Skip to main content

Biomechanical comparison of meniscus-suture constructs for pullout repair of medial meniscus posterior root tears

Abstract

Background

Transtibial pullout repair for posterior meniscus root tear is widely performed to restore meniscal function. However, it is sometimes technically difficult to pass the suture through the posterior medial meniscus root in narrow joint space. To address this limitation, a new suture technique using an all-inside meniscal suture device through the tibial tunnel was proposed. The purpose of the present study was to compare the biomechanical properties of a meniscus-suture construct prepared using an all-inside meniscal suture device and those of the construct prepared using conventional suture techniques.

Methods

A total of 18 fresh-frozen porcine medial menisci were used and randomly divided into three groups according to the type of suturing technique applied. Three suturing methods were evaluated: suturing with all-inside meniscal suture device, single simple suture, and double simple sutures. All specimens were subjected to cyclic loading of 300 cycles followed by a load-to-failure test. The displacement after cyclic loading, the ultimate failure load, and the mode of failure were evaluated.

Results

There was no significant difference among the three suturing techniques regarding both displacement after cyclic loading and ultimate failure load. Suture breakage was the most common failure mode in each group.

Conclusions

The biomechanical properties of meniscus-suture construct with the all-inside meniscal suture device were equivalent to those obtained using conventional suture techniques. Our results suggest that pullout repair using the all-inside meniscal suture device through the tibial tunnel could serve as an alternative suture technique for the repair of posterior meniscus root tears.

Background

The meniscus plays an important role in load distribution and joint stability. To maintain a healthy articular cartilage, it is crucial to maintain the hoop strain mechanism and circumferential integrity of the meniscus. Once a medial meniscus root tear occurs, the contact pressure in the medial compartment increases to the level similar to that observed following total meniscectomy (Allaire et al., 2008; Padalecki et al., 2014). Thus, meniscus root tears should be repaired anatomically to preserve meniscal function and prevent subsequent cartilage degeneration (Johannsen et al., 2012).

Meniscus root repair is widely performed to restore meniscal function (Feucht et al., 2015a; LaPrade & LaPrade, 2014; LaPrade et al., 2017; Lee et al., 2014a). While both transtibial pullout repair and suture anchor repair are well-established techniques for the management of posterior meniscus root tears, transtibial pullout repair is more commonly used and has better clinical outcomes (Ahn et al., 2007; Lee et al., 2014b; Petersen et al., 2014). However, it is sometimes technically demanding to pass the suture through the posterior medial meniscus root in a narrow joint space. To address this limitation, we propose a new suture technique that involves using an all-inside meniscal suture device through the tibial tunnel. First, the tibial tunnel is created in the same fashion as for the conventional pullout repair. Then, the all-inside meniscal suture device is inserted into the meniscus through the tibial tunnel, and repair is conducted entirely within the joint space. While this novel technique is advantageous in that it enables repairing the meniscus even if the joint space is narrow, the biomechanical properties of this suture method remains unclear.

In the present study, we aimed to compare the biomechanical properties of meniscus-suture constructs using an all-inside meniscal suture device against those obtained using conventional suture techniques. We hypothesized that the mechanical properties of meniscus-suture constructs using an all-inside meniscal suture device would be similar to those using conventional suture techniques.

Methods

Study design

In this study, we used porcine menisci to evaluate the mechanical properties of meniscus-suture construct, which represents a well-established practice in biomechanical research (Feucht et al., 2015b; Forkel et al., 2017). This study was approved by our institution’s internal review board. All applicable international, national, and institutional guidelines for the care and use of animals were followed.

Specimen preparation

A total of 18 fresh-frozen porcine medial menisci were used. The mean age of the pigs at the time the specimens were obtained was 24 weeks (age range, 23–25 weeks). All specimens were kept frozen at − 20 °C and thawed at 4 °C for 24 h in a refrigerator before biomechanical testing. After removing the patella, patellar tendon, muscles, and ligaments, the medial meniscus was sharply resected from the bony attachment to the tibia at the anterior and posterior root with a scalpel and completely detached from the tibial insertion. The specimens were randomly divided into three groups (six specimens per group) according to the type of suture technique applied.

Suturing technique

A No. 2 polyester suture was whip-stitched to the anterior horn of each medial meniscus. Three suturing techniques for the medial meniscus posterior root were tested. To minimize the effect of other factors on the comparison of suturing techniques, we used the same suture materials and the same number of suture strands to pull the meniscus in each specimen.

