Skip to main content

Table 6 p-Values obtained from the t-tests comparing the previously tested implants to the FlexitSystem

From: Mechanical strength of a new plate compared to six previously tested opening wedge high tibial osteotomy implants

Groups

Ultimate load

Medial displ. at ultimate load [mm]

Lateral displ. at ultimate load [mm]

valgus-malrotation at ultimate load

Lateral stiffness at ultimate load

TomoFix std

>  0.05

>  0.05

>  0.05

<  0.05

>  0.05

PEEKPower

>  0.05

>  0.05

>  0.05

>  0.05

>  0.05

iBalance

>  0.05

>  0.05

>  0.05

>  0.05

>  0.05

TomoFix sm

>  0.05

>  0.05

>  0.05

>  0.05

>  0.05

Contour Lock

>  0.05

>  0.05

<  0.05

>  0.05

>  0.05

Activmotion

>  0.05

>  0.05

>  0.05

>  0.05

>  0.05

  1. Mean values were compared. All statistical tests were performed two sided. Statistical significance was considered at p <  0.05.