Skip to main content

Table 2 Mean and standard deviation (SD) of the measured total micromotion between the femoral-component and the anterior and posterior femoral cortex for the different reconstructive techniques and load cases. P-values obtained from T-tests to test differences between means of the measured micromotions. Finite element (FE) model total micromotion results and relative difference (Dif.) to the measured

From: Femoral revision knee Arthroplasty with Metaphyseal sleeves: the use of a stem is not mandatory of a structural point of view

Load case

Cortex side

Experimental (DIC)

Finite element model

Total micromotion (μm)

P - values

Total micromotion (μm)

Femoral component

(A)

Sleeve (B)

Sleeve + Stem (C)

Femoral component (A) versus Sleeve (B)

Femoral component (A) versus Sleeve + Stem (C)

Femoral component (A)

Sleeve (B)

Sleeve + Stem (C)

Mean

SD

Mean

SD

Mean

SD

Value

Dif.

FE-DIC

Value

Error

FE-DIC

Value

Dif.

FE-DIC

Axial load

Anterior

10

6

7

3

7

3

NS

NS

11

+ 1

9

+ 2

8

+ 1

Posterior

39

12

35

8

32

5

NS

NS

42

+ 3

38

+ 3

30

−2

Internal-External moment

Anterior

12

6

9

4

8

3

NS

NS

10

−2

7

-2

9

+ 1

Posterior

10

3

7

3

8

3

NS

NS

12

+ 2

9

+ 2

7

−1

Axial load + Internal-External moment

Anterior

11

4

11

5

14

4

NS

NS

9

+ 2

10

−1

8

−6

Posterior

32

9

28

9

32

8

NS

NS

37

+ 5

33

+ 1

29

−7