Robotic-assisted total knee arthroplasty improves the rotational mismatch between femoral and tibial components, but not the forgotten joint score 12: a single-center retrospective cohort study

Abstract

Purpose

The primary aim of this study was to compare postoperative short-term patient reported outcome measurements (PROMs) and rotational mismatch between femoral and tibial following conventional jig-based total knee arthroplasty (Conv-TKA) versus robotic-assisted TKA (RA-TKA) using three-dimensional computed tomography (3DCT) measurements.

Methods

This retrospective, consecutive case–control trial included 83 patients with varus osteoarthritis of the knee undergoing Conv-TKA versus RA-TKA using bi-cruciate stabilized TKA. The rotational mismatch of the femoral and tibial components between the two groups were compared using 3DCT measurements. PROMs (2011 Knee Society Score (KSS), forgotten joint score-12 (FJS-12), patella score were compared in patients between 1 and 2 years postoperatively.

Results

The two groups did not exhibit significant differences in any of the following preoperative factors: age at surgery, body mass index (BMI), preoperative range of motion (ROM), hip-knee-ankle (HKA) angle. There were no significant differences in postoperative HKA angle and tibial rotation angle, the absolute values of the femoral rotational angle and rotational mismatch were significantly smaller in the RA-TKA group than in the Conv-TKA group (both p < 0.01). Neither Postoperative PROMs (2011 KSS: pain, patient satisfaction, patient expectation, advanced activities score) nor patella score differed significantly between the groups, but FJS-12 was significantly better in the Conv-TKA group than in the RA-TKA group (p < 0.01).

Conclusions

RA-TKA did not improve FJS-12 compared to Conv-TKA, but did result in more accurate rotational alignment of femoral component and rotational mismatch between the femoral and tibial components.

Level of evidence

IV.

Introduction

Total knee arthroplasty (TKA) is one of the most successful treatments for reducing pain and improving function in patients with end-stage knee osteoarthritis. Some national registries with long-term follow-up data have shown that TKA implants have a survival rate of over 90% at 10 years [1,2,3,4]. However, 15 to 20% patients are not satisfied with their new joint, and up to one-third of patients report that their joint does not feel normal after TKA [5, 6].

The keys to success in TKA are accurate osteotomy, mechanically alignment (MA) of the components, and good soft tissue balance. In 1976, Insall J et al. [7] first coined the terms flexion gap and extension gap. To achieve balance between these two gaps, they advocated the classic method of bone resection and aforementioned soft tissue releases.

In recent years, with advances in preoperative three-dimensional computed tomography (3DCT) planning [8], patient-matched-instruments [9], computed assisted surgery, and manual and digital soft tissue balancing tools, the first three elements can now be achieved more accurately. Manual soft tissue balancing tools such as balancer are difficult to use for freehand osteotomies where joint pressures are fine-tuned for measurement after osteotomy, while a digital soft tissue balancing tool using the VERASENSE® sensor-guided balancing technology (Orthosensor Inc) did not improve range of motion (ROM) and patient reported outcome measurements (PROMs) at 2 years postoperatively, despite increased operative time and cost, and is not recommended for routine use [10].

Recently, a variety of semiactive robotic-assisted (RA) systems for orthopaedic surgery (with or without images, and using different cutting systems and planning methods) have been promoted worldwide. As for component accuracy, RA surgical techniques have given surgeons intra-operative options to improve accuracy [11,12,13,14]. One of the most important features is that some RA systems show data about the gaps all over ROM.

Previous reports have improved the accuracy of implant placement but have not evaluated the rotational mismatch between femoral and tibial components in RA surgical technique.

Regarding patient reported outcome measurements (PROMs), a meta-analysis of outcome data demonstrated that robot-arm TKA (MAKO: robotic interactive orthopedic arm system, Stryker, Fort Lauderdale, Florida, USA) resulted in significantly better PROMs than conventional jig-based (Conv)-TKA after short- to mid-term follow up [15]. In contrast, RA-arm TKA resulted in a qualitatively higher Knee Society (KS) composite function score at 1 year postoperatively than Conv-TKA, although the difference was not statistically significant [16]. On the other hand, two studies of an image-free handheld RA-TKA (Blue Belt Navio surgical system., Navio, Smith & Nephew, Plymouth, MN, USA) found no significant differences between RA-TKA and Conv-TKA at 12 or 20 months after surgery for any of the following PROMs; the Short Form-12 score, Westren Ontario and MacMaster Universities Osteoarthritis Index (WOMAC) score and Knee Society Score (KSS) functional score [17, 18]. However, these studies did not use uniform prosthetic designs, and PROMs such as patient satisfaction and patient expectations were not adequately evaluated.