Six specimens were sutured using the single simple suture technique (single simple suture group), according to the following protocol: one 2–0 Ultra High Molecular Weight Polyethylene suture (Ultrabraid®, Smith & Nephew Endoscopy, Andover, MA, USA) was passed through the medial meniscus posterior horn 5-mm away from the posterior edge of the specimen (LaPrade et al., 2015a) (Fig. 1a).

Fig. 1
figure 1

Suture techniques. (a) Single simple suture group. A 2–0 Ultrabraid suture was passed through the medial meniscus posterior horn 5-mm away from the posterior edge. (b) Double simple sutures group. Two separate 2–0 Ultrabraid sutures were passed through the medial meniscus posterior horn 5 mm away from the posterior edge of the specimen, with a distance of 5 mm between the two sutures. (c) All-inside meniscal suture device group (1) Superior side. The first PEEK anchors of the two FasT-Fix 360® devices were installed through the medial meniscus posterior horn 5-mm away from the posterior edge, with a distance of 5 mm between the two anchors. (2) Inferior side. Before removal of remaining anchors

Another six specimens were sutured at the medial meniscus posterior horn using the double simple suture technique (double simple suture group), according to the following protocol: two separate 2–0 Ultrabraid® sutures were passed through the medial meniscus posterior horn 5 mm away from the posterior edge of the specimen, with a distance of 5 mm between the two sutures. Each suture was fastened and one end of each suture was cut to use two strands in the testing, while all four strands were used in the double simple suture in the previous study (Rosslenbroich et al., 2013) (Fig. 1b).

The remaining six specimens were sutured at the medial meniscus posterior horn using two all-inside meniscal suture device with 2–0 Ultrabraid® sutures (FasT-Fix 360®; Smith & Nephew Endoscopy, Andover, MA, USA) (all-inside suture device group), according to the following protocol: the first polyether ether ketone (PEEK) anchors of the two FasT-Fix 360® devices were installed through the medial meniscus posterior horn 5-mm away from the posterior edge of the specimen, with a distance of 5 mm between the two anchors, while both of the remaining anchors were cut and removed (Fig. 1c). We used the two FasT-Fix 360® devices to match the number of strands in the single simple suture and double simple suture groups.

Biomechanical testing

Biomechanical testing was conducted on a material testing machine (AUTOGRAPH AG-IS; SHIMADZU, Kyoto, Japan). The sutures stitched the anterior horn were fastened to the upper clamp. The sutures introduced to the posterior horn were tightened over the metal jig with Endobutton® (Smith & Nephew Endoscopy, Andover, MA, USA) (Fig. 2).

Fig. 2
figure 2

Set-up for biomechanical testing. The sutures stitched the anterior horn were fastened to the upper clamp. The sutures introduced to the posterior horn were tightened over the metal jig

One dot was marked on the meniscus and the other was marked on the suture introduced to the posterior horn. Afterwards, a cyclic loading test was performed at a loading force between 5 and 20 N and a rate of 200 mm/min for 300 cycles. Following the cyclic loading test, a load-to-failure test was performed at a rate of 5 mm/min. These procedures were recorded using a video recorder (HDR-CX370V; SONY, Tokyo, Japan). Instead of measuring the distance between the clamps, all measurements were performed using image analysis software (DIPP Motion Pro 2D; DITECT, Tokyo, Japan). This software automatically tracks and captures the dot. It can measure the distance between the dot marked on the meniscus and the reference line. Thus, we could measure the actual displacement caused by loading, excluding the effect of slippage between clamps as well as the stress-relaxation response of the soft tissue. For image analysis, the reference point was fixed at 10 mm below the meniscus edge at the display of the software and we measured the distance from the reference point to a small dot marked 3 mm anteriorly to the suture or anchor (Fig. 3). The displacement caused by a loading force of 5 N was defined as the difference between the distance measured at the first cycle and the distance measured at the last cycle. Additionally, the ultimate failure load and the mode of failure were recorded.

