Skip to main content

Table 1 Material Properties of Scaffold Designsa

From: Augmented repair of radial meniscus tear with biomimetic electrospun scaffold: an in vitro mechanical analysis

 

Aligned

Random

Biomimetic

Ultimate Stress (MPa)

 Parallel

12.9 ± 4.3*, ***

3.4 ± 1.1***

8.5 ± 1.9*, ***

 Perpendicular

1.2 ± 0.3****

3.8 ± 1.0***

5.1 ± 1.0***

Ultimate Strain (mm/mm)

 Parallel

0.40 ± 0.03*

3.28 ± 1.49****

0.34 ± 0.08

 Perpendicular

3.69 ± 1.30

3.00 ± 1.46

0.40 ± 0.07***

Modulus (MPa)

 Parallel

93.6 ± 33.9*, *****

16.9 ± 9.7

68.7 ± 14.7*, *****

 Perpendicular

2.7 ± 0.5***

18.5 ± 5.4***

39.4 ± 11.6****

Yield Stress (MPa)

 Parallel

4.9 ± 1.5*, *****

1.7 ± 0.5

4.0 ± 1.3**, ******

 Perpendicular

0.4 ± 0.1****

1.7 ± 0.4***

2.2 ± 0.4***

Yield Strain (mm/mm)

 Parallel

0.06 ± 0.01*****

0.11 ± 0.04

0.07 ± 0.01*****

 Perpendicular

0.12 ± 0.07

0.10 ± 0.04

0.07 ± 0.02

  1. aFor a given scaffold type, significant difference when comparing parallel vs. perpendicular direction, * p < 0.001, ** p < 0.05; Significantly different from both scaffolds, *** p < 0.05, **** p < 0.001; Significantly different from random scaffold, ***** p < 0.01, ****** p < 0.05