The purpose of this study was to compare the RM between femoral and tibial component and PROMs of patients who underwent an image-free handheld RA-TKA and Conv-TKA using bi-cruciate stabilized TKA (BCS) (Journey II BCS; Smith & Nephew. Inc. Memphis, TN, USA).

We hypothesized that RA-TKA improves RM and short-term postoperative PROMs more than Conv-TKA.

Methods

The study design was approved by the Ethics Review Committee (21010). All patients who participated provided written informed consent. Between 2019 and 2020, this retrospective case control study enrolled consecutive 53 patients who underwent TKA using an image-free handheld RA surgical system (Robot group) and between 2018 and 2019, this retrospective case control study enrolled consecutive 41 patients who performed TKA using a conventional manual surgical procedure (Manual group). The patients were not randomized and the same surgeon used the BCS prosthesis in both groups. The inclusion criteria were substantial pain, loss of function due to varus-type osteoarthritis of the knee and availability of complete data over 1 year of postoperative follow-up. Exclusion criteria included valgus-type osteoarthritis of the knee, rheumatoid arthritis, previous hip or knee arthroplasty surgery, severe bony defects requiring bone graft or augmentation, revision TKA, lumbar region problems, and active knee joint infection.

Preoperative patient demographics including age, gender, body mass index (BMI), and operation time, were closely matched in both groups. Matching criteria are summarized in Table 1.

Table 1 Preoperative patient demographics data

Surgical procedure

Conv-TKAs were performed using conventional manual surgical procedures with the measured resection technique. After inflating a tourniquet to 300 mmHg at the beginning of the procedure, a subvastus arthrotomy was performed. A distal femoral osteotomy was performed at the valgus angle of the femur; this angle was measured between the mechanical axis and the anatomical axis using an intramedullary resection guide during preoperative 3DCT planning of the entire lower extremity. Rotational alignment of femoral component was aligned parallel to the surgical epicondyle axis and perpendicular to the whiteside line. An anterior or posterior referencing technique was used for the anterior and posterior femoral cut. The size of the femoral component was determined based on the anteroposterior length of the femur. The medial joint gap in extension and at 90 degrees of flexion was kept constant, regardless of the size selected or the external rotation angle. If the flexion gap was loose, the femoral sizing guide could be moved posteriorly so that the anterior notch did not occur in the frontal cortical bone. If the flexion gap was too tight, the guide can be moved anteriorly. The differences between the extension gap and the flexion gaps of the medial compartment were reduced as much as possible. An extramedullary resection guide was used for proximal tibial osteotomy. The angle of the osteotomy was perpendicular to the mechanical axis, and the tibial posterior tilt had an 85-degrees orientation. The landmark used to determine the rotational alignment of the tibia was Akagi’s line, defined as a straight line from the middle of the posterior cruciate ligament and the medial border of the patellar tendon attachment site [19]. Furthermore, Akagi's Line was perpendicular to the surgical epicondylar axis with the knee in extension [20], which is important because the BCS-TKA results in a guided motion designed to induce medial pivot movement. Finally, the position of the tibial component was fine-tuned and determined by the ROM method [12, 21, 22] (Fig. 1a, c). Therefore, excessive external rotation alignment of the femoral component was avoided due to rotational mismatch.

Fig. 1
figure 1

We created a unique pinhole (black arrowhead) in the anteromedial base plate to perform this accuracy (a). The tibial fixation to determine the rotational alignment was similar in both Robot (b: right knee) and Manual groups (c: left knee). The determination of the rotational alignment of the tibial component used the range of motion (ROM) method, depending on the rotational alignment of the femoral component and soft tissue balancing [18, 23, 24]

The patella was resurfaced. In medial ligament balance in extension, we released the deep medial collateral ligament (d-MCL) and removed osteophytes within 1 cm of the joint line against osteophytes attached to the d-MCL, which affect the extension gap. The superficial layer of the MCL, semimembranosus, and posterior oblique ligament were not released.