Fig. 3
figure 3

Measurement of displacement. One dot was marked on the meniscus (a). The reference point was fixed at 10 mm below the meniscus edge (b). The distance from the reference point to a small dot marked 3 mm anteriorly to the suture or anchor was measured

Statistical analysis

One-way analysis of variance (ANOVA) was performed to compare the overall effect of suture technique on displacement and ultimate failure load, followed by Tukey’s post-hoc test to detect significant differences. The Fisher’s exact test was used to evaluate the differences in mode of failure. P-values of less than 0.05 were considered to indicate statistical significance. Power analysis indicated that the minimum sample size necessary to detect a meaningful difference among three independent groups was six menisci per group.

Results

Displacement after cyclic loading

No suture breakage was observed during a cyclic loading test. There was no significant difference among the three suture techniques in terms of displacement after cyclic loading (Table 1).

Table 1 Displacement after cyclic load and ultimate failure load

Ultimate failure load

There was no significant difference among the three techniques regarding the ultimate failure load (Table 1).

Mode of failure

Suture breakage was observed in four specimens of the all-inside suture device and single simple suture groups as well as in five specimens of the double simple suture group. One specimen in the all-inside suture device group failed because of anchor pullout (Table 2). All the suture breakages occurred on the inner sutures in both the all-inside suture device and double simple suture groups. There was no significant difference among the three suture techniques regarding the mode of failure.

Table 2 Mode of failure

Discussion

The principal finding of the present study is that the biomechanical properties of meniscus-suture construct obtained using the all-inside meniscal suture device were equivalent to those obtained using the single simple suture and double simple suture techniques. Thus, the results of our study suggest that the pullout repair of the posterior meniscus root tear with the all-inside meniscal suture device through the tibial tunnel can serve as an alternative to the repair performed with conventional suture techniques.

The maximum failure load of the native medial meniscus posterior root has been reported at 359–678 N (Ellman et al., 2014; Kopf et al., 2011; Mitchell et al., 2016). Meniscal root repair with insufficient strength of meniscus-suture construct might result in failure before completion of the healing. Stärke et al. (Starke et al., 2013) clarified the tensile forces acting on the repaired medial meniscus posterior root under different loading conditions in cadaveric knees, reporting the highest mean tensile force at 60.1 N. Similarly, Kim and Joo (Kim & Joo, 2013) reported a mean pullout failure force of 71.6 N for simple vertical suture introduced to the medial meniscus posterior horn. In this context, one may conclude that no suture technique described in the present or previous studies can restore the native strength of the medial meniscus posterior root (Ellman et al., 2014). However, such biomechanical studies evaluated only time-zero stability (i.e., immediately after repair, before completion of the healing), and the minimal strength of the meniscus-suture construct required to ensure stability of the repaired medial meniscus posterior root. The meniscal healing process is likely affected not only by biomechanical factors related to the repair but also by biological factors intrinsic to the meniscus. Therefore, the repaired meniscal root may acquire enough fixation strength after completion of the healing if sufficient initial stabilization has been achieved.

Several suture techniques have been developed for the repair of the posterior meniscus root tear. Feucht et al. (Feucht et al., 2013) compared the biomechanical properties of meniscus-suture constructs among four different suture techniques currently used for transtibial pullout repair of posterior meniscus root tears. In this study, the modified Mason-Allen suture technique provided superior biomechanical properties of the meniscus-suture construct compared to those using double simple suture or horizontal mattress suture. On the other hand, LaPrade et al. (LaPrade et al., 2015b) reported that the modified Mason-Allen suture and the double simple suture were not significantly different in terms of the biomechanical properties. However, the authors admitted that the modified Mason-Allen suture is more technically demanding, and suggested that the simple suture technique might represent the clinical standard for posterior meniscus root repair because of its technical simplicity and acceptable biomechanical outcomes. Therefore, we compared the biomechanical properties of single or double simple suture and that of the all-inside meniscal suture device.

Rosslenbroich et al. (Rosslenbroich et al., 2013) compared the biomechanical properties of transtibial pullout repair using either single or double simple suture, and reported that ultimate failure load of the double suture was almost twice as high as that of the single suture. However, this previous study employed two strands for single simple suture and four strands for double simple sutures, whereas we compared three techniques employing the same number of sutures and the same suture materials, finding no significant difference regarding the ultimate failure load. Our results thus describe the effect of the suture technique itself. The discrepancy between our results and those of Rosslenbroich’s study may be due to the difference of the biomechanical properties of the suture materials themselves, suggesting that stronger suture materials may contribute to achieving superior mechanical properties of the repair.

