Table 6 p-Values obtained from the t-tests comparing the previously tested implants to the FlexitSystem

From: Mechanical strength of a new plate compared to six previously tested opening wedge high tibial osteotomy implants

GroupsUltimate loadMedial displ. at ultimate load [mm]Lateral displ. at ultimate load [mm]valgus-malrotation at ultimate loadLateral stiffness at ultimate load
TomoFix std>  0.05>  0.05>  0.05<  0.05>  0.05
PEEKPower>  0.05>  0.05>  0.05>  0.05>  0.05
iBalance>  0.05>  0.05>  0.05>  0.05>  0.05
TomoFix sm>  0.05>  0.05>  0.05>  0.05>  0.05
Contour Lock>  0.05>  0.05<  0.05>  0.05>  0.05
Activmotion>  0.05>  0.05>  0.05>  0.05>  0.05
  1. Mean values were compared. All statistical tests were performed two sided. Statistical significance was considered at p <  0.05.