSI guides for correction osteotomies around the knee under fluoroscopic control is a safe procedure in the hand of low volume osteotomy surgeons. In contrast to that, omitting fluoroscopic control consequences in a lack of accuracy. The use of fluoroscopic control during PSI guided correction osteotomies around the knee is highly recommended.

References

  1. Amis AA (2013) Biomechanics of high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 21:197–205

    Article  Google Scholar 

  2. Arnal-Burró J, Pérez-Mañanes R, Gallo-Del-Valle E, Igualada-Blazquez C, Cuervas-Mons M, Vaquero-Martín J (2017) Three dimensional-printed patient-specific cutting guides for femoral varization osteotomy: Do it yourself. Knee 24:1359–1368

    Article  Google Scholar 

  3. Bae DK, Song SJ, Yoon KH (2009) Closed-wedge high tibial osteotomy using computer-assisted surgery compared to the conventional technique. J Bone Joint Surg Br 91:1164–1171

    Article  CAS  Google Scholar 

  4. Boonen B, Kerens B, Schotanus MG, Emans P, Jong B, Kort NP (2016) Inter-observer reliability of measurements performed on digital long-leg standing radiographs and assessment of validity compared to 3D CT-scan. Knee 23:20–24

    Article  CAS  Google Scholar 

  5. Chaouche S, Jacquet C, Fabre-Aubrespy M, Sharma A, Argenson JN, Parratte S et al (2019) Patient-specific cutting guides for open-wedge high tibial osteotomy: safety and accuracy analysis of a hundred patients continuous cohort. Int Orthop. https://doi.org/10.1007/s00264-019-04372-4

    Article  PubMed  Google Scholar 

  6. Dessyn E, Sharma A, Donnez M, Chabrand P, Ehlinger M, Argenson JN et al (2020) Adding a protective K-wire during opening high tibial osteotomy increases lateral hinge resistance to fracture. Knee Surg Sports Traumatol Arthrosc 28:751–758

    Article  Google Scholar 

  7. Dexel J, Fritzsche H, Beyer F, Harman MK, Lutzner J (2017) Open-wedge high tibial osteotomy: incidence of lateral cortex fractures and influence of fixation device on osteotomy healing. Knee Surg Sports Traumatol Arthrosc 25:832–837

    Article  Google Scholar 

  8. Dickschas J, Staubli A, Harrer J (2019) Osteotomies around the knee. Z Orthop Unfall 157:203–218

    Article  Google Scholar 

  9. Donnez M, Ollivier M, Munier M, Berton P, Podgorski JP, Chabrand P et al (2018) Are three-dimensional patient-specific cutting guides for open wedge high tibial osteotomy accurate? An in vitro study. J Orthop Surg Res 13:171

    Article  Google Scholar 

  10. Elattar O, Swarup I, Lam A, Nguyen J, Fragomen A, Rozbruch SR (2017) Open wedge distal femoral osteotomy: accuracy of correction and patient outcomes. HSS J 13:128–135

    Article  Google Scholar 

  11. Gulagaci F, Jacquet C, Ehlinger M, Sharma A, Kley K, Wilson A et al (2019) A protective hinge wire, intersecting the osteotomy plane, can reduce the occurrence of perioperative hinge fractures in medial opening wedge osteotomy. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-019-05806-7

    Article  PubMed  Google Scholar 

  12. Hankemeier S, Hufner T, Wang G, Kendoff D, Zeichen J, Zheng G et al (2006) Navigated open-wedge high tibial osteotomy: advantages and disadvantages compared to the conventional technique in a cadaver study. Knee Surg Sports Traumatol Arthrosc 14:917–921

    Article  CAS  Google Scholar 

  13. Hart D, Wall BF (2004) UK population dose from medical X-ray examinations. Eur J Radiol 50:285–291

    Article  CAS  Google Scholar 

  14. Jacquet C, Chan-Yu-Kin J, Sharma A, Argenson JN, Parratte S, Ollivier M (2018) More accurate correction using “patient-specific” cutting guides in opening wedge distal femur varization osteotomies. Int Orthop. https://doi.org/10.1007/s00264-018-4207-1

    Article  PubMed  Google Scholar 

  15. Jacquet C, Sharma A, Fabre M, Ehlinger M, Argenson JN, Parratte S et al (2019) Patient-specific high-tibial osteotomy’s ‘cutting-guides’ decrease operating time and the number of fluoroscopic images taken after a Brief Learning Curve. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-019-05637-6

    Article  PubMed  Google Scholar 

  16. Jo HS, Park JS, Byun JH, Lee YB, Choi YL, Cho SH et al (2018) The effects of different hinge positions on posterior tibial slope in medial open-wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 26:1851–1858

    Article  Google Scholar 

  17. Kim YC, Yang JH, Kim HJ, Tawonsawatruk T, Chang YS, Lee JS et al (2018) Distal femoral varus osteotomy for valgus arthritis of the knees: systematic review of open versus closed wedge osteotomy. Knee Surg Relat Res 30:3–16

