Skip to main content

Table 1 Elongation and load data for all cortical suspension devices

From: Biomechanical testing of fixed and adjustable femoral cortical suspension devices for ACL reconstruction under high loads and extended cyclic loading

 

Device type

Elongation during cyclic loading

(mm)

Elongation at failure

(mm)

Ultimate failure load

(N)

Mean ± SD

Range

Mean ± SD

Range

Mean ± SD

Range

G-Lok

Fixed-loop

1.46 ± 0.25

(1.02 – 1.65)

3.17 ± 0.33

(2.63 – 3.51)

2178 ± 118

(2075 – 2367)

UltraButton

Adjustable-loop

2.66 ± 0.28

(2.32 – 3.07)

4.20 ± 0.33

(3.75 – 4.53)

1903 ± 81

(1798 – 1998)

RigidLoop Adjustable

Adjustable-loop

1.51 ± 0.16

(1.38 – 1.71)

2.88 ± 0.20

(2.76 – 3.24)

1835 ± 179

(1529 – 1975)

ProCinch RT

Adjustable-loop

1.60 ± 0.09

(1.50 – 1.73)

2.78 ± 0.08

(2.70 – 2.92)

1456 ± 137

(1322 – 1668)

  1. Abbreviations: SD standard deviation