Image-free handheld RA surgeries were performed using the Navio RA system [12, 18, 21, 25,26,27]. After inflating a tourniquet to 300 mmHg at the beginning of the procedure, a subvastus arthrotomy was performed. Planning of prosthesis position and bone resection was determined intraoperatively considering soft tissue balance. Osteophytes on the femur and tibia were resected. The tracker fixation of the femur was proximal and anterior to the medial epicondyle and mid-shaft femur, while that of the tibia extended from the anteromedial tibia to 6 cm from the wound using bury pin threads (4.0 mm). Landmark registration, ROM, varus-valgus laxity mapping, and anatomy of the femoral condyle and tibial plateau was mapped by ‘‘painting” the surfaces with an optical probe. A virtual model of the knee was thus created. The surgeon intraoperatively determined the volume of bone removal and planned the prosthesis size, alignment, and position.

Based on previous reports using a balancer, the femoral rotation position was determined so that the medial joint gap was almost constant from extension to 90 degrees of flexion and the lateral joint gap was loose at 90 degrees of flexion [21, 28]. Arthritic cartilage and bone were then methodically removed using the handheld sculptor. The fixation of the tibial component to determine the rotational alignment was similar in the Manual group, regardless of RA surgery (Fig. 1a, b). In this technique, the rotational alignment of the tibial component is determined through conformity to the femoral component when the knee is put through a series of full flexion–extension cycles [24]. The RA surgical technique continuously tracked the position of the patient’s lower limb and the progress of bone resection using a navigation system camera [12, 18, 21, 25,26,27].

The drainage tube was removed and physical therapy was initiated in both the Robot and Manual groups at first day after surgery. Full weight-bearing was not restricted and patients were allowed to walk with or without assistive devices. All the patients followed the same postoperative rehabilitation protocol.

Pre- and postoperative three- dimensional computed tomography images

The preoperative plans were developed using 3DCT data of the entire extremity in all cases. Using reference points, preoperative CT images were automatically fused to postoperative images by matching bone surfaces. After matching, the femoral-tibial component template from a computer- aided design model was manually superimposed on the implant image to match their contours.

Based on the positioning of the templates, the coronal and axial alignments of both components were measured with reference to the coordinate system using ZedView (ZedKnee; LEXI Co., Ltd., Tokyo, Japan). Postoperative CT scans were obtained at 4-weeks in both groups. The postoperative hip-knee-ankle (HKA) angle was also measured as the leg axis between the anatomical axis of the femur and tibia in the coronal plane defined by the coordinate systems. The postoperative 3DCT images of the femur and tibia were superimposed onto those of the preoperative 3DCT plan using ZedView software (Fig. 2) [12, 22, 26, 27].

Fig. 2
figure 2

The preoperative three-dimensional computed tomography image (3DCT) plan and a postoperative 3DCT image are shown. 3D computer-aided design data of femoral and tibial components were fit to the 3DCT image using six parameters, specifically the coronal, sagittal, and axial alignment of the femoral and tibial prostheses. The blue components indicate the preoperative plan image; the green and yellow components indicate the postoperative component

Rotational alignment of the femoral and tibial components relative to the bone landmarks and mismatch between the components were measured using 3DCT images. The femoral component rotational angle (+ : external rotation, -: internal rotation). Figure 3 was defined as the line perpendicular to the surgical epicondylar axis, the line between the lowest point of the medial epicondyle and the midpoint of the lateral epicondyle [29]. The tibial component rotational angle (+ : external rotation, -: internal rotation). Figure 3 was defined as Akagi’s line, the line between the center of the posterior cruciate ligament and the medial border of the tibial tuberosity [19]. On each postoperative 3DCT image, a positive value represents external rotation of the femoral and tibial components, while a negative value represents internal rotation. The absolute value of the angular divergence of the femoral component relative to the tibial component is defined as the rotational mismatch (Fig. 3) [30].

Fig. 3
figure 3

The postoperative three-dimensional computed tomography image (3DCT) image are shown. a Femoral component rotational angle. Axial 3DCT image of the left femur. The surgical epicondylar axis (SEA) connects the lowest point of the medial epicondyle to the midpoint of the lateral epicondyle. The prosthetic posterior condylar axis (PCA) connects the medial and lateral prosthetic posterior condylar surfaces. The femoral component rotational angle was defined as the angle between the SEA and the PCA. b, C Tibial component rotational angle. Axial 3DCT image of the left tibia. Akagi’s line connects the center of the posterior cruciate ligament and the medial border of the tibial tuberosity. The tibial component rotational angle was defined as the angle between the centerline of the tibial component and Akagi’s line. A positive value represents external rotation, and a negative value represents internal rotation of the femoral and tibial components. The absolute value of the angular divergence of the femoral component relative to the tibial component is defined as the rotational mismatch [27]. E/R; external rotation, I/R; internal rotation

A previous study showed that interclass correlation coefficients for 3DCT evaluations in the coronal, sagittal, and axial planes were 0.901, 0.899, and 0.881 for the femur and 0.924, 0.911, and 0.899 for the tibia, respectively. Intraclass correlation coefficients for the coronal, sagittal, and axial planes were 0.956, 0.903, and 0.878 for the femur and 0.918, 0.815, and 0.896 for the tibia, respectively [12, 27].