During the load-to-failure test, most specimens in the all-inside suture device and double simple suture groups exhibited suture breakage but the two sutures were never broken. This may suggest that there exists an imbalance between the two strands, as the fiber arrangement of the meniscus was not parallel to the direction of the tensile force. In our study, the inner sutures always failed before the outer sutures, indicating that the inner suture was exposed to higher loads.

The present study has several limitations. First, we did not use the bone tunnel model, whereby the button-bone interface regulates loading displacement of the meniscus-suture complex in the repair construct (Cerminara et al., 2014). We just modified the bone tunnel model to use a metal fixation device for minimizing the deformity. Thus, we could evaluate the meniscus-suture interface directly. Second, we did not apply the transtibial technique in this study. This was done to arrange the condition of meniscus-suture constructs of each specimen and accurately evaluate the biomechanical properties as they have been in previous studies (Kopf et al., 2011; LaPrade et al., 2015b). Third, the direction of the suture pull was not physiological. Our biomechanical testing protocol was designed to simulate a worst-case scenario, in which the load to the meniscus was applied parallel to the meniscal fibers. This might not reflect in vivo loading of the meniscus, which also involves shear and compressive forces. However, we believe that this model is useful because it allowed us to evaluate the influence of the suture method independently from that of other factors. Fourth, the porcine meniscus was used instead of a human meniscus because most human cadavers available to us came from elderly donors and thus likely exhibited degenerative menisci; under these circumstances, using the menisci of adult, healthy pigs were preferable because it reduced the variation in biomechanical properties across different specimens, allowing us to focus on the effect of the suture technique itself.

In this porcine study, we found that the biomechanical properties of meniscus-suture construct with the all-inside meniscal suture device were equivalent to those with conventional suture techniques. Our results suggest that pullout repair using the all-inside meniscal suture device through the tibial tunnel could serve as an alternative suture technique for the repair of the posterior meniscus root.

References

  • Ahn JH, Wang JH, Yoo JC, Noh HK, Park JH (2007) A pull out suture for transection of the posterior horn of the medial meniscus: using a posterior trans-septal portal. Knee Surg Sports Traumatol Arthrosc 15:1510–1513

    Article  Google Scholar 

  • Allaire R, Muriuki M, Gilbertson L, Harner CD (2008) Biomechanical consequences of a tear of the posterior root of the medial meniscus. Similar to total meniscectomy. J Bone Joint Surg Am 90:1922–1931

    Article  Google Scholar 

  • Cerminara AJ, LaPrade CM, Smith SD, Ellman MB, Wijdicks CA, LaPrade RF (2014) Biomechanical evaluation of a transtibial pull-out meniscal root repair: challenging the bungee effect. Am J Sports Med 42:2988–2995

    Article  Google Scholar 

  • Ellman MB, LaPrade CM, Smith SD, Rasmussen MT, Engebretsen L, Wijdicks CA et al (2014) Structural properties of the meniscal roots. Am J Sports Med 42:1881–1887

    Article  Google Scholar 

  • Feucht MJ, Bigdon S, Mehl J, Bode G, Muller-Lantzsch C, Sudkamp NP et al (2015a) Risk factors for posterior lateral meniscus root tears in anterior cruciate ligament injuries. Knee Surg Sports Traumatol Arthrosc 23:140–145

    Article  Google Scholar 

  • Feucht MJ, Grande E, Brunhuber J, Burgkart R, Imhoff AB, Braun S (2013) Biomechanical evaluation of different suture techniques for arthroscopic transtibial pull-out repair of posterior medial meniscus root tears. Am J Sports Med 41:2784–2790

    Article  Google Scholar 

  • Feucht MJ, Grande E, Brunhuber J, Rosenstiel N, Burgkart R, Imhoff AB et al (2015b) Biomechanical evaluation of different suture materials for arthroscopic transtibial pull-out repair of posterior meniscus root tears. Knee Surg Sports Traumatol Arthrosc 23:132–139

    Article  Google Scholar 

  • Forkel P, Foehr P, Meyer JC, Herbst E, Petersen W, Brucker PU et al (2017) Biomechanical and viscoelastic properties of different posterior meniscal root fixation techniques. Knee Surg Sports Traumatol Arthrosc 25:403–410

    Article  Google Scholar 

  • Johannsen AM, Civitarese DM, Padalecki JR, Goldsmith MT, Wijdicks CA, LaPrade RF (2012) Qualitative and quantitative anatomic analysis of the posterior root attachments of the medial and lateral menisci. Am J Sports Med 40:2342–2347

