Table 3 Total operative time (minutes) between the learning raTKA, mastered raTKA, and mTKA groups. Reported as least square (LS) mean and standard error (SE)

From: The initial learning curve for the ROSA® Knee System can be achieved in 6-11 cases for operative time and has similar 90-day complication rates with improved implant alignment compared to manual instrumentation in total knee arthroplasty

Comparison

LS mean (Std err.)

95% CI

P Value

Learning

102.40 (1.89)

[98.68; 106.12]

 

Mastered

86.51 (1.24)

[84.07; 88.95]

 

mTKA

73.34 (1.03)

[71.30; 75.38]

 

Difference

 Learning vs. Mastered

15.89 (2.26)

[11.44;20.34]

0.001

 Learning vs. mTKA

29.06 (2.15)

[24.81; 33.31]

0.001

 Mastered vs. mTKA

13.17 (1.61)

[9.99;16.35]

0.001