Patient reported outcome measurements

PROMs were assessed 1 and 2 years postoperatively using four sections (symptom, patient satisfaction, patient expectation and advanced activities) of the 2011 KSS [31], 12 items of the Forgotten Joint Score (FJS-12) [32], and the patella score [33].

Statistical analysis

Means and standard deviations were used to describe the data. Student’s t-test or the Wilcoxon nonparametric test were implemented. Categorical variables were compared using the Fisher exact test. All statistical analyses were performed with SPSS version 24.0 software (SPSS Inc, Chicago, IL). Statistical significance was set at a p value of less than 0.01. A power analysis using the G*Power 3 analysis program [34] was performed using an α error of 0.01 and a 1 − β error of 0.80 (Type II error is no more than 20%) to compare the means between the two groups, and it indicated that a sample of 26 knees was sufficient to detect differences between the RA and Manual groups.

Results

Preoperative patient demographics datas are shown in Tables 1 and 2. There were no significant differences between the two groups in terms of age, gender, BMI or follow up duration (Table 1). There were no significant differences between the two groups in terms of preoperative ROM, 1989 Knee Society Knee and Function score, and HKA angle (Table 2). The Robot group presented a significantly smaller rate of outliers for the femoral axial alignment than the Manual group (p < 0.01, Table 3). The absolute values in rotational mismatch were less in the Robot group than Manual group (p < 0.01, Table 3).

Table 2 Preoperative range of motion (ROM), Knee society 1989 (knee and function score), and hip-knee-ankle (HKA) angle
Table 3 The values in the femoral and tibial component rotational angle and absolute values in the rotational mismatch of the femoral and tibial component in postoperative 3DCT image

Postoperative ROM, 2011 KSS subscale scores (symptom, patient satisfaction, patient expectation, and advanced activities), FJS-12 score and patella score are shown in Table 4. Postoperative PROMs (symptom, patient satisfaction, patient expectation, advanced activities) and patella score were not significantly different between the two groups, but FJS-12 score was significantly higher in the Manual group than in RA group (p< 0.01).

Table 4 Postoperative range of motion (ROM), postoperative 2011 Knee society score (symptom, patient satisfaction, patient expectation and activity) and Patella score

Discussion

The main finding of the present study was (1) RA-TKA reduced the outliers in terms of rotational alignment of the femoral prosthesis and rotational mismatch between femoral and tibial components, compared to Conv-TKA, but (2) RA-TKA did not show better short-term postoperative improvement than Conv-TKA with respect to the FJS-12.

This is the first study to compare an image-free handheld RA-TKA and Conv-TKA in term of rotational mismatch using pre- and postoperative 3DCT data and detailed PROMs such as the 2011 KSS, FJS-12, and patella score. Previous reports comparing imageless RA TKA and Conv-TKA [17, 18] included Journey II BCS and Legion Posterior-Stabilized prostheses with different prosthetic designs in both groups. Regarding PROMs, only WOMAC, SF-12, 1989 Knee society functional score and 2011 Knee society score have not been examined. It is a strong point that this study was able to evaluate the postoperative alignments accurately with the 3DCT measurements using CAD software.

As for RA-TKA component accuracy attained using Navio, Bollars et al. reported that RA-TKA allowed the surgeon to accurately achieve the planned mechanical axis with significantly fewer outliers than Conv-TKA [35], and Navio TKA resulted in accurate alignment in more than 93% of cases [36]. Both Navio and CORI TKA (CORI surgical system., Smith & Nephew, Plymouth, MN, USA) demonstrated high levels of component alignment accuracy and ease of use [37]. All of these reports assessed the component accuracy of RA procedures using 2D radiographic measurements, and did not evaluate rotational alignment or rotational mismatch between femoral and tibial components. Several studies reported that despite the success of computer navigation and related technology in reducing outliers in coronal and sagittal prosthesis alignments, the error in axial rotation has not been reduced [38, 39]. Mahoney et al. revealed that the robotic-arm assisted TKA using MAKO (MAKO: robotic interactive orthopedic arm system, Stryker, Fort Lauderdale, Florida, USA) demonstrated greater accuracy for tibial component alignment, femoral component rotation and tibial slope and provided greater 3D accuracy to plan for various component positioning parameters [16].