    Article  Google Scholar 

  18. Lee DK, Kim KK, Ham CU, Yun ST, Kim BK, Oh KJ (2018) The learning curve for biplane medial open wedge high tibial osteotomy in 100 consecutive cases assessed using the cumulative summation method. Knee Surg Relat Res 30:303–310

    Article  Google Scholar 

  19. Lobenhoffer P, Agneskirchner JD (2003) Improvements in surgical technique of valgus high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 11:132–138

    Article  Google Scholar 

  20. Mannan A, Smith TO (2016) Favourable rotational alignment outcomes in PSI knee arthroplasty: a level 1 systematic review and meta-analysis. Knee 23:186–190

    Article  CAS  Google Scholar 

  21. Munier M, Donnez M, Ollivier M, Flecher X, Chabrand P, Argenson JN et al (2017) Can three-dimensional patient-specific cutting guides be used to achieve optimal correction for high tibial osteotomy? Pilot study. Orthop Traumatol Surg Res 103:245–250

    Article  CAS  Google Scholar 

  22. Pérez-Mañanes R, Burró JA, Manaute JR, Rodriguez FC, Martín JV (2016) 3D surgical printing cutting guides for open-wedge high tibial osteotomy: do it yourself. J Knee Surg 29:690–695

    Article  Google Scholar 

  23. Saragaglia D, Chedal-Bornu B, Rouchy RC, Rubens-Duval B, Mader R, Pailhe R (2016) Role of computer-assisted surgery in osteotomies around the knee. Knee Surg Sports Traumatol Arthrosc 24:3387–3395

    Article  CAS  Google Scholar 

  24. Saragaglia D, Roberts J (2005) Navigated osteotomies around the knee in 170 patients with osteoarthritis secondary to genu varum. Orthopedics 28:s1269-1274

    Article  Google Scholar 

  25. Schegerer A, Loose R, Heuser LJ, Brix G (2019) Diagnostic reference levels for diagnostic and interventional X-ray procedures in Germany: update and handling. Rofo 191:739–751

    Article  Google Scholar 

  26. Schroter S, Ihle C, Elson DW, Dobele S, Stockle U, Ateschrang A (2016) Surgical accuracy in high tibial osteotomy: coronal equivalence of computer navigation and gap measurement. Knee Surg Sports Traumatol Arthrosc 24:3410–3417

    Article  CAS  Google Scholar 

  27. Shi J, Lv W, Wang Y, Ma B, Cui W, Liu Z et al (2019) Three dimensional patient-specific printed cutting guides for closing-wedge distal femoral osteotomy. Int Orthop 43:619–624

    Article  Google Scholar 

  28. Staubli AE, Jacob HA (2010) Evolution of open-wedge high-tibial osteotomy: experience with a special angular stable device for internal fixation without interposition material. Int Orthop 34:167–172

    Article  Google Scholar 

  29. Takeuchi R, Ishikawa H, Kumagai K, Yamaguchi Y, Chiba N, Akamatsu Y et al (2012) Fractures around the lateral cortical hinge after a medial opening-wedge high tibial osteotomy: a new classification of lateral hinge fracture. Arthroscopy 28:85–94

    Article  Google Scholar 

  30. Van den Bempt M, Van Genechten W, Claes T, Claes S (2016) How accurately does high tibial osteotomy correct the mechanical axis of an arthritic varus knee? A systematic review. Knee 23:925–935

    Article  Google Scholar 

  31. Victor J, Premanathan A (2013) Virtual 3D planning and patient specific surgical guides for osteotomies around the knee: a feasibility and proof-of-concept study. Bone Joint J 95-b:153–158

    Article  CAS  Google Scholar 

  32. Yoon SD, Zhang G, Kim HJ, Lee BJ, Kyung HS (2016) Comparison of cable method and miniaci method using picture archiving and communication system in preoperative planning for open wedge high tibial osteotomy. Knee Surg Relat Res 28:283–288

    Article  Google Scholar 

Download references

Acknowledgements

ME, PS, and MP planned this study. PS and HH performed the experimental surgeries. MH, CF and PS conducted the radiographic analysis and statistical analysis. PS, MH, ME, HH and MP drafted the manuscript. All authors read and approved the final manuscript.

Funding

Open Access funding enabled and organized by Projekt DEAL. No funding was received for this study.

Author information

Authors and Affiliations

Authors

Contributions

The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Peter Savov.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the local ethic committee (#8294_BO_K_2019).

Competing interests

All authors declare that they have no conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Savov, P., Hold, M., Petri, M. et al. CT based PSI blocks for osteotomies around the knee provide accurate results when intraoperative imaging is used. J EXP ORTOP 8, 47 (2021). https://doi.org/10.1186/s40634-021-00357-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40634-021-00357-8

Keywords