Influence of peracetic acid-ethanol sterilisation on the biomechanical properties of human meniscus transplants

Abstract

Purpose

Meniscus allograft transplantation (MAT) is a possible treatment for patients suffering with pain after meniscectomy. Here, peracetic acid (PAA) sterilised meniscus transplants were investigated on whether they would provide an adequate alternative to fresh-frozen transplants in their viscoelastic and mechanical properties.

Methods

In this analysis, 31 menisci donors (26 male and 5 female) were included. The average donor age was 49.87 years, ranging from 32 to 65 years. Menisci of matched pairs of knees underwent chemical sterilisation while counterparts were left fresh-frozen. Stiffness and load to failure were determined via suture retention. Further menisci were analysed while attached to the tibial bone block using a novel test device to mimic physiological load distribution. Meniscus relaxation, stiffness and failure loads were determined. Histology and biphasic properties of the menisci were examined and results were analysed using paired t-tests.

Results

A novel custom built test device allowed the application of physiological loads for suture retention testing and revealed no significant differences between PAA sterilised (14.85 ± 4.46 N/mm, 50.49 ± 17.01 N) and fresh-frozen (18.26 ± 4.46 N/mm, 59.49 ± 21.07 N) regarding stiffness and failure load, respectively. Furthermore, initial 200 N loading showed significantly higher strain in sterilised menisci (18.87 ± 1.56) compared to fresh frozen (13.81 ± 1.04). Load relaxation experiments demonstrated significantly lower relaxation for sterilised menisci (77.71 ± 1.62) compared to fresh-frozen (89.11 ± 1.00, p-value < 0.0001).

Conclusion

Peracetic acid sterilised human menisci performed equally to fresh-frozen counterparts in a suture retention test and in physiological failure testing providing an adequate alternative. However, meniscus relaxation, biphasic properties and strain were shown to be significantly different between the groups. A common problem of MAT is graft extrusion or shrinkage, therefore the parameters measured here should be considered and may influence meniscus extrusion after transplantation.

Level of evidence

n/a (experimental study)

Introduction

The meniscus is an essential part of the complex knee mechanics as it acts as a shock absorber by resisting tibiofemoral pressure [19, 27]. It stabilizes the knee, improves congruency and further provides lubrication [17, 34, 52]. Severe meniscal injuries are often treated by partial or total meniscectomy. Mensicus allograft transplantation provides treatment for post-meniscectomy syndrome in the presence of a previous total or sub-total meniscectomy [48]. When left untreated, a meniscectomy is associated with early cartilage damage and in many cases leaving no other option than knee arthroplasty [13, 45]. Especially for young and active patients, who, with todays life expectancy, will undergo several cycles of revision arthroplasty, meniscus transplantation therefore provides a valuable alternative [33, 36, 37, 48].

Currently, three options are available for meniscal replacement such as the collagen meniscus implant (CMI) derived from a bovine collagen [20], Actifit, a polyurethane scaffold [28] and 3D printed scaffolds [16]. Additionally, tendon autografts are also used. However, for meniscus substitution, allografts have been used and historically supplied as fresh-frozen (FF), deep-frozen, cryopreserved or lyophilized grafts [38, 51].

Sterilization of allografts is important to ensure tissue safety and sterility. Two sterilisation methods are widely used to ensure tissue safety, gamma irradiation and ethylene oxide [6, 21, 24, 57]. Gamma irradiation is associated with negative effects on the biomechanical properties of meniscal tissue [6, 15, 49, 56, 57]. Ethylene oxide may induce an inflammatory response [24]. A more recent approach is supercritical CO2 which seems to outperform gamma irradiation with regards to biomechanical properties of meniscus transplants [5, 47]. However, validation of its sterility assurance level has not yet been confirmed.

Peracetic acid (PAA) ethanol sterilisation is a validated procedure to reliably eliminate bacterial, fungal and viral contaminations as well as spores [41, 42]. PAA sterilisation did not show an effect on the biomechanics of cortical bone [22]. Nevertheless, PAA sterilised cancellous bone has inferior biomechanical properties but can be successfully vitalized using mesenchymal stromal cells [43, 54].

In 2008, Scheffler and colleagues performed an in vivo sheep study comparing the remodelling and biomechanical behaviour of PAA sterilised and non-sterilised long flexor tendons for anterior cruciate ligament (ACL) reconstruction and analysed at 6 and 12 weeks [50]. The authors demonstrated that the remodelling of PAA-tendons was delayed. Hence, the failure properties of PAA-grafts were inferior to those of FF grafts at the given time points. Two major limitations of that study included the uncontrolled weight bearing of the animals and the selected early time points. In a previous study, the authors did not find any significant differences in the viscoelastic and mechanical properties of bone-patellar tendon-bone (BPTB) allografts when comparing PAA sterilised and fresh-frozen counterparts in vitro [51]. To the authors’ best knowledge, no study has focused on the PAA sterilisation of human meniscus transplants. Therefore, the aim of the present study is to investigate the influence of PAA sterilisation on the viscoelastic and mechanical properties of human meniscus transplants. We hypothesise that PAA sterilisation does not negatively affect the failure properties of meniscus transplants, but might display differences in the viscoelastic properties.