In this study, the rotational angle for the femoral component was 2.4 ± 2.6° and 9.2 ± 2.6° for the RA group and Manual group cohorts, respectively (p < 0.01), and RA-TKA reduced the rotational mismatch outliers between femoral and tibial components compared to Conv-TKA (p < 0.01).

Kono et al. revealed that femoral component in low-PROMs group had more axial external rotation than did that in high-PROMs group after BCS-TKA [40]. We have previously reported that achieving tightness in the medial gap and looseness in the lateral gap at 90° of flexion results in improved postoperative Patient-Reported Outcome Measures (PROMs) when utilizing Journey II BCS with a balancer [21, 27]. Therefore, the RA surgical technique, which determines prosthetic alignment by considering the soft tissue envelope and incorporates an intraoperative joint-balancing procedure, has shown greater reduction in outliers in rotational alignment of the femoral component and less rotational mismatch between the femoral and tibial components compared to the Manual group.

Fujita et al. demonstrated that rotational mismatch between the femoral and tibial components and the postoperative clinical outcomes at one year postoperatively, including flexion angle, objective indicators, functional activity scores and total 2011 KSS score, were negatively correlated, and therefore when a BCS system is used for conventional TKA, surgeons should avoid excessive rotational mismatch [41].

In this study, the RA group did not show any improvement in FJS-12 scores at 1 to 2 years postoperatively compared to the Manual group. The reason was that several patients could not forget the discomfort of the belly pin thread insertion site; Navio and CORI used a thicker 4 mm bury pin threads, while the other semi-active RA techniques used a 3.2 mm bury pin thread (Table 5). East Asia have short height and small morphology, as a result, two patients developed intraoperative and postoperative iatrogenic fractures due to bury pin threads (Fig. 4).

Table 5 The thickness of berry pin thread thickness in semi-active robotic assisted surgery
Fig. 4
figure 4

Intra and postoperative fractures during robotic assisted total knee arthroplasty using bury pin threads (4.0 mm). A, B 78-year-old woman in cast immobilization for tibial tibial diaphysis fracture (white arrow) at 1 month after surgery, C, D 89-year-old woman with intraoperative fracture of medial condyle of femur (white arrow) with solid screw and anchor fixation. A, C AP view. B, D lateral view

Carlson et al. revealed that patients can expect marked improvement in the FJS-12 score during the first year after Conv-TKA, followed by slight continued improvement between 2 and 3 years, and a decline after 4 years [42]. In a multicenter study using Mako surgical system, Joo et al. reported that there were significant gradual improvements in PROMs from baseline preoperatively to 1–2 years and then to > 2 years of follow-up [43]. On the other hand, RA-TKA with Navio achieved a better FJS-12 score at 2 years postoperatively than Conv-TKA [23]. Based on these results, we expect that the results of this study will also show positive longitudinal changes in the FJS-12.

Our study has some limitations. First, the sample size was small and the follow-up period was short. In addition, the subjects enrolled only patients with primary osteoarthritis of the knee with varus deformity. Therefore, the results of the present study are not necessarily applicable to other diseases and deformities. Second, 3DCT measurements were taken in the supine position, so the lower limb alignment under weight-bearing conditions was not measured. In the pre- and postoperative supine position, a plate for dorsalis flexing of the ankle joint is placed on the sole of the foot [12, 22, 26, 27]. Fourth, when determining the soft tissue balance of individual patients in RA technique, the surgeon should be aware that the osteophytes of medial posterior femoral condyle and medial posterior tibial plateau will affect the postoperative soft tissue balance with MJG in extension and flexion. Future larger prospective studies with mid- and long-term PROMs are required to further substantiate our findings.

The clinical relevance of the present study is that validated 3DCT measurement showed that an image-free RA-TKA system reduced the outliers of rotational mismatch between femoral and tibial components, compared to Conv-TKA, but RA-TKA did not improve FJS-12 compared to Conv-TKA between 1 and 2 years postoperatively.

Availability of data and materials

The datasets used and analysed during the current study are available from the corresponding author.

