Author | Year | Journal | Model | Fixed-loop devices | Adjustable-loop devices | Company | Sample size | Preconditioning | Cyclic loading | Pull-to-failure (mm/min) | Elongation during cyclic loading (mm) | Ultimate failure load (N) | ||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
No. of cycles | Applied load (N) | Freq. (Hz) | No. of cycles | Applied load (N) | Freq. (Hz) | |||||||||||
Min. Max. | Min. Max. | Mean ± SD | Mean ± SD | |||||||||||||
Current study | Device only | G-Lok | Stryker | 5 | 10 | 20 - 70 | 1 | 5000 | 20 - 520 | 1 | 20 | 1.46 ± 0.25 | 2178 ± 118 | |||
UltraButton | Smith & Nephew | 5 | 2.66 ± 0.28 | 1903 ± 81 | ||||||||||||
RigidLoop | DePuy | 5 | 1.51 ± 0.16 | 1835 ± 179 | ||||||||||||
ProCinch | Stryker | 5 | 1.60 ± 0.09 | 1456 ± 137 | ||||||||||||
Ahmad | 2004 | AJSM | Porcine construct | Endobutton | Smith & Nephew | 8 | 1000 | 50 - 250 | 1 | 20 cm/min | 1.75 ± 0.97 | 864 ± 164 | ||||
Barrow | 2014 | AJSM | Endobutton | Smith & Nephew | 6 | 10 | 10 - 50 | 1 | 4500 | 10 - 250 | 1 | 20 | 1.34 ± 0.03 | 1529 ± 26 | ||
ToggleLoc | Biomet | 6 | 5.76 ± 0.35 | 1652 ± 45 | ||||||||||||
Tightrope | Arthrex | 6 | 42.45 ± 7.01 | 809 ± 53 | ||||||||||||
Chang | 2018 | Arthroscopy | Device only | Endobutton | Smith & Nephew | 6 | 10 | 10 - 75 | 0.1 | 4500 | 100 - 400 | 0.5 | 50 | 0.74 | 1410 | |
Tightrope | Arthrex | 6 | 1.91 | 925 | ||||||||||||
Porcine construct | Endobutton | Smith & Nephew | 6 | 10 | 10 - 75 | 0.1 | 4500 | 100 - 400 | 0.5 | 50 | 14.88 | 843 | ||||
Tightrope | Arthrex | 6 | 15.65 | 888 | ||||||||||||
Conner | 2010 | Arthroscopy | Porcine construct | Endobutton | Smith & Nephew | 8 | 2000 | 50 - 450 | 1 | 30 mm/s | 3.55 ± 0.57 | 1191 ± 150 | ||||
ToggleLoc | Biomet | 8 | Data not given. | |||||||||||||
EZLoc | Biomet | 8 | 5.46 ± 0.95 | 913 ± 82 | ||||||||||||
Johnson | 2015 | AJSM | Endobutton | Smith & Nephew | 8 | 10 | 10 - 75 | 0.1 | 1000 | 100 - 400 | 0.5 | 50 | 1.05 ± 0.05 | 1530 ± 180 | ||
XO Button | ConMed Linvatec | 8 | 1.65 ± 0.43 | 2218 ± 114 | ||||||||||||
Rigidloop | DePuy | 8 | 1.09 ± 0.16 | 1976 ± 229 | ||||||||||||
Tightrope | Arthrex | 8 | 2.2 ± 0.62 | 784 ± 45 | ||||||||||||
ToggleLoc | Biomet | 8 | 3.69 ± 2.39 | 1995 ± 217 | ||||||||||||
Noonan | 2016 | Arthroscopy | Device only | Endobutton | Smith & Nephew | 5 | 25 | 50 - 250 | 4500 | 50 - 250 | 0.52 ± 0.08 | 1384 ± 71 | ||||
Tightrope | Arthrex | 5 | 0.96 ± 0.07 | 886 ± 39 | ||||||||||||
Porcine construct | Endobutton | Smith & Nephew | 5 | 25 | 50 - 250 | 1000 | 50 - 250 | 3 ± 0.30 | 866 ± 102 | |||||||
Tightrope | Arthrex | 5 | 2.7 ± 0.50 | 786 ± 166 | ||||||||||||
Petre | 2013 | AJSM | Device only | Endobutton | Smith & Nephew | 5 | 10 | 10 - 50 | 0.1 | 1000 | 50 - 250 | 0.5 | 50 | 0.11 ± 0.03 | 1456 ± 130 | |
XO Button | ConMed Linvatec | 5 | 0.35 ± 0.06 | 2230 ± 252 | ||||||||||||
Tightrope | Arthrex | 5 | 0.3 ± 0.04 | 841 ± 55 | ||||||||||||
ToggleLoc | Biomet | 5 | 0.82 ± 0.18 | 1561 ± 112 | ||||||||||||
Porcine construct | Endobutton | Smith & Nephew | 10 | 10 | 10 - 50 | 0.1 | 1000 | 50 - 250 | 0.5 | 50 | 1.88 ± 0.25 | 1456 ± 101 | ||||
XO Button | ConMed Linvatec | 10 | 1.82 ± 0.23 | 1748 ± 140 | ||||||||||||
Tightrope | Arthrex | 10 | 2.74 ± 0.39 | 859 ± 43 | ||||||||||||
ToggleLoc | Biomet | 10 | 3.34 ± 1.28 | 1334 ± 81 | ||||||||||||
Rylander | 2014 | Clinical Biomechanics | Porcine construct | G-Lok | Stryker | 10 | 1000 | 50 - 250 | 1 | 20 | 614 ± 176 | |||||
Endobutton | Smith & Nephew | 10 | 717 ± 128 | |||||||||||||
ToggleLoc | Biomet | 10 | 560 ± 101 | |||||||||||||
RetroButton | Arthrex | 10 | 526 ± 160 |