Material and methods

Donor tissue and specimen preparation

All allografts were provided by a non-profit tissue bank (German Institute for Cell and Tissue Replacement, DIZG gemeinnützige GmbH). All tissues are acquired from non-profit tissue recovery partners after informed consent. Sterilisation was performed using a validated, GMP-compliant process. Sterile allografts are approved as medicinal products under §21 of the German Medicinal Products Act (license number: PEI.H.03360.01.1). Knees from a single donor are thawed at 2–8 °C and menisci are exposed and visually evaluated after removing remnants of blood, fat, muscles and connective tissue. Previously injured menisci or knees showing signs of severe cartilage degeneration were excluded. The tibial plateau was cut 10–15 mm distal to the articular surface using a band saw. One of the two menisci was cut close to the root attachments for further determination of suture retention and histology leaving the other still attached to the tibial plateau. As depicted in Fig. 1e the plateau was cut to 12 mm width and maximum 10 mm anterior height without compromising the meniscal structure. The posterior as well as the contralateral bone blocks were adjusted to equal heights. A total of 31 menisci pairs from 26 male and 5 female donors were included in this analysis. The average donor age was 49.87 years and ranged from 32 to 65 years. For the different testing parameters, the menisci were all analysed pairwise.

Fig. 1
figure 1

A suture loop (a) separates a collagen bundle with a cross section of 0.5 mm2. The collagen bundle was channeled through a plate with an elongated hole that the meniscus was pulled against thereby loading the collagen bundle only. A novel test device shown in (b) and (c) was constructed to mimic physiological load distribution on menisci, which are still attached to the tibial bone. Hemispheres were chosen according to the meniscus’ size and pulled down for load application. d shows a meniscus still attached to a bone block. The distance between the inner meniscal vertex and the bone block is referred to as L0. The width of the bone block is 12 mm. e Equal anterior (ha) and posterior (hb) heights of the bone block were ensured during the preparation process

Peracetic acid-ethanol sterilisation

All menisci pairs were thawed in their packaging and placed in containers filled with sterile physiological saline solution. Menisci were left to thaw for at least 30 min at room temperature. Once thawed one underwent sterilisation, the other was kept at room temperature. Both groups contained almost equal numbers of right and left menisci to account for side specific differences. For sterilisation, tissues were fully submerged in validated tissue-preserving sterilisation solution (2% peracetic acid, 96% ethanol, aqua ad iniectabilia; ratio v/v/v | 2/1/1) and incubated under constant agitation, at low pressure and at room temperature for 4 h. Subsequently, tissues were rinsed in a washing process using aqua ad iniectabilia. All menisci were stored at − 30 °C for further analysis.

Suture retention

Suture retention tests were performed with 13 pairs of menisci (mean age 50.38 years, range 37–63). Suture loops were placed around equal meniscus cross-sections (0.5 mm2) using a 0.6 mm suture material (see Fig. 1a). The collagen bundles were pulled to failure through an elongated hole providing a barrier for the rest of the meniscus as previously performed with a collagen meniscus implant (CMI) [17]. A 200 N ElectroForce® TestBench (Bose, USA) was used to determine stiffness and failure load (preload: 10 N, speed: 0.05 mm/s).

Physiological testing

Pairwise physiological testing was performed with 22 pairs of menisci (mean age 50.59 years, range 32–65). A total of 14 medial and 8 lateral menisci pairs were tested pairwise to their counterparts. A newly developed test device (Fig. 1b and c) was constructed to mimic physiological load distribution on the femoral condyle contact area. Prior to the mechanical testing, the menisci were thawed at 35 ± 2 °C for at least 15 min followed by measuring the distance between meniscal vertex and the bone block (L0) using a caliper (Fig. 1d). Bone blocks were then fixed and hemispheres of matched size were placed between the meniscal vertex and the bone block. Mechanical tests were performed at 35 ± 2 °C in 0.9% saline solution. Temperature was confirmed using a Hamster data logger (Elpro, Germany).

Mechanical tests were conducted in three stages using a Zwick Z010 test machine (Zwick, Germany). First, the menisci were loaded to 200 N at 2%/s applying a preload of 5 N. The crossbar’s positions at 5 and 200 N were recorded and used as the lower and upper limit during preconditioning, respectively. Menisci were preconditioned 9 times at 0.1 Hz with the load-time-graph following a cosine function. At the peak of the 10th cycle, the position was held constant for 45 min. Load relaxation was determined.