References

  1. Australian Orthopaedic Association (2017) Hip, knee & shoulder arthroplasty: annual report 2017. Natl Jt Replace Regist S48-S50. https://aoanjrr.sahmri.com/documents/10180/397736/Hip%2CKnee%26ShoulderArthroplasty. Accessed 11 Oct 2018.

  2. Harris AI, Christen B, Malcorps JJ, O’Grady CP, Kopjar B, Sensiba PR, Vandenneucker H, Huang BK, Cates HE, Hur J, Marra DA (2019) Midterm performance of a guided-motion Bicruciate-stabilized total knee system: results from the international study of over 2000 consecutive primary total knee arthroplasties. J Arthroplasty 34(7S):S201–S208

    Article  PubMed  Google Scholar 

  3. National Joint Registry (2017) 14th annual report - national joint registry for England, Wales, Northern Ireland and the Isle of Man. vol 1821. http://www.njrreports.org.uk/Portals/0/PDFdownloads/NJR14thAnnualReport.pdf. Accessed 11 Oct 2018.

  4. Siddigi A, Levine BR, Springer BD (2022) Highlights of the 2021 American joint replacement registry annual report. Arthroplast Today 13:205–207

    Article  Google Scholar 

  5. Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KD (2009) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res 468:57–63

    Article  PubMed Central  Google Scholar 

  6. Coles LG, Gheduzzi S, Miles AW (2014) In vitro method for assessing the biomechanics of the patellofemoral joint following total knee arthroplasty. J Eng Med 228(12):1217–1226

    Article  CAS  Google Scholar 

  7. Insall JN, Ranawat CS, Scott WN, Walker P (1976) Total condylar knee replacement: preliminary report 1976. Clin Orthop 120:149–154

    Google Scholar 

  8. Miura M, Hagiwara S, Nakamura J, Wako Y, Kawarai Y, Ohtori S (2018) Interobserver and intraobserver reliability of computed tomography-based three-dimensional preoperative planning for primary total knee arthroplasty. J Arthroplasty 33:1572–1578

    Article  PubMed  Google Scholar 

  9. Pfitzner T, Abdel MP, von Roth P, Perka C, Hommel H (2014) Small improvements in mechanical axis alignment achieved with MRI versus CT-based patient-specific instruments in TKA: a randomized clinical trial. Clin Orthop Relat Res 472:2913–2922

    Article  PubMed  PubMed Central  Google Scholar 

  10. Sarpong NO, Held MB, Grosso MJ, Herndon CL, Santos W, Lakra A, Shah RP, Cooper HJ, Geller JA (2022) No benefit to sensor-guided balancing compared with freehand balancing in TKA: a randomized controlled trial. Clin Orthop Relat Res 480(8):1535–1544

    Article  PubMed  PubMed Central  Google Scholar 

  11. Hampp EL, Chughtai M, Scholl LY, Sodhi N, Stoker MB, Jacofsky DJ, Mont MA (2019) Robotic-arm assisted total knee arthroplasty demonstrated greater accuracy and precision to plan compared with manual techniques. J Knee Surg 32(3):239–250

    Article  PubMed  Google Scholar 

  12. Kaneko T, Igarashi T, Takada K, Yoshizawa S, Ikegami H, Musha Y (2021) Robotic-assisted total knee arthroplasty improves the outlier of rotational alignment of the tibial prosthesis using 3DCT measurements. Knee 31:64–76

    Article  PubMed  Google Scholar 

  13. Pietrzak JRT, Rowan FE, Kayani B, Donaldson MJ, Huq SS, Haddad FS (2019) Preoperative CT-based three-dimensional templating in robot-assisted total knee arthroplasty more accurately predicts implant sizes than two-dimensional templating. J Knee Surg 32(7):642–648

    Article  CAS  PubMed  Google Scholar 

  14. Savov P, Tuecking LR, Windhagen H, Ehmig J, Ettinger M (2021) Imageless robotic handpiece-assisted total knee arthroplasty: a learning curve analysis of surgical time and alignment accuracy. Arch Orthop Trauma Surg 141(12):2119–2128

    Article  PubMed  PubMed Central  Google Scholar 

  15. Zhang J, Ndou WS, Ng N, Gaston P, Simpson PM, Macpherson GJ, Patton JT, Clement ND (2022) Robotic-arm assisted total knee arthroplasty is associated with improved accuracy and patient reported outcome: a systematic reviewe and meta-analysis. Knee Surg Sports Traumatol Arthrosc 30(8):2677–2695