    Article  Google Scholar 

  • Kim YM, Joo YB (2013) Pullout failure strength of the posterior horn of the medial meniscus with root ligament tear. Knee Surg Sports Traumatol Arthrosc 21:1546–1552

    Article  Google Scholar 

  • Kopf S, Colvin AC, Muriuki M, Zhang X, Harner CD (2011) Meniscal root suturing techniques: implications for root fixation. Am J Sports Med 39:2141–2146

    Article  Google Scholar 

  • LaPrade CM, LaPrade MD, Turnbull TL, Wijdicks CA, LaPrade RF (2015a) Biomechanical evaluation of the transtibial pull-out technique for posterior medial meniscal root repairs using 1 and 2 transtibial bone tunnels. Am J Sports Med 43:899–904

    Article  Google Scholar 

  • LaPrade RF, LaPrade CM, EW. J (2014) Recent advances in posterior meniscal root repair techniques. J Am Acad Orthop Surg 23:71–76

    Article  Google Scholar 

  • LaPrade RF, LaPrade CM, Ellman MB, Turnbull TL, Cerminara AJ, Wijdicks CA (2015b) Cyclic displacement after meniscal root repair fixation: a human biomechanical evaluation. Am J Sports Med 43:892–898

    Article  Google Scholar 

  • LaPrade RF, Matheny LM, Moulton SG, James EW, Dean CS (2017) Posterior meniscal root repairs: outcomes of an anatomic Transtibial pull-out technique. Am J Sports Med 45:884–891

    Article  Google Scholar 

  • Lee DW, Ha JK, Kim JG (2014a) Medial meniscus posterior root tear: a comprehensive review. Knee Surg Relat Res 26:125–134

    Article  Google Scholar 

  • Lee DW, Kim MK, Jang HS, Ha JK, Kim JG (2014b) Clinical and radiologic evaluation of arthroscopic medial meniscus root tear refixation: comparison of the modified Mason-Allen stitch and simple stitches. Arthroscopy 30:1439–1446

    Article  Google Scholar 

  • Mitchell R, Pitts R, Kim YM, Matava MJ (2016) Medial meniscal root avulsion: a biomechanical comparison of 4 different repair constructs. Arthroscopy 32:111–119

    Article  Google Scholar 

  • Padalecki JR, Jansson KS, Smith SD, Dornan GJ, Pierce CM, Wijdicks CA et al (2014) Biomechanical consequences of a complete radial tear adjacent to the medial meniscus posterior root attachment site: in situ pull-out repair restores derangement of joint mechanics. Am J Sports Med 42:699–707

    Article  Google Scholar 

  • Petersen W, Forkel P, Feucht MJ, Zantop T, Imhoff AB, Brucker PU (2014) Posterior root tear of the medial and lateral meniscus. Arch Orthop Trauma Surg 134:237–255

    Article  Google Scholar 

  • Rosslenbroich SB, Borgmann J, Herbort M, Raschke MJ, Petersen W, Zantop T (2013) Root tear of the meniscus: biomechanical evaluation of an arthroscopic refixation technique. Arch Orthop Trauma Surg 133:111–115

    Article  Google Scholar 

  • Starke C, Kopf S, Lippisch R, Lohmann CH, Becker R (2013) Tensile forces on repaired medial meniscal root tears. Arthroscopy 29:205–212

    Article  Google Scholar 

Download references

Acknowledgements

We would like to thank Editage (www.editage.jp) for English language editing.

Funding

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

Author information

Authors and Affiliations

Authors

Contributions

SO, TM and KS participated in the design of the study. SO, RI, YT and KN dissected the porcine knees. SO analyzed the data. TM and KS helped draft the manuscript. KS and TY unified this study. KS directed this study design. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Shinichiro Okimura.

Ethics declarations

Ethics approval and consent to participate

This study was approved by our institution’s internal review board. All applicable international, national, and institutional guidelines for the care and use of animals were followed.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Okimura, S., Mae, T., Tachibana, Y. et al. Biomechanical comparison of meniscus-suture constructs for pullout repair of medial meniscus posterior root tears. J EXP ORTOP 6, 17 (2019). https://doi.org/10.1186/s40634-019-0186-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40634-019-0186-4