After a recovery phase at 4 °C in 0.9% saline solution over night, 8 out of 22 meniscus pairs (men age 44 years, range 32–64) underwent a 10 cycle preconditioning according to day 1 and were tested again for their relaxation behaviour on 5 consecutive days. An Inspekt table blue 10 kN machine (Hegewald & Peschke, Germany) was used to perform repetitive tests. One to three days after relaxation testing, menisci were loaded to failure following a 10 cycle preconditioning similar to day 1. Stiffness and load to failure were assessed.

Moisture content

Moisture content of 18 menisci pairs composed of 9 medial and 9 lateral pairs (mean age 54.88 years, range 43–65) was evaluated by thermogravimetric analysis using the HX204 Moisture Analyser (Mettler Toledo, Germany). Similar segments of meniscus pairs were dissected resulting in 0.2 to 0.8 g specimens. Specimens were gently dabbed prior to the measurement using a lint-free tissue. After preheating the analyser to 100 °C, moisture content was assessed using a standard drying program, a temperature of 105 °C and a constant weight criterion of 0.1%/5 min.

Histological analysis

Meniscus segments of equal size and origin of 2 pairs of menisci (n = 4) were sent for histological evaluation (Morphisto GmbH, Germany). Briefly, fixed tissues were dehydrated and paraffin embedded. Longitudinal and cross sections of 5 μm thickness were hematoxylin and eosin (H&E) stained and mounted on slides. Microscopic analysis was performed using a DMIL microscope (Leica, Germany). Micrographs of similar zones were compared with regard to their general appearance and the presence of interspaces (lacunae).

Statistical analysis

Data are presented as mean ± standard error of mean (SEM). Statistical significance was set with p <  0.05. Gaussian distributions were confirmed by D’Agostino-Pearson tests and the data were analysed using paired student’s t-tests. (*: p ≤ 0.05; **: p ≤ 0.01; ***: p ≤ 0.001; ****: p ≤ 0.0001). Unless stated otherwise significant differences between the two groups are shown in brackets.

Results

Suture retention analysis

Pairwise comparison of 13 meniscus pairs revealed no significant difference between PAA sterilised (14.85 ± 4.46 N/mm, 50.49 ± 17.01 N) and fresh-frozen (18.26 ± 4.46 N/mm, 59.49 ± 21.07 N) menisci regarding stiffness and load to failure, respectively (Fig. 2).

Fig. 2
figure 2

Stiffness (a) and load to failure (b) determined by suture retention analysis of 13 pairs of menisci. No significant difference between peracetic acid (PAA) sterilised and fresh-frozen (FF) menisci could be demonstrated with regards to stiffness and load to failure

Physiological testing

Using a novel custom built test device that allows the application of physiological loads, 22 pairs of menisci were tested for viscoelastic and failure properties. Prior to biomechanical evaluation, meniscus size (L0) was determined. Menisci showed a significant size reduction after treatment with peracetic acid (18.30 ± 0.66 for PAA compared to FF 19.39 ± 0.65, p-value 0.0126) (Fig. 3a). Strain measurements at 200 N revealed a significantly higher strain for PAA (18.87 ± 1.56) treated menisci in comparison to the untreated controls (13.81 ± 1.04, p-value < 0.0001) (Fig. 3b). After the application of 200 N the menisci in the two groups displayed similar sizes, PAA (21.87 ± 0.69) and FF (21.73 ± 0.62) p-value 0.7360 (Fig. 3c). For detailed results see Table 1.

Fig. 3
figure 3

Twenty two pairs of menisci of which one meniscus was PAA treated and the other was left untreated (FF), were biomechanically analysed using a novel meniscus fixture. Prior to the testing, meniscus length was measured (a). Strain was measured and meniscal size was calculated after the application of a 200 N load (b and c). While keeping the strain constant for 45 min, the tissue starts to relax (d). On a subset of 8 menisci this testing was performed in 5 consecutive days (e). Two groups sharing the same character are not significantly different (p < 0.05)

Table 1 Results of physiological and biomechanical comparison of peracetic acid sterilised (PAA) versus untreated fresh frozen (FF) menisci. Twenty two meniscus pairs were examined with regards to size, strain and size at 200 N, relaxation properties as well as stiffness and load to failure. Values are presented as MEAN ± SEM

Relaxation testing assessed viscoelastic properties. Lower relaxation was observed for PAA treated menisci at day 1 of physiological testing (Fig. 3d). Additional testing of a subgroup of 8 menisci on 5 consecutive days confirmed the significant differences in relaxation, PAA (77.71 ± 1.62) and FF (89.11 ± 1.00) p-value < 0.0001 (Fig. 3e). Furthermore, PAA treated menisci showed a significantly lower relaxation on day 1 compared to the following days. No significant differences in either stiffness or failure load were demonstrated using the physiological test setup (Fig. 4). Here, failure always occurred through bony avulsion.