    Article  PubMed  Google Scholar 

  16. Mahoney O, Kinsey T, Sodhi N, Mont MA, Chen AF, Orozco F, Hozack W (2020) Improved component placement accuracy with robotic-arm assisted total knee arthroplasty. J Knee Surg 35(3):337–344

    PubMed  Google Scholar 

  17. Held MB, Grosso MJ, Gazgalis SNO, Boddapati V, Neuwirth A, Geller JA (2021) Improved compartment balancing using a robot-assisted total knee arthroplasty. Arthroplast Today 7:130–134

    Article  PubMed  PubMed Central  Google Scholar 

  18. Held MB, Gazgalis A, Neuwirth AL, Shah RP, John Cooper H, Geller JA (2022) Imageless robotic-assisted total knee arthroplasty leads to similar 24-month WOMAC scores as compared to conventional total knee arthroplasty: a retrospective cohort study. Knee Surg Sports Traumatol Arthrosc 8:2631–2638

    Article  Google Scholar 

  19. Akagi M, Oh M, Nonaka T, Tsujimoto H, Asano T, Hamanishi C (2004) An anteroposterior axis of the tibia for total knee arthroplasty. Clin Orthop Relat Res 420:213–219

    Article  Google Scholar 

  20. Akagi M, Mori S, Nishimura NA, Asano T, Hamanishi C (2005) Variability of extraarticular tibial rotation references for total knee arthroplasty. Clin Orthop Relat Res 436:172–176

    Article  Google Scholar 

  21. Kaneko T, Kono N, Mochizuki Y, Hada M, Toyoda T, Ikegami H, Musha Y (2017) Bi-cruciate substituting total knee arthroplasty improved medio-lateral instability in mid-flexion range. J Orthopaedics 14:201–206

    Article  PubMed  Google Scholar 

  22. Kaneko T, Kono N, Mochizuki Hada M, Sunakawa T, Ikegami H, Musha Y (2018) The influence of compressive forces across the patellofemoral joint on patient-reported outcome after bi-cruciate stabilized total knee arthroplasty. Bone Joint J 100-B(12):1585–1591

    Article  CAS  PubMed  Google Scholar 

  23. Eerens W, Bollars P, Henckes ME, Schotanus M, Mievis J, Janssen D (2022) Improved joint awareness two years after total knee arthroplasty with a handheld image-free robotic system. Acta Orthop Belg 88(1):47–52

    Article  PubMed  Google Scholar 

  24. Ikeuchi M, Yamanaka N, Okanoue Y, Ueta E, Tani T (2007) Determining the rotational alignment of the tibial component at total knee replacement: a comparison of two techniques. J Bone Joint Surg Br 89(1):45–49

    Article  CAS  PubMed  Google Scholar 

  25. Batailler C, White N, Ranaldi FM, Neyret P, Servien E, Lustig S (2019) Improved implant position and lower revision rate with robotic-assisted unicompartmental knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 27(4):1232–1240

    Article  PubMed  Google Scholar 

  26. Kaneko T, Igarashi T, Yoshizawa S, Takada K, Ikegami H, Musha Y (2021) Robotic-assisted total knee arthroplasty for distal femur fracture with lateral knee osteoarthritis. Case Rep Orthop 2021:5576955

    PubMed  PubMed Central  Google Scholar 

  27. Kaneko T, Yamamoto A, Takada K, Yoshizawa S (2022) Intraoperative joint balancing procedure using an imageless robotic assisted technique does not necessarily result in kinematically aligned bicruciate stabilized total knee arthroplasty. J Robot Surg 17(2):447–456

    Article  PubMed  Google Scholar 

  28. Inui H, Taketomi S, Yamagami R, Shirakawa N, Kawaguchi K, Tanaka S (2019) The relationship between soft-tissue balance and intraoperative kinematics of guided motion total knee arthroplasty. J Knee Surg 32:91–96

    Article  PubMed  Google Scholar 

  29. Berger RA, Rubash HE, Seel MJ, Thompson WH, Crossett LS (1993) Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop Relat Res 286:40–47

    Article  Google Scholar 

  30. Lützner J, Kirschner S, Günther KP, Harman MK (2012) Patients with no functional improvement after total knee arthroplasty show different kinematics. Int Orthop 36:1841–1847

    Article  PubMed  PubMed Central  Google Scholar 

  31. Scuderi GR, Bourne RB, Noble PC, Benjamin JB, Lonner JH, Scott WN (2012) The new knee society knee scoring system. Clin Orthop Relat Res 470(1):3–19

