Skip to main content

Table 7 Summary of fatigue failure tests

From: Mechanical strength of a new plate compared to six previously tested opening wedge high tibial osteotomy implants

Specimens

Maximal load [N]

Vertical stiffness

KV [N/mm]

Lateral stiffness

KL [N/mm]

Number of cycles

Failure types

TomoFix std. 3

1280

1350

2000

>  60,000

2

TomoFix std. 4

1440

2000

2500

>  80,000

2

TomoFix std. 5

1760

2500

2200

>  120,000

2

PEEKPower 3

1440

2000

2500

>  80,000

2

PEEKPower 4

1280

1950

2140

>  60,000

2

PEEKPower 5

1440

2785

2250

>  80,000

2

iBalance 3

1760

4000

3600

>  120,000

2,4

iBalance 4

1760

3000

3400

>  120,000

2

iBalance 5

1920

3000

2952

>  140,000

2

iBalance 6

1760

3500

2500

>  120,000

1,2

TomoFix sm 3

1280

2200

2000

>  60,000

2,3

TomoFix sm 4

1280

1750

1500

>  60,000

2,3

TomoFix sm 5

1760

2000

2300

>  120,000

1,2

Contour Lock 3

2400

2100

4400

>  200,000

2

Contour Lock 4

1760

2300

2400

>  120,000

2

Contour Lock 5

2400

2700

2600

>  200,000

1,2,3

Activmotion 3

2240

2500

6300

>  180,000

2

Activmotion 4

2240

2500

2900

>  180,000

2

Activmotion 5

1600

2500

4750

>  100,000

2

Activmotion 6

1600

3100

5100

>  100,000

2

FlexitSystem 3

1760

1070

2050

>  120,000

2

FlexitSystem 4

1600

960

1630

>  100,000

2

FlexitSystem 5

1440

950

1580

>  80,000

2

FlexitSystem 6

1920

960

1730

>  140,000

2

  1. Maximal load, vertical & lateral stiffnesses, Min number of cycles (all values prior to failure) and failure types. The values of the groups I to VI were retrieved from our previous studies and reported here for the sake of comparison.