Fig. 4
figure 4

Meniscus pairs of 22 donors were analysed for failure properties. Menisci that were still attached to a tibial bone block, were fixed in a novel device to mimic physiological load distribution. Following 10 preconditioning cycles, stiffness (a) and load to failure (b) were determined comparing PAA sterilised menisci to their untreated controls

Moisture content

Moisture content was determined thermogravimetrically. PAA-sterilisation was found to lower the moisture content from 72.74 ± 2.56% to 68.47 ± 2.32% (p <  0.0001).

Histological analysis

H&E stained histological sections of paraffin embedded samples showed widened lacunae in PAA treated menisci (Fig. 5).

Fig. 5
figure 5

Representative longitudinal sections of PAA treated (a) and fresh-frozen (b) menisci are shown after H&E staining. The scale bar depicts 200 μm and n = 4. Evenly distributed lacunae were present in both types of menisci. Lacunae were wider in PAA treated tissue

Discussion

The meniscus has many biomechanical roles that vary from load transmission [53], stability [29, 52], joint lubrication [46], reduction of contact stress by providing an increased contact area [25, 58] and knee congruity [58]. Various factors can affect MAT performance, some examples are graft size, fixation technique, level of activity and age [1, 4, 11, 32, 44]. Analysis into different biomechanical parameters comparing PAA sterilised and non-sterilised menisci were conducted. In the present study, a biomechanical equivalence between PAA sterilised and non-sterilised menisci with regards to stiffness and load to failure in the two different test setups was demonstrated. Furthermore, suture retention tests are normally used to describe optimal meniscus repair techniques. Referring to root attachment fixation, sutures are loaded alongside the collagen fibers, leading to failure loads lower than 200 N [26, 35]. Here, the load was applied perpendicularly to the fiber’s orientation. Pretests (unpublished data) only showed suture failures after testing entire meniscus cross sections. Therefore, partial cross sections (0.5 mm2) were tested. Suture fixation is an important biomechanical parameter in successful meniscal repair and is required to withstand the different loads applied [39]. Failure loads of longitudinal repair techniques vary widely in the literature ranging from 57.8 to 161 N based on the experimental set up [2, 7, 31, 50]. The load to failure values presented here (PAA sterilised 50.49 ± 17.01 N and fresh-frozen 59.49 ± 21.07 N) are at the lower end compared to the values documented in the literature. However, differences in experimental set ups were observed and varied from porcine, sheep and calf models. The differences in influencing factors such as properties of the suture material, area of the sutured meniscal cross section and the mechanical test procedure itself complicate a thorough comparison of the suture retention results. Stiffness measurements conducted here displayed values of 14.85 ± 4.46 N/mm for PAA and 18.26 ± 4.46 N/mm for fresh- frozen allografts, similar to that reported in the literature ranging from (2.2–67.8 N/mm) [7, 31, 50].

Additional biomechanical testing was conducted and investigated the stiffness and load to failure of menisci. These parameters are of importance as stiffness is defined as the resistance to deformation under applied forces and is important for successful MAT [5]. Consistent with the suture retention results, stiffness and load to failure were not significantly different between PAA treated and untreated menisci using the novel physiological test setup. In 2010, Hauch et al. determined the failure properties of human meniscus root attachments showing slightly lower values for average load to failure of 500.6 ± 232.5 N and an almost identical stiffness of 180.5 ± 59.5 N/mm [23]. The group tested one root attachment at a time, whereas in the present study tensile stress was applied on both root attachments. Ellman et al. also investigated single meniscal root attachments biomechanically [12]. Ellman pointed out the role of tissue preparation. Resection of supplemental fibers has a negative effect on the failure properties of meniscal attachments. Native roots were tested with ultimate failure loads ranging from approximately 500 to 660 N. According to the observations of both Ellman and Hauch failure always occurred through bony avulsion as observed here in this study.

In 2005, Scheffler and colleagues tested PAA sterilised human BPTB grafts. No significant differences in stiffness and load to failure were detected between sterilised BPTBs and the untreated counterparts [51]. In contrast to the failure properties, evaluation of time dependent biomechanical properties showed an effect of PAA treatment. On the one hand, sterilised menisci were smaller than the untreated ones but on the other hand they stretched more when subjected to the same load of 200 N. The higher initial strain compensates for the size reduction. After calculation of the size of the menisci at the first load peak, there are no significant differences between PAA and untreated menisci indicating preserved structural integrity of PAA menisci.

Sterilised menisci relaxed less compared to fresh frozen menisci. This was confirmed on five consecutive days. According to Hauch and colleagues the relaxation could affect the position of the meniscus graft within the joint [23]. Meniscus extrusion could be enhanced in meniscal replacements showing high relaxation. Thus, meniscal extrusion is associated with knee malalignment, cartilage damage and meniscal tears [10]. An extrusion of more than 3 mm is hypothesized to facilitate biomechanical changes and osteoarthritis [8,9,10, 30, 37]. PAA treated menisci show less load relaxation after a 10 cycle preconditioning phase, suggesting superior extrusion properties compared to untreated menisci.