    Article  PubMed  Google Scholar 

  32. Behrend H, Giesinger K, Giesinger JM, Kuster MS (2012) The “forgotten joint” as the ultimate goal in joint arthroplasty: validation of a new patient- reported outcome measure. J Arthroplasty 27(3):430–436

    Article  PubMed  Google Scholar 

  33. Fellar JA, Bartlett RJ, Lang DM (1996) Patellar resurfacing versus retention in total knee arthroplasty. J Bone Joint Surg Br 78(2):226–228

    Article  Google Scholar 

  34. Faul F, Erdfelder E, Lang AG, Buchner A (2007) G*Power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods 39:175–191

    Article  PubMed  Google Scholar 

  35. Bollars P, Boeckxstaens A, Mievis J, Kalaai S, Schotanus MGM, Janssen D (2020) Preliminary experience with an image-free handheld robot for total knee arthroplasty: 77 cases compared with a matched control group. Eur J Orthop Surg Traumatol 30:723–729

    Article  CAS  PubMed  Google Scholar 

  36. Collins K, Agius PA, Fraval A, Petterwood J (2021) Initial experience with the NAVIO robotic-assisted total knee replacement-coronal alignment accuracy and the learning curve. J Knee Surg 35(2):1295–1300

    PubMed  Google Scholar 

  37. Sicat CS, Chow JC, Kaper B, Mitra R, Xie J, Schwarzkopf R (2021) Component placement accuracy in two generations of handheld robotics-assisted knee arthroplasty. Arch Orthop Trauma Surg 141:2059–2067

    Article  PubMed  Google Scholar 

  38. Lutzner J, Krummenauer F, Wolf C, Gunther KP, Kirschner S (2008) Computer assisted and conventional total knee replacement: a comparative, prospective, randomised study with radiological and CT evaluation. J Bone Joint Surg Br 90:1039–1044

    Article  CAS  PubMed  Google Scholar 

  39. Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C (2007) A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am 89:236–243

    Article  PubMed  Google Scholar 

  40. Kono K, Inui H, Tomita T, Yamazaki T, Taketomi S, Yamagami R, Kawaguchi K, Tanaka S (2022) The higher patient-reported outcome measure group had smaller external rotation of the femur in bicruciate-stabilized total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 30(4):1292–1299

    Article  PubMed  Google Scholar 

  41. Fujita M, Matsumoto T, Nakano N, Ishida K, Kuroda Y, Maeda T, Hayashi S, Kuroda R (2022) Rotational mismatch between femoral and tibial components should be avoided in JOURNEY II bi-cruciate stabilized total knee arthroplasty. Knee 38:69–75

    Article  PubMed  Google Scholar 

  42. Carlson VR, Post ZD, Orozco FR, Davis DM, Lutz RW, Ong AC (2018) When does the knee feel normal again: a cross-sectional study assessing the forgotten joint score in patients after total knee arthroplasty. J Arthroplasty 33(3):700–703

    Article  PubMed  Google Scholar 

  43. Joo PY, Chen AF, Richards J, Law TY, Taylor K, Marchand K, Roche M, Mont MA, Malkani AL (2022) Clinical results and patient-reported outcomes following robotic-assisted primary total knee arthroplasty: a multicentre study. Bone Jt Open 3(8):589–595

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Conflict of interest statement

The authors declare no competing interests. This author, their immediate family, and any research foundation with which they are affiliated did not receive any financial payments or other benefits from any commercial entity related to the subject of this article.

Ethical review committee statement

The hospital ethics committee approved the study protocol, and patients provided informed consent for participation in the study.

Funding

No founding were received.

Author information

Authors and Affiliations

Authors

Contributions

AY and TK authors have contributed to the development of the research questions and study design. KT and SY developed the first and subsequent drafts of the manuscript. All authors reviewed and approved the manuscript.

Corresponding author

Correspondence to Takao Kaneko.

Ethics declarations

Ethics approval and consent to participate

Not required.

Consent for publication

Not required.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yamamoto, A., Kaneko, T., Takada, K. et al. Robotic-assisted total knee arthroplasty improves the rotational mismatch between femoral and tibial components, but not the forgotten joint score 12: a single-center retrospective cohort study. J EXP ORTOP 10, 133 (2023). https://doi.org/10.1186/s40634-023-00705-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40634-023-00705-w

Keywords