Table 2 Axial and torsion test results

From: Biomechanical testing of distal femur osteotomy plate fixation techniques: the role of simulated physiological loading

   

Total runs

Mean

SD

Axial tests

150 N

NEW plate

700

0.085651

0.054997

  

OLD plate

700

0.087737

0.056193

 

800 N

NEW plate

700

0.088803

0.048292

  

OLD plate

700

0.096411

0.044069*

Torsion tests

0 N

NEW plate

700

0.077263

0.022999

  

OLD plate

700

0.080334

0.025255*

 

150 N

NEW plate

700

0.076817

0.02146

  

OLD plate

700

0.078167

0.023

 

800 N

NEW Plate

700

0.076932

0.021548

  

OLD plate

700

0.077137

0.02275

  1. *Statistically significantly less motion with the anatomical MDF plate design.
  2. Results for the axial and torsion test runs, movement is in mm for the axial tests and in degrees for the torsion tests. ‘NEW plate’ = anatomical MDF plate, ‘OLD plate’ = antecedent MDF plate.