The meniscus, an extracellular matrix is mainly composed of water. It contains approximately 70% water (liquid phase) and a 30% solid phase. The solid phase mainly consists of collagenous proteins which are accompanied by a low percentage of proteoglycans such as glycosaminoglycans, which are key players with regards to viscoelastic properties of the tissue [3, 17]. These viscoelastic properties depend on the biphasic composition of the meniscal tissue [14]. They have water retention capability and high resistance to compressive loads [3, 18, 40]. Therefore, the moisture content was investigated with both lateral and medial menisci. In this study, the water content of PAA treated menisci was significantly lower compared to untreated menisci, which can account for the significantly lower and slower relaxation of PAA treated menisci revealed by relaxation tests.

A degradation of proteoglycans can explain both, the lower residual moisture and the altered viscoelastic properties of PAA sterilised menisci. However, 5% PAA does not reduce the glycosaminoglycan content of tendons [57].

The reduced water content might also be a consequence of the macroscopic structural changes seen in PAA sterilised tissues. In 2011, Woon et al. treated acellularized intrasynovial tendons with 5% PAA for 4 h to enhance reseeding capability. They too observed larger interspaces between the collagen bundles. PAA treated tendons showed greater reseeding, no cytotoxicity and were not inferior to untreated controls with regards to ultimate tensile stress and elastic modulus [59]. Vascular access channels generated by trephination are described as meniscal repair facilitators [3, 18]. The interspaces between collagen bundles of sterilised menisci found in this study might accelerate the reseeding process as they facilitate cell-to-tissue contact.

In contrast, Scheffler et al. performed biomechanical tests with PAA sterilised anterior cruciate ligament (ACL) constructs 6 and 12 weeks after transplantation into sheep. PAA treated ACL grafts were significantly weaker and less stiff [50]. In contrast to the menisci used in the present study, the tendons used in the Scheffler study were delipidised with chloroform, methanol and ultrasound.

The present study has certain limitations. Here, the physiological test device only accounted for one major meniscus transplantation technique, the bone block fixation but did not include other methods such as the soft tissue technique. Also, meniscus grafts will certainly be transplanted using the keyhole technique too. Aged related changes to the meniscus have been previously reported [55]. This study also included mensicus harvested from aged donors, however the pairwise comparisons conducted here have minimized any potential bias. Additionally the biomechanical testing conducted here cannot be directly compared to clinical success and would need to be investigated further. Differences in experiemental set up used here and in the literature make it difficult to directly compare this work. The present study is experimental and the positive effects of PAA sterilised menisci regarding meniscal extrusion would need to be clinically verified.

Conclusion

Human PAA sterilised menisci performed equally to fresh-frozen counterparts in a suture retention test and in physiological failure testing. Meniscus relaxation, biphasic properties and strain were shown to be significantly different between the two groups. Treated menisci displayed lower relaxation and moisture content and higher strain compared to untreated menisci. PAA sterilised menisci might exhibit properties positively influencing meniscus extrusion after transplantation.

Abbreviations

ACL:

Anterior Cruciate Ligament

BPTB:

Bone - Patellar Tendon – Bone

CMI:

Collagen Meniscus Implant

FF:

Fresh-Frozen

H&E:

Hematoxylin and eosin

PAA:

Peracetic Acid

References

  1. Alhalki MM, Howell SM, Hull ML (1999) How three methods for fixing a medial meniscal autograft affect tibial contact mechanics. Am J Sports Med 27:320–328

    Article  CAS  PubMed  Google Scholar 

  2. Aşık M, Şener N (2002) Failure strength of repair devices versus meniscus suturing techniques. Knee Surg Sports Traumatol Arthrosc 10:25–29

    Article  PubMed  Google Scholar 

  3. Brindle T, Nyland J, Johnson DL (2001) The meniscus: review of basic principles with application to surgery and rehabilitation. J Athl Train 36:160–169

    CAS  PubMed  PubMed Central  Google Scholar 

  4. Brophy RH, Matava MJ (2012) Surgical options for meniscal replacement. JAAOS-J Am Acad Orthopaedic Surg 20:265–272

    Article  Google Scholar 

  5. Bui D, Lovric V, Oliver R, Bertollo N, Broe D, Walsh WR (2015) Meniscal allograft sterilisation: effect on biomechanical and histological properties. Cell Tissue Bank 16:467–475

    Article  CAS  PubMed  Google Scholar 

  6. Campbell DG, Li P (1999) Sterilization of HIV with irradiation: relevance to infected bone allografts. Aust N Z J Surg 69:517–521

    Article  CAS  PubMed  Google Scholar 

  7. Chang JH, Shen HC, Huang GS, Pan RY, Wu CF, Lee CH et al (2009) A biomechanical comparison of all-inside meniscus repair techniques. J Surg Res 155:82–88

    Article  PubMed  Google Scholar 

  8. Choi CJ, Choi YJ, Lee JJ, Choi CH (2010) Magnetic resonance imaging evidence of meniscal extrusion in medial meniscus posterior root tear. Arthroscopy 26:1602–1606

    Article  PubMed  Google Scholar 

  9. Costa CR, Morrison WB, Carrino JA (2004) Medial meniscus extrusion on knee MRI: is extent associated with severity of degeneration or type of tear? AJR Am J Roentgenol 183:17–23

    Article  PubMed  Google Scholar 

  10. Crema MD, Roemer FW, Felson DT, Englund M, Wang K, Jarraya M et al (2012) Factors associated with meniscal extrusion in knees with or at risk for osteoarthritis: the multicenter osteoarthritis study. Radiology 264:494–503

    Article  PubMed  PubMed Central  Google Scholar 

  11. Dienst M, Greis PE, Ellis BJ, Bachus KN, Burks RT (2007) Effect of lateral meniscal allograft sizing on contact mechanics of the lateral tibial plateau: an experimental study in human cadaveric knee joints. Am J Sports Med 35:34–42

    Article  PubMed  Google Scholar 

  12. Ellman MB, LaPrade CM, Smith SD, Rasmussen MT, Engebretsen L, Wijdicks CA et al (2014) Structural properties of the meniscal roots. Am J Sports Med 42:1881–1887

    Article  PubMed  Google Scholar 

  13. Fairbank TJ (1948) Knee joint changes after meniscectomy. J Bone Joint Surg Br 30b:664–670

    Article  CAS  PubMed  Google Scholar 

  14. Favenesi J, Shaffer J, Mow V (1983) Biphasic mechanical properties of knee meniscus. Trans Orthop Res Soc 8:57

    Google Scholar 

  15. Fideler BM, Vangsness JRCT, Lu B, Orlando C, Moore T (1995) Gamma irradiation: effects on biomechanical properties of human bone-patellar tendon-bone allografts. Am J Sports Med 23:643–646

    Article  CAS  PubMed  Google Scholar 

  16. Filardo G, Petretta M, Cavallo C, Roseti L, Durante S, Albisinni U et al (2019) Patient-specific meniscus prototype based on 3D bioprinting of human cell-laden scaffold. Bone Joint Res 8:101–106

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  17. Fithian DC, Kelly MA, Mow VC (1990) Material properties and structure-function relationships in the menisci. Clin Orthop Relat Res 252:19–31

    Article  Google Scholar 

  18. Fox JM, Rintz KG, Ferkel RD (1993) Trephination of incomplete meniscal tears. Arthroscopy 9:451–455

    Article  CAS  PubMed  Google Scholar 

  19. Fukubayashi T, Kurosawa H (1980) The contact area and pressure distribution pattern of the knee. A study of normal and osteoarthrotic knee joints. Acta Orthop Scand 51:871–879

    Article  CAS  PubMed  Google Scholar 

  20. Ghodbane SA, Brzezinski A, Patel JM, Plaff WH, Marzano KN, Gatt CJ et al (2019) Partial meniscus replacement with a collagen-hyaluronan infused three-dimensional printed polymeric scaffold. Tissue Eng A 25:379–389

    Article  CAS  Google Scholar 

  21. Gibbons MJ, Butler DL, Grood ES, Bylski-Austrow DI, Levy MS, Noyes FR (1991) Effects of gamma irradiation on the initial mechanical and material properties of goat bone-patellar tendon-bone allografts. J Orthop Res 9:209–218

    Article  CAS  PubMed  Google Scholar 

  22. Haimi S, Vienonen A, Hirn M, Pelto M, Virtanen V, Suuronen R (2008) The effect of chemical cleansing procedures combined with peracetic acid-ethanol sterilization on biomechanical properties of cortical bone. Biologicals 36:99–104

    Article  CAS  PubMed  Google Scholar 

  23. Hauch KN, Villegas DF, Haut Donahue TL (2010) Geometry, time-dependent and failure properties of human meniscal attachments. J Biomech 43:463–468

    Article  PubMed  Google Scholar 

  24. Jackson DW, Windler GE, Simon TM (1990) Intraarticular reaction associated with the use of freeze-dried, ethylene oxide-sterilized bone-patella tendon-bone allografts in the reconstruction of the anterior cruciate ligament. Am J Sports Med 18:1–10 discussion 10-11

    Article  CAS  PubMed  Google Scholar 

  25. Kettelkamp DB, Jacobs AW (1972) Tibiofemoral contact area-determination and implications. JBJS 54:349–356

    Article  CAS  Google Scholar 

  26. Kopf S, Colvin AC, Muriuki M, Zhang X, Harner CD (2011) Meniscal root suturing techniques: implications for root fixation. Am J Sports Med 39:2141–2146

