Skip to main content

Table 3 Axial and torsion failure test results

From: Biomechanical testing of distal femur osteotomy plate fixation techniques: the role of simulated physiological loading

Femur

NEW axial

OLD

NEW Torsion

OLD

1

4313.10

4240.50

31.00*

32.10*

2

4242.90

4193.50

33.40

37.20

3

4011.20

5002.10

29.50

26.50

4

3607.60

4084.80

  

Average

4043.70

4380.23

31.30

31.93

Significance

P = 0.27

 

P = 0.78

 

(p < 0.05 = sign)

   
  1. Results for the axial and torsion failure tests, axial force is in N/mm, torsion force is Nm/°; differences are not statistically significant.
  2. *Femurs in which the drive-shaft of the MTS came loose.
  3. ‘NEW’ = anatomical MDF plate, ‘OLD’ = antecedent MDF plate.