Table 4 Static tests summary: Displacements, valgus-malrotation of the tibial head and their corresponding crack and ultimate loads, including mean values and standard deviations (SD)

From: Static and fatigue strength of a novel anatomically contoured implant compared to five current open-wedge high tibial osteotomy plates

Specimen Crack / Ultimate load [kN] Medial displ. at crack/ ultimate load [mm] Lateral displ. at crack/ ultimate load [mm] valgus-malrotation at crack/ ultimate load (°) Lateral stiffness at crack/ ultimate load [kN/mm] Failure types
Activmotion 1 - / 8.9 - / 1.3 - / 2.5 - / 0.6 - / 3.6 2
Activmotion 2 3.7 / 7.5 0.7 / 2.1 2.6 / 5.1 0.9 / 1.4 1.4 / 1.5 1 and 2
 Mean: - / 8.2 - / 1.7 - / 3.8 - / 1.0 - / 2.6  
 SD ±: - / 1.0 - / 0.4 - / 1.3 - / 0.4 - / 1.1