    Article  PubMed  Google Scholar 

  27. Kurosawa H, Fukubayashi T, Nakajima H (1980) Load-bearing mode of the knee joint: physical behavior of the knee joint with or without menisci. Clin Orthop Relat Res 149:283–290

  28. Leroy A, Beaufils P, Faivre B, Steltzlen C, Boisrenoult P, Pujol N (2017) Actifit® polyurethane meniscal scaffold: MRI and functional outcomes after a minimum follow-up of 5 years. Orthop Traumatol Surg Res 103:609–614

    Article  CAS  PubMed  Google Scholar 

  29. Levy IM, Torzilli P, Warren R (1982) The effect of medial meniscectomy on anterior-posterior motion of the knee. JBJS 64:883–888

    Article  CAS  Google Scholar 

  30. Magee T (2008) MR findings of meniscal extrusion correlated with arthroscopy. J Magn Reson Imaging 28:466–470

    Article  PubMed  Google Scholar 

  31. Masoudi A, Beamer BS, Harlow ER, Manoukian OS, Walley KC, Hertz B et al (2015) Biomechanical evaluation of an all-inside suture-based device for repairing longitudinal meniscal tears. Arthroscopy 31:428–434

    Article  PubMed  Google Scholar 

  32. Matava MJ (2007) Meniscal allograft transplantation: a systematic review. Clin Orthop Relat Res 455:142–157

    Article  PubMed  Google Scholar 

  33. McCormick F, Harris JD, Abrams GD, Hussey KE, Wilson H, Frank R et al (2014) Survival and reoperation rates after meniscal allograft transplantation: analysis of failures for 172 consecutive transplants at a minimum 2-year follow-up. Am J Sports Med 42:892–897

    Article  PubMed  Google Scholar 

  34. McDermott ID, Masouros SD, Amis AA (2008) Biomechanics of the menisci of the knee. Curr Orthop 22:193–201

    Article  Google Scholar 

  35. Mitchell R, Pitts R, Kim YM, Matava MJ (2016) Medial meniscal root avulsion: a biomechanical comparison of 4 different repair constructs. Arthroscopy 32:111–119

    Article  PubMed  Google Scholar 

  36. Moens K, Dhollander A, Moens P, Verdonk K, Verdonk R, Almqvist KF et al (2014) Meniscal transplantation: still experimental surgery? A review. Acta Orthop Belg 80:403–413

    CAS  PubMed  Google Scholar 

  37. Noyes FR, Barber-Westin SD (2015) A systematic review of the incidence and clinical significance of postoperative meniscus transplant extrusion. Knee Surg Sports Traumatol Arthrosc 23:290–302

    Article  PubMed  Google Scholar 

  38. Peters G, Wirth CJ (2003) The current state of meniscal allograft transplantation and replacement. Knee 10:19–31

    Article  PubMed  Google Scholar 

  39. Post WR, Akers SR, Kish V (1997) Load to failure of common meniscal repair techniques: effects of suture technique and suture material. Arthroscopy 13:731–736

    Article  CAS  PubMed  Google Scholar 

  40. Proctor CS, Schmidt MB, Whipple RR, Kelly MA, Mow VC (1989) Material properties of the normal medial bovine meniscus. J Orthop Res 7:771–782

    Article  CAS  PubMed  Google Scholar 

  41. Pruss A, Baumann B, Seibold M, Kao M, Tintelnot K, von Versen R et al (2001) Validation of the sterilization procedure of allogeneic avital bone transplants using peracetic acid-ethanol. Biologicals 29:59–66

    Article  CAS  PubMed  Google Scholar 

  42. Pruss A, Göbel UB, Pauli G, Kao M, Seibold M, Mönig HJ et al (2003) Peracetic acid-ethanol treatment of allogeneic avital bone tissue transplants--a reliable sterilization method. Ann Transplant 8:34–42

    PubMed  Google Scholar 

  43. Rauh J, Despang F, Baas J, Liebers C, Pruss A, Gelinsky M et al (2014) Comparative biomechanical and microstructural analysis of native versus peracetic acid-ethanol treated cancellous bone graft. Biomed Res Int 2014:784702

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  44. Rodeo SA (2001) Meniscal allografts—where do we stand? Am J Sports Med 29:246–261

    Article  CAS  PubMed  Google Scholar 

  45. Rongen JJ, Rovers MM, van Tienen TG, Buma P, Hannink G (2017) Increased risk for knee replacement surgery after arthroscopic surgery for degenerative meniscal tears: a multi-center longitudinal observational study using data from the osteoarthritis initiative. Osteoarthr Cartil 25:23–29

    Article  CAS  Google Scholar 

  46. Rosvold JM, Darcy SP, Peterson RC, Achari Y, Corr DT, Marchuk LL et al (2011) Technical issues in using robots to reproduce joint specific gait. J Biomech Eng 1:133

