Skip to main content

Table 2. Overview of change in calf circumference, the amount of saline infusion needed, the maximum pressurization, and the Intraclass Correlation Coefficient (ICC) with the respective confidence interval (CI) after running the decay model per needle type per post mortem human specimen

From: Applicability of devices available for the measurement of intracompartmental pressures: a cadaver study

 

Δ calf circumference (centimeters)

Saline infusion (ml)

Maximum Pressurization

ICC(95% Confidence Interval)*

>100 mmHg

<100 mmHg

Catheters

Slit Catheter, C2Dx

1

222

X

 

0.834 (0.593-0.932)c

Venflon Safety Pro, 22 gauge, BD

1

401

 

X

0.968 (0.898-0.988)b

Venflon Safety Pro, 20 gauge, BD

4

378

X

 

0.996 (0.984-0.999)a

Venflon Safety Pro, 18 gauge, BD

1

347

X

 

0.996 (0.990-0.998)a

Venflon Safety Pro, 17 gauge, BD

2

420

X

 

0.893 (0.809-0.948)b

Venflon Safety Pro, 16 gauge, BD

1

302

X

 

0.990 (0.969-0.996)a

Venflon Safety Pro, 14 gauge, BD

4

308

 

X

0.863 (0.696-0.941)c

Intranule, 18 gauge, Vygon

2

358

X

 

0.994 (0.976-0.998)a

Intranule, 16 gauge, Vygon

1

448

X

 

0.896 (0.607-0.965)c

Intranule, 14 gauge, Vygon

3

647

 

X

0.992 (0.984-0.997)a

Intranule, 13 gauge, Vygon

2

564

 

X

0.948 (0.897-0.976)b

Straight needles

Microlance needle, 25 gauge, BD

2

248

X

 

0.881 (0.790-0.943)b

Microlance needle, 23 gauge, BD

2

222

X

 

0.877 (0.765-0.943)b

Microlance needle, 21 gauge, BD

3

335

 

X

0.855 (0.717-0.934)c

Microlance needle, 18 gauge, BD

1

145

X

 

0.943 (0.732-0.982)c

Microlance needle, 16 gauge, BD

2

116

X

 

0.994 (0.977-0.998)a

Sonoplex facet tip, 22 gauge, Pajunk

3

347

X

 

0.838 (0.722-0.920)c

Sonoplex facet tip, 21 gauge, Pajunk

4

417

X

 

0.919 (0.840-0.963)b

Sonoplex facet tip, 20 gauge, Pajunk

0

234

X

 

0.978 (0.956-0.990)a

Side-ported needles

Side-ported needle, C2Dx

3

526

X

 

0.989 (0.973-0.995)a

Sonoplex Sprotte tip, Pajunk

1

157

X

 

0.890 (0.777-0.950)b

Side-ported needle, Stryker

1

370

X

 

0.998 (0.993-0.999)a

  1. * Excellent reliability with CI >0.900 is represented in a, good reliability with CI 0.900-0.750 in b, moderate reliability with CI 0.750-0.500 in c, and poor reliability with CI < 0.500 in red