  47. Russell N, Rives A, Pelletier MH, Wang T, Walsh WR (2015) The effect of supercritical carbon dioxide sterilization on the anisotropy of bovine cortical bone. Cell Tissue Bank 16:109–121

    Article  CAS  PubMed  Google Scholar 

  48. Saltzman BM, Bajaj S, Salata M, Daley EL, Strauss E, Verma N et al (2012) Prospective long-term evaluation of meniscal allograft transplantation procedure: a minimum of 7-year follow-up. J Knee Surg 25:165–175

    Article  PubMed  Google Scholar 

  49. Samitier G, Alentorn-Geli E, Taylor DC, Rill B, Lock T, Moutzouros V et al (2015) Meniscal allograft transplantation. Part 1: systematic review of graft biology, graft shrinkage, graft extrusion, graft sizing, and graft fixation. Knee Surg Sports Traumatol Arthrosc 23:310–322

    Article  PubMed  Google Scholar 

  50. Scheffler SU, Gonnermann J, Kamp J, Przybilla D, Pruss A (2008) Remodeling of ACL allografts is inhibited by peracetic acid sterilization. Clin Orthop Relat Res 466:1810–1818

    Article  PubMed  PubMed Central  Google Scholar 

  51. Scheffler SU, Scherler J, Pruss A, von Versen R, Weiler A (2005) Biomechanical comparison of human bone-patellar tendon-bone grafts after sterilization with peracetic acid ethanol. Cell Tissue Bank 6:109–115

    Article  CAS  PubMed  Google Scholar 

  52. Shoemaker SC, Markolf KL (1986) The role of the meniscus in the anterior-posterior stability of the loaded anterior cruciate-deficient knee. Effects of partial versus total excision. J Bone Joint Surg Am 68:71–79

    Article  CAS  PubMed  Google Scholar 

  53. Shrive N, O'connor J, Goodfellow J (1978) Load-bearing in the knee joint. Clin Orthop Relat Res 131:279–287

  54. Stiehler M, Seib FP, Rauh J, Goedecke A, Werner C, Bornhauser M et al (2010) Cancellous bone allograft seeded with human mesenchymal stromal cells: a potential good manufacturing practice-grade tool for the regeneration of bone defects. Cytotherapy 12:658–668

    Article  CAS  PubMed  Google Scholar 

  55. Tsujii A, Nakamura N, Horibe S (2017) Age-related changes in the knee meniscus. Knee 24:1262–1270

    Article  PubMed  Google Scholar 

  56. Vangsness CT Jr, Dellamaggiora RD (2009) Current safety sterilization and tissue banking issues for soft tissue allografts. Clin Sports Med 28:183–189 vii

    Article  PubMed  PubMed Central  Google Scholar 

  57. Vangsness CT Jr, Garcia IA, Mills CR, Kainer MA, Roberts MR, Moore TM (2003) Allograft transplantation in the knee: tissue regulation, procurement, processing, and sterilization. Am J Sports Med 31:474–481

    Article  PubMed  Google Scholar 

  58. Walker PS, Erkman MJ (1975) The role of the menisci in force transmission across the knee. Clin Orthop Relat Res 109:184–192

  59. Woon CY, Pridgen BC, Kraus A, Bari S, Pham H, Chang J (2011) Optimization of human tendon tissue engineering: peracetic acid oxidation for enhanced reseeding of acellularized intrasynovial tendon. Plast Reconstr Surg 127:1107–1117

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgments

We thank Dr. Arnd Hoburg for his valuable contribution regarding the design and construction of the physiological test system. Our acknowledgement also go to Prof. Britt Wildemann for her scientific advice and Mr. Jan-Erik Ode for his guidance and support during the first set of experiments at his lab.

Author information

Authors and Affiliations

Authors

Contributions

VE: experimental design, conduction of study, analysis of results, writing the manuscript. JG: conduction of the study. JB: experimental design, analysis of results. NA: review of manuscript. The authors read and approved the final manusript.

Corresponding author

Correspondence to Norus Ahmed.

Ethics declarations

Ethics approval and consent to participate

Tissue grafts in Germany are classed and regulated as medicinal products under the German Medicinal Products Act (Arzneimittelgesetz, AMG, federal law, Federal Republic of Germany). All the tissues were acquired from non-profit tissue donation organizations after informed consent according to the German AMG and TPG (Transplantation Act, Transplantationsgesetz, federal law, Federal Republic of Germany) regulations. Stringent serological screening is then followed by tissue preparation under controlled conditions. The tissue is then subjected to a validated sterilization process.

Consent for publication

All authors read and approved the final manuscript.

Competing interests

The authors VE, JG and NA, JCB are employees of the German Institute for Cell and Tissue Replacement (DIZG gemeinnützige GmbH), a non-profit provider of